High-performance liquid chromatographic method for simple and rapid determination of linezolid in human plasma

被引:48
|
作者
Boak, Lauren M. [1 ]
Li, Jian [1 ]
Nation, Roger L. [1 ]
Rayner, Craig R. [1 ]
机构
[1] Monash Univ, Facil Antiinfect Drug Dev & Innovat, Dept Pharm Practice, Victorian Coll Pharm, Parkville, Vic 3052, Australia
关键词
linezolid; HPLC; plasma;
D O I
10.1002/bmc.597
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple high-performance liquid chromatographic (HPLC) method was developed and validated for rapid quantification of linezolid in human plasma. Protein precipitation using a mixture of 5% trichloroacetic acid and methanol (3:1, v/v) provided a straightforward method of sample preparation and the internal standard eperezolid was employed. A concentration range from 0.20 to 40.0 mg/L was utilized to construct calibration curves, and analysis of low- (0.40 mg/L), medium- (7.50 mg/L) and high-quality (25.0 mg/L) control samples revealed excellent reproducibility (<= 3.88%) and accuracy (<= 4.20%). The recovery of both linezolid and eperezolid was approximately 100%. No interference was observed at the retention times of linezolid and eperezolid from blank plasma or eight commonly used antibiotics. Tests confirmed the stability of linezolid in plasma during three freeze-thaw cycles, on the bench during 24 h and during long-term storage of frozen plasma for up to 12 weeks; in extracts it was stable in the HPLC autosampler over 12 h. Overall, this assay offers a highly simplistic approach to quantifying linezolid in plasma, and would be well suited to clinical pharmacokinetic, pharmacodynamic and toxicodynamic analyses. Copyright (c) 2005 John Wiley & Sons, Ltd.
引用
收藏
页码:782 / 786
页数:5
相关论文
共 50 条
  • [21] High-performance liquid chromatographic method for the determination of gabapentin in human plasma
    Zhu, Z
    Neirinck, L
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 779 (02): : 307 - 312
  • [22] RAPID HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF PROPRANOLOL IN PLASMA
    SMITH, KA
    WOOD, S
    CROUS, M
    ANALYST, 1987, 112 (04) : 407 - 409
  • [23] RAPID HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC DETERMINATION OF VANCOMYCIN IN HUMAN PLASMA
    LUKSA, J
    MARUSIC, A
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1995, 667 (02): : 277 - 281
  • [24] A simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma and application to a pharmacokinetic study
    Yu, Liyan
    Xiang, Bingren
    Zhan, Ying
    ARZNEIMITTEL-FORSCHUNG-DRUG RESEARCH, 2008, 58 (04): : 199 - 202
  • [25] Simple high-performance liquid chromatographic method for the determination of acyclovir in human plasma using fluorescence detection
    Peh, KK
    Yuen, KH
    JOURNAL OF CHROMATOGRAPHY B, 1997, 693 (01): : 241 - 244
  • [26] SIMPLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF MEDROXYPROGESTERONE ACETATE IN HUMAN-PLASMA
    READ, J
    MOULD, G
    STEVENSON, D
    JOURNAL OF CHROMATOGRAPHY, 1985, 341 (02): : 437 - 444
  • [27] A SENSITIVE AND SIMPLE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR THE DETERMINATION OF MITOXANTRONE IN PLASMA
    SLORDAL, L
    ANDERSEN, A
    WARREN, DJ
    THERAPEUTIC DRUG MONITORING, 1993, 15 (04) : 328 - 333
  • [28] Simple and sensitive high-performance liquid chromatographic method for the determination of mycophenolic acid in plasma
    Na-Bangchang, K
    Supasyndh, O
    Supaporn, T
    Banmairuroi, V
    Karbwang, J
    JOURNAL OF CHROMATOGRAPHY B, 2000, 738 (01): : 169 - 173
  • [29] High-performance liquid chromatographic method for the determination of mycophenolate mofetil in human plasma
    Tsina, I
    Kaloostian, M
    Lee, R
    Tarnowski, T
    Wong, B
    JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 681 (02): : 347 - 353
  • [30] Enantioselective high-performance liquid chromatographic method for the determination of baclofen in human plasma
    Hefnawy, MM
    Aboul-Enein, HY
    TALANTA, 2003, 61 (05) : 667 - 673