摘要
A new automated permuting capacitor device (APCD) for calibration of inductive voltage dividers (IVDs) with four-terminal-pair definition was designed and tested. The permuting capacitors composed of 11-SMD capacitors were housed in a bulk metal in a symmetrical arrangement. Thus, the APCD is compact and the temperature of the APCD could be controlled within several mKs. Using coaxial microwave switches, the APCD was permuted quickly and automatically using electrical signals during the calibration process. This prevented the APCD from being thermally perturbed by manual operation. Owing to the improved stability of the device, the effects of stray capacitance and voltage coefficient of the APCD were evaluated accurately. The calibration of an IVD using the permuting capacitor method with this APCD in the frequency range 1-2 kHz with accuracy comparable with that of the straddling method is presented in this paper.
引用
收藏
页码:2271 / 2278
页数:8
相关论文
共 50 条
  • [1] Automated permuting capacitor device for calibration of high-precision impedance infrastructure
    Dudzik, Kordian
    Musiol, Krzysztof
    Kampik, Marian
    PRZEGLAD ELEKTROTECHNICZNY, 2022, 98 (11): : 127 - 130
  • [2] Calibration of IVDs at frequencies up to 1MHz by permuting capacitors
    Awan, SA
    Kibble, BP
    Robinson, IA
    IEE PROCEEDINGS-SCIENCE MEASUREMENT AND TECHNOLOGY, 2000, 147 (04) : 193 - 195
  • [3] Digital Bridge Voltage Ratio Calibration using an Automated Permuting Capacitive Device
    Shin, Seong Su
    Kim, Wan-Seop
    Kim, Dan Bee
    2024 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS, CPEM 2024, 2024,
  • [4] Automated calibration for assembly device installation based on Plug & Produce concept
    Kikuchi, H
    Maeda, Y
    Sugi, M
    Arai, T
    INITIATIVES OF PRECISION ENGINEERING AT THE BEGINNING OF A MILLENNIUM, 2001, : 729 - 733
  • [5] CAPACITOR MICROPHONE RECIPROCITY CALIBRATION
    CORNEY, AC
    METROLOGIA, 1975, 11 (01) : 25 - 32
  • [6] THE WIDEBAND CALIBRATION OF CAPACITOR MICROPHONES
    GIBBINGS, DLH
    GIBSON, AV
    METROLOGIA, 1984, 20 (03) : 95 - 99
  • [7] Automated Pressure Calibration of Blood Pressure Measuring Device Calibrator To Realize Its Traceability
    Turk, Ahmet
    Hamarat, Abdullah
    2019 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2019,
  • [8] CAPACITOR CALIBRATION IN A RATIO TRANSFORMER BRIDGE
    KOCH, M
    ISA TRANSACTIONS, 1970, 9 (03) : 196 - &
  • [9] A gang capacitor testing device
    Graham, VM
    PROCEEDINGS OF THE INSTITUTE OF RADIO ENGINEERS, 1928, 16 (10): : 1401 - 1403
  • [10] Comparison of Calibration vs Non-calibration Techniques in the Automated Capture Of Eye Movement Data: Initial Validation of the ROADIE Device
    Southerland, Andrew
    Hassan, Mohamed
    Aldridge, Chad
    Zhuang, Yan
    McMurry, Timothy
    Rohde, Gustavo
    NEUROLOGY, 2021, 96 (15)