Variation in defibrillation modes after synchronized cardioversion: A potential hazard

被引:0
|
作者
Monsieurs, KG [1 ]
Coady, EM [1 ]
机构
[1] UNIV ANTWERP HOSP,UIA,DEPT INTENS CARE,B-2650 EDEGEM,BELGIUM
关键词
defibrillation; standardization; synchronized; cardioversion; training;
D O I
10.1016/S0300-9572(97)00019-1
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We conducted a survey of all defibrillators currently on the European market that offer synchronized cardioversion. The mode after cardioversion can be classed into one of four categories: (1) remain in synchronized mode; (2) defer to asynchronized mode; (3) 'intelligent'; and (4) configurable. The different modes and the lack of standardization present a potential hazard because they cause confusion among the user of the defibrillator. Uniformity should be obtained after broad agreement between clinicians and manufacturers of defibrillators. (C) 1997 Elsevier Science Ireland Ltd.