On the Monte Carlo simulation of small-field micro-diamond detectors for megavoltage photon dosimetry

被引:56
|
作者
Andreo, Pedro [1 ]
Palmans, Hugo [2 ,3 ]
Marteinsdottir, Maria [1 ]
Benmakhlouf, Hamza [1 ]
Carlsson-Tedgren, Asa [1 ]
机构
[1] Karolinska Univ Hosp, Dept Med Phys, SE-17176 Stockholm, Sweden
[2] Natl Phys Lab, Acoust & Ionising Radiat Div, Teddington TW11 0LW, Middx, England
[3] EBG MedAustron GmbH, Dept Med Phys, A-2700 Wiener Neustadt, Austria
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 01期
关键词
small field dosimetry; Monte Carlo; output correction factors; detector geometry accuracy; CHAMBERS;
D O I
10.1088/0031-9155/61/1/L1
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Monte Carlo (MC) calculated detector-specific output correction factors for small photon beam dosimetry are commonly used in clinical practice. The technique, with a geometry description based on manufacturer blueprints, offers certain advantages over experimentally determined values but is not free of weaknesses. Independent MC calculations of output correction factors for a PTW-60019 micro-diamond detector were made using the EGSnrc and PENELOPE systems. Compared with published experimental data the MC results showed substantial disagreement for the smallest field size simulated (5 mm x 5 mm). To explain the difference between the two datasets, a detector was imaged with x rays searching for possible anomalies in the detector construction or details not included in the blueprints. A discrepancy between the dimension stated in the blueprints for the active detector area and that estimated from the electrical contact seen in the x-ray image was observed. Calculations were repeated using the estimate of a smaller volume, leading to results in excellent agreement with the experimental data. MC users should become aware of the potential differences between the design blueprints of a detector and its manufacturer production, as they may differ substantially. The constraint is applicable to the simulation of any detector type. Comparison with experimental data should be used to reveal geometrical inconsistencies and details not included in technical drawings, in addition to the well-known QA procedure of detector x-ray imaging.
引用
收藏
页码:L1 / L10
页数:10
相关论文
共 50 条
  • [21] Characterization of a microSilicon diode detector for small-field photon beam dosimetry
    Akino, Yuichi
    Fujiwara, Masateru
    Okamura, Keita
    Shiomi, Hiroya
    Mizuno, Hirokazu
    Isohashi, Fumiaki
    Suzuki, Osamu
    Seo, Yuji
    Tamari, Keisuke
    Ogawa, Kazuhiko
    JOURNAL OF RADIATION RESEARCH, 2020, 61 (03) : 410 - 418
  • [22] ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY
    BEDDAR, AS
    MASON, DJ
    OBRIEN, PF
    MEDICAL PHYSICS, 1994, 21 (07) : 1075 - 1079
  • [23] Estimation of the cable effect in megavoltage photon beam by measurement and Monte Carlo simulation
    Yamauchi, Ryohei
    Igari, Mitsunobu
    Kasai, Yuya
    Hariu, Masatsugu
    Suda, Yuhi
    Kawachi, Toru
    Katayose, Tetsurou
    Mizuno, Norifumi
    Miyasaka, Ryohei
    Saitoh, Hidetoshi
    MEDICAL PHYSICS, 2020, 47 (10) : 5324 - 5332
  • [24] Effect of angular dependence for small-field dosimetry using seven different detectors
    Kawata, Kohei
    Ono, Tomohiro
    Hirashima, Hideaki
    Tsuruta, Yusuke
    Fujimoto, Takahiro
    Nakamura, Mitsuhiro
    Nakata, Manabu
    MEDICAL PHYSICS, 2023, 50 (03) : 1274 - 1289
  • [25] Small-field radiotherapy photon beam output evaluation: Detectors reviewed
    Lam, S. E.
    Bradley, D. A.
    Khandaker, M. U.
    RADIATION PHYSICS AND CHEMISTRY, 2021, 178
  • [26] Precise Monte Carlo Simulation of Single-Photon Detectors
    Stipcevic, Mario
    Gauthier, Daniel J.
    ADVANCED PHOTON COUNTING TECHNIQUES VII, 2013, 8727
  • [27] The influence of detector size relative to field size in small-field photon-beam dosimetry using synthetic diamond crystals as sensors
    Ade, N.
    Nam, T. L.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 113 : 6 - 13
  • [28] Evaluation of effective field size characteristics for small megavoltage photon beam dosimetry
    Keivan, H.
    Maskani, R.
    Shahbazi-Gahrouei, D.
    Shanei, A.
    Pandesh, S.
    Sereshke, E. Tarighati
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2022, 20 (01): : 163 - 168
  • [29] The practical application of scintillation dosimetry in small-field photon-beam radiotherapy
    Burke, Elisa
    Poppinga, Daniela
    Schoenfeld, Andreas A.
    Harder, Dietrich
    Poppe, Bjoern
    Looe, Hui Khee
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2017, 27 (04): : 324 - 333
  • [30] Monte Carlo modelling of diode detectors for small field MV photon dosimetry: detector model simplification and the sensitivity of correction factors to source parameterization
    Cranmer-Sargison, G.
    Weston, S.
    Evans, J. A.
    Sidhu, N. P.
    Thwaites, D. I.
    PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (16): : 5141 - 5153