Determination of procaine hydrochloride by ion-pairing flow injection analysis with piezoelectric detection

被引:9
|
作者
Mo, ZH [1 ]
Luo, J [1 ]
Long, XH [1 ]
Xia, ZL [1 ]
机构
[1] CHONGQING UNIV,COLL CHEM & CHEM ENGN,CHONGQING 630044,PEOPLES R CHINA
来源
关键词
D O I
10.1007/s002160050469
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A fast and simple method for the determination of procaine hydrochloride by ion-pairing flow injection analysis has been developed. It is based on the formation of an ion-pair with sodium dodecyl phenylsulphonate and piezoelectric detection. The calibration curve was linear between 0.02 and 2.00 mg/mL, with a detection limit of 0.01 mg/mL, a relative standard deviation of 0.3% (10 replicates) and a sampling frequency of 120/h. The proposed method has been satisfactorily applied to the determination of procaine hydrochloride in pharmaceutical preparations.
引用
收藏
页码:556 / 558
页数:3
相关论文
共 50 条
  • [31] Piezoelectric detection of ion pairs between sulphonate and catecholamines for flow injection analysis of pharmaceutical preparations
    Mo, ZH
    Long, XH
    Zhang, MJ
    TALANTA, 1999, 48 (03) : 643 - 648
  • [32] Analysis of dye components using MECC and ion-pairing chromatography
    Jeong, Hyuk
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2006, 19 (01): : 31 - 38
  • [33] ANALYSIS OF FOOD-ADDITIVES BY ION-PAIRING ELECTROKINETIC CHROMATOGRAPHY
    NG, CL
    LEE, HK
    LI, SFY
    JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1992, 30 (05) : 167 - 170
  • [34] IS CETHEXONIUM A VERSATILE ION-PAIRING REAGENT IN HPLC ANALYSIS OF GLUCURONIDES
    LEROY, P
    LIU, HF
    NICOLAS, A
    ANNALES DE BIOLOGIE CLINIQUE, 1988, 46 (07) : 549 - 549
  • [35] Flow injection spectrophotometric determination of promazine hydrochloride and thioridazine hydrochloride
    Karpinska, J
    Misiuk, W
    Puzanowska-Tarasiewicz, H
    INDIAN JOURNAL OF CHEMISTRY SECTION A-INORGANIC BIO-INORGANIC PHYSICAL THEORETICAL & ANALYTICAL CHEMISTRY, 1998, 37 (12): : 1135 - 1139
  • [36] Control of impurities in diphenhydramine hydrochloride by an ion-pairing, reverse-phase liquid chromatographic method
    Henderson, L
    Miller, JHM
    Skellern, GG
    JOURNAL OF PHARMACY AND PHARMACOLOGY, 2001, 53 (03) : 323 - 331
  • [37] DETERMINATION OF NEOMYCIN IN MILK BY REVERSED PHASE ION-PAIRING LIQUID-CHROMATOGRAPHY
    SHAIKH, B
    JACKSON, J
    JOURNAL OF LIQUID CHROMATOGRAPHY, 1989, 12 (08): : 1497 - 1515
  • [38] Piezoelectric flow injection analysis biosensor for the detection of Salmonella typhimurium
    Babacan, S
    Pivarnik, P
    Letcher, S
    Rand, A
    JOURNAL OF FOOD SCIENCE, 2002, 67 (01) : 314 - 320
  • [39] DETERMINATION OF AMMONIUM ION IN RAINWATER AND FOGWATER BY FLOW-INJECTION ANALYSIS WITH CHEMILUMINESCENCE DETECTION
    HU, XC
    TAKENAKA, N
    TAKASUNA, S
    KITANO, M
    BANDOW, H
    MAEDA, Y
    HATTORI, M
    ANALYTICAL CHEMISTRY, 1993, 65 (23) : 3489 - 3492
  • [40] Flow injection potentiometric determination of clobutinol hydrochloride
    Issa, YM
    Zayed, SIM
    TALANTA, 2006, 69 (02) : 481 - 487