Assessment of a new point-of-care system for detection of prostate specific antigen

被引:4
|
作者
Rausch, Steffen [1 ]
Hennenlotter, Joerg [1 ]
Wiesenreiter, Josef [1 ]
Hohneder, Andrea [1 ]
Heinkele, Julian [1 ]
Schwentner, Christian [1 ]
Stenzl, Arnulf [1 ]
Todenhofer, Tilman [1 ,2 ]
机构
[1] Univ Tubingen, Dept Urol, D-72076 Tubingen, Germany
[2] Univ British Columbia, Vancouver Prostate Ctr, Vancouver, BC V6H 3Z6, Canada
来源
BMC UROLOGY | 2016年 / 16卷
关键词
Point-of-care; Prostate cancer; Prostate specific antigen; CANCER PATIENTS; TOTAL PSA; HYPERPLASIA; MEN; PREDICTOR; RELAPSE; VOLUME; BLOOD; RISK;
D O I
10.1186/s12894-016-0119-9
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: Measurement of the prostate specific antigen (PSA) remains an important tool in prostate cancer (PC) diagnosis. Due to limited availability of laboratory devices in an outpatient setting, compact and easy-to-handle point-of-care (POC) systems are desirable. Recently, a chip for PSA measurement on the concile (R) Omega 100 POC reader platform was introduced. To investigate the clinical applicability, we evaluated the system in a consecutive cohort of patients undergoing PSA measurement in our outpatient clinic. Methods: Between 07/2014 and 01/2015, PSA was analyzed in a total of 198 patients by the POC reader system and in parallel by an Immulite 2000 (R) and Centaur (R) standard laboratory system, respectively. By standard (Immulite (R)) measurement, 67 (34,2 %) had PSA > 4 ng/ml and 131 (65,8 %) had PSA = 4 ng/ml. Results were correlated by linear regression analyses for all patients and within PSA subgroups. For patients with available prostate histology after PSA measurement (n = 68), receiver-operating characteristic curves were created and area under the curve (AUC), sensitivity and specificity for the prediction of PC at best cut-off value were calculated. Results: The coefficients of determination (r(2)) for the POC device compared to laboratory testing were 0.72 (Immulite (R)) and 0.63 (Centaur (R)), respectively (both p < 0.0001). In the PSA range of = 4 ng/ml, the observed correlations were 0.75 and 0.70, respectively. For the POC test system, AUC for detection of PC was calculated with 0.745 while the standard laboratory tests showed 0.778 (Immulite (R)) and 0.771 (Centaur (R)). At best cut-off of 3.64 ng/ml, PSA analysis by the POC system showed a sensitivity of 85.7 % and a specificity of 66.7 %. Conclusions: The POC system obtained good concordance to elaborate laboratory measurement. In a screening scenario, the system provides quick and reliable PSA measurement, especially in the PSA range up to 4 ng/ml.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] Towards Point-of-Care Insulin Detection
    Soffe, Rebecca
    Nock, Volker
    Chase, James Geoffrey
    ACS SENSORS, 2019, 4 (01) : 3 - 19
  • [32] Point-of-care testing of group A streptococcal antigen: performance evaluated by external quality assessment
    Nissinen, A.
    Stranden, P.
    Myllys, R.
    Takkinen, J.
    Bjorkman, Y.
    Leinikki, P.
    Siitonen, A.
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2009, 28 (01) : 17 - 20
  • [33] Performance of a Point-of-Care Test for the Rapid Detection of SARS-CoV-2 Antigen
    Stroemer, Annabelle
    Rose, Ruben
    Schaefer, Miriam
    Schoen, Frieda
    Vollersen, Anna
    Lorentz, Thomas
    Fickenscher, Helmut
    Krumbholz, Andi
    MICROORGANISMS, 2021, 9 (01) : 1 - 11
  • [34] Gonococcal antigen detection by Immunochromatographic assay: A reliable and point-of-care diagnostic test for gonorrhoea
    Alam, Shah
    Salam, Abdus
    PAKISTAN JOURNAL OF MEDICAL SCIENCES, 2012, 28 (01) : 130 - 134
  • [35] A comprehensive comparison of laboratory and point-of-care assays for the detection of Aspergillus antigen from serum
    Schub, T.
    Suerbaum, S.
    Wagener, J.
    Dichtl, K.
    MYCOSES, 2021, 64 : 6 - 6
  • [36] A new point-of-care test for the rapid antimicrobial susceptibility assessment of uropathogens
    Arienzo, Alyexandra
    Murgia, Lorenza
    Cellitti, Valentina
    Ferrante, Valeria
    Stalio, Ottavia
    Losito, Francesca
    Gallo, Valentina
    Tomassetti, Federica
    Marino, Rossella
    Cristofano, Flavia
    Orru, Michela
    Visca, Paolo
    Di Somma, Salvatore
    Silvestri, Lorena
    Ziparo, Vincenzo
    Antonini, Giovanni
    PLOS ONE, 2023, 18 (07):
  • [37] A Portable Microfluidic System for Point-of-Care Detection of Multiple Protein Biomarkers
    Li, Nan
    Shen, Minjie
    Xu, Youchun
    MICROMACHINES, 2021, 12 (04)
  • [38] Microfluidic system for rapid detection of contagious bovine pleuropneumonia at the point-of-care
    Nepper, Julia F.
    Liljander, Anne
    Irizarry, Pedro J. Resto
    Zan, Jindong
    Jores, Joeerg
    Weibel, Douglas B.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2014, 247
  • [39] Detection System for Point-Of-Care Multiplexed Bead-Based Immunoassays
    de Haan, K.
    Dou, J.
    Aitchison, J. S.
    30TH ANNUAL CONFERENCE OF THE IEEE PHOTONICS SOCIETY (IPC), 2017, : 673 - 674
  • [40] Neurodegenerative diseases early detection and monitoring system for point-of-care applications
    Thakur, Rajiv
    Saini, Anil Kumar
    Taliyan, Rajeev
    Chaturvedi, Nidhi
    MICROCHEMICAL JOURNAL, 2025, 208