Optimization of dialysate flow in on-line hemodiafiltration

被引:8
|
作者
Maduell, Francisco [1 ]
Ojeda, Raquel [1 ]
Arias-Guillen, Marta [1 ]
Fontsere, Nestor [1 ]
Vera, Manel [1 ]
Masso, Elisabeth [1 ]
Gomez, Miguel [1 ]
Rodas, Lida [1 ]
Bazan, Giannina [1 ]
Jimenez-Hernandez, Mario [1 ]
Pineiro, Gaston [1 ]
Rico, Nayra [2 ]
机构
[1] Hosp Clin Barcelona, Dept Nefrol & Trasplante Renal, E-08036 Barcelona, Spain
[2] Hosp Clin Barcelona, Dept Bioquim, E-08036 Barcelona, Spain
来源
NEFROLOGIA | 2015年 / 35卷 / 05期
关键词
Dialysate flow; On-line hemodiafiltration; Efficacy; Convective volume; TRANSFER-AREA COEFFICIENTS; ALL-CAUSE MORTALITY; MASS-TRANSFER; CONVECTIVE VOLUME; CLINICAL-EVIDENCE; HEMODIALYSIS; UREA; EFFICIENCY; DIALYZER; RATES;
D O I
10.1016/j.nefro.2015.06.019
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Currently, on-line hemodiafiltration (HDF-OL) is the most effective technique. Several randomized studies and meta-analyses have shown a reduced mortality and a direct association with convective volume has been reported. At present, it has not been established if an increased dialysate flow (Qd) results in improved results in terms of convective and depurative efficiency. We aim at assessing the effects of Qd variations on convective volume and its depurative capacity in patients on HDF-OL. Material and methods: A total of 59 patients (45 men and 14 women) from a HDF-OL programme in which a monitor 5008 Cordiax with self-replacement was used, were enrolled. Patients were assessed in 5 sessions with post-dilutional HDF-OL, using helixone-based dialyzers, with only Qd being changed (300, 400, 500, 600 and 700 ml/min). Serum levels of urea (60 Da), creatinine (113 Da),beta(2)-microglobulin (11,800 Da), myoglobin (17,200 Da) and alpha(1)-microglobulin (33,000 Da) were measured at the beginning and at the end of each session, in order to estimate the percent reduction of such solutes. Results: An increased dialysate volume per session was observed, from 117.9 +/- 6.4 L with Qd 300 ml/min to 232.4 +/- 12 L with Qd 700 ml/min. No changes were found in replacement volume or convective volume. Regarding diffusion, Qd increase was associated to a significantly increased dialysis dose, with an increased Kt from 68 +/- 6.9 L with Qd 300 ml/min to 75.5 +/- 7.3 L with Qd 700 ml/min (p < 0,001), and a gradually increased percent reduction in urea associated to increased Qd with significantly lower levels being found with Qd 300 ml/min. No changes were found in other measured substances. Conclusion: Qd variations in HDF-OL do not change convective volume. A higher Qd was associated to a slightly increased urea clearance with no change being observed for medium and large molecules. Qd optimisation to the minimal level assuring an adequate dialysis dose and allowing water and dialysate use to be rationalised should be recommended. (C) 2015 The Authors. Published by Elsevier Espana, S.L.U. on behalf of Sociedad Espanola de Nefrologia.
引用
收藏
页码:473 / 478
页数:6
相关论文
共 50 条
  • [1] Dialysate/infusate composition and infusion mode in on-line hemodiafiltration
    Pedrini, LA
    De Cristofaro, V
    Pagliari, B
    Ruggiero, P
    HEMODIALYSIS TECHNOLOGY, 2002, 137 : 344 - 349
  • [2] DIALYSATE CONSUMPTION AND DIALYSIS DOSE IN ON-LINE HEMODIAFILTRATION WITH AUTOMATED COUPLING DIALYSATE/BLOOD FLOW FUNCTION (AUTOFLOW FUNCTION)
    Park, Yongki
    Jeon, Jimin
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 206 - 207
  • [3] Centralized on-line hemodiafiltration system utilizing purified dialysate as substitution fluid
    Sato, T
    Koga, N
    ARTIFICIAL ORGANS, 1998, 22 (04) : 285 - 290
  • [4] Centralized on-line Hemodiafiltration system utilizing purified dialysate as substitution fluid
    Kidney Center, Koga Hospital, Kurume, Japan
    不详
    Artif. Organs, 4 (285-290):
  • [5] Evaluation of pre- and postdilutional on-line hemodiafiltration adequacy by partial dialysate quantification and on-line urea monitor
    Lim, PS
    Lee, HP
    Kho, B
    Yu, YL
    Chang, SC
    Lin, YY
    Yang, CC
    Lin, YT
    Wang, TH
    Kuo, SY
    Lin, LC
    BLOOD PURIFICATION, 1999, 17 (04) : 199 - 205
  • [6] On-line hemodiafiltration
    Ward, RA
    NIEREN-UND HOCHDRUCKKRANKHEITEN, 1999, 28 (01) : 18 - 24
  • [7] Effects of on-line hemodiafiltration regimens and dialysate composition on serum concentrations of magnesium and calcium ions
    Tanaka, Yusaku
    Shima, Hisato
    Hatonari, Ryosuke
    Okada, Daisuke
    Michiwaki, Hiroyuki
    Wariishi, Seiichiro
    Tao, Tomohiro
    Minakuchi, Jun
    RENAL REPLACEMENT THERAPY, 2021, 7 (01)
  • [8] MID-TERM EVALUATION OF CITRATE DIALYSATE-REINFUSATE IN ON-LINE POSTDILUTIONAL HEMODIAFILTRATION
    Calabrese, Giovanni
    Mancuso, Domenico
    Mazzotta, Antonio
    Vagelli, Giuseppe
    Balenzano, Chiara
    Steckiph, Denis
    Bertucci, Andrea
    Della Volpe, Mario
    Gonella, Marco
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2014, 29 : 217 - 217
  • [9] Effects of on-line hemodiafiltration regimens and dialysate composition on serum concentrations of magnesium and calcium ions
    Yusaku Tanaka
    Hisato Shima
    Ryosuke Hatonari
    Daisuke Okada
    Hiroyuki Michiwaki
    Seiichiro Wariishi
    Tomohiro Tao
    Jun Minakuchi
    Renal Replacement Therapy, 7
  • [10] On-line hemodiafiltration at home
    Vega, Almudena
    Abad, Soraya
    Macias, Nicolas
    Aragoncillo, Ines
    HEMODIALYSIS INTERNATIONAL, 2018, 22 (02) : E33 - E35