Failure of the uncoated titanium ProxiLock femoral hip prosthesis

被引:16
|
作者
Luites, J. W. H.
Spruit, M.
van Hellemondt, G. G.
Horstmann, W. G.
Valstar, E. R.
机构
[1] St Maartensklin Res Dev & Educ, OrthoRes Unit, NL-9500 GM Nijmegen, Netherlands
[2] St Maartensklin Res Dev & Educ, Dept Orthopaed, NL-9500 GM Nijmegen, Netherlands
[3] Leiden Univ, Med Ctr, Dept Orthopaed, Leiden, Netherlands
[4] Delft Univ Technol, Fac Mech Maritime & Mat Engn, Dept Biomech, Delft, Netherlands
关键词
D O I
10.1097/01.blo.0000224011.12175.83
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
New prostheses should be evaluated for stability and clinical performance. In a prospective randomized clinical trial, we implanted 22 titanium (Ti) and 20 hydroxyapatite-coated (HA) ProxiLock femoral hip prostheses during total hip arthroplasty in 42 patients. The patients were followed for 24 months with clinical, radiographic and radiostereometric analysis. Full weightbearing was allowed immediately postoperatively. One patient with a titanium stem was lost to followup. During the first two months, 34 of the 41 stems subsided and/or rotated towards retroversion, regardless of stem type. At the 24-month followup 35 of the 41 prostheses were either fully stabilized (16 HA and 11 Ti stems) or had clinical irrelevant migration (four HA and four Ti stems). Six Ti prostheses showed continuous migrations with maximums of 4.7 mm translation and 12.2 degrees retroversion; four of these were revised, the other two had no clinical complaints. Clinical relevance: The migration pattern we found indicates insufficient primary fixation of the ProxiLock stem in an immediate full weightbearing protocol. The HA coating improves the secondary stability of the prosthesis compared to the uncoated stem. Early migration is associated with an increased risk of possible future loosening and revision, and therefore we discontinued the use of this prosthesis.
引用
收藏
页码:79 / 86
页数:8
相关论文
共 50 条
  • [21] The engineering design of the MARGRON® femoral hip prosthesis system
    Frost, RB
    Sekel, R
    Kandel, L
    JOURNAL OF ENGINEERING DESIGN, 2003, 14 (02) : 115 - 127
  • [22] A novel technique for the explantation of a cemented femoral hip prosthesis
    Hillier, D.
    Kazi, H. A.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2017, 99 (02) : 176 - 177
  • [23] Intrapelvic dislocation of trial femoral prosthesis in hip arthroplasty
    Sivakumar, Brahman Shankar
    Suzuki, Arnold
    Lorentzos, Peter
    Suthersan, Mayuran
    Dave, Jaykar
    ANZ JOURNAL OF SURGERY, 2018, 88 (1-2) : E97 - E98
  • [24] Stress Analysis and Shape Design of a Femoral Hip Prosthesis
    YANG Zheng-ming1
    ChineseJournalofBiomedicalEngineering, 2011, 20 (02) : 67 - 74
  • [25] ABOUT DAMAGE OF THE FEMORAL HEAD OF TOTAL HIP PROSTHESIS
    Capitanu, L.
    Badita, L. L.
    Bursuc, D. C.
    JOURNAL OF THE BALKAN TRIBOLOGICAL ASSOCIATION, 2012, 18 (04): : 642 - 653
  • [26] Total hip arthroplasty with femoral osteotomy and modular prosthesis for proximal femoral deformity
    Xiaowen Deng
    Jun Liu
    Tao Qu
    Xusheng Li
    Ping Zhen
    Qiuming Gao
    Yun Xue
    Peng Liu
    Guoding Cao
    Xiaole He
    Journal of Orthopaedic Surgery and Research, 14
  • [27] Total hip arthroplasty with femoral osteotomy and modular prosthesis for proximal femoral deformity
    Deng, Xiaowen
    Liu, Jun
    Qu, Tao
    Li, Xusheng
    Zhen, Ping
    Gao, Qiuming
    Xue, Yun
    Liu, Peng
    Cao, Guoding
    He, Xiaole
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (01)
  • [28] Premature failure of a femoral hip replacement
    Oravcova, Monika
    Palcek, Peter
    Chalupova, Maria
    MATERIALS TODAY-PROCEEDINGS, 2016, 3 (04) : 1114 - 1117
  • [29] EXPERIMENTAL RESULTS OF TITANIUM SHAFTS FOR HIP-PROSTHESIS
    STOCK, D
    GOTTSTEIN, J
    GRISS, P
    WINTER, M
    HEIMKE, G
    ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 1983, 121 (04): : 385 - 385
  • [30] Failure Analysis of Cementless Hip Joint Prosthesis
    Dhandapani, N.
    Babu, A. Gnanavel
    Sivasankar, M.
    MATERIALS, INDUSTRIAL, AND MANUFACTURING ENGINEERING RESEARCH ADVANCES 1.1, 2014, 845 : 403 - 407