COMPARISON OF GLUCOSE POINT-OF-CARE VALUES WITH LABORATORY VALUES IN CRITICALLY ILL PATIENTS

被引:25
|
作者
Shearer, Anna [1 ]
Boehmer, Marilyn [1 ]
Closs, Melanie [1 ]
Dela Rosa, Rosalina [1 ]
Hamilton, Jean [1 ]
Horton, Karen [1 ]
McGrath, Rose [1 ]
Schulman, Christine
机构
[1] Providence St Vincents Med Ctr, Intens Care Unit, Portland, OR 97225 USA
关键词
INTENSIVE INSULIN THERAPY; ARTERIAL-BLOOD; METERS; HEMATOCRIT; CAPILLARY; ACCURACY; REFLECTANCE; AGREEMENT; MONITORS; PH;
D O I
10.4037/ajcc2009448
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background Blood from central venous or arterial catheters as well as from fingersticks is often used for point-of-care glucose testing. Objectives To compare glucose values obtained with a point-of-care device for catheter and fingerstick blood samples with values obtained by clinical laboratory analysis of a catheter blood sample. Methods A method-comparison design was used. Point-of-care values for fingerstick and catheter blood samples were compared with laboratory values for blood from catheters in a convenience sample of 63 critically ill patients. Device bias and precision were calculated and graphed according to the Bland-Altman method, and a t test was used to determine differences in glucose values for the 2 methods. Results Laboratory glucose values for blood from a catheter sample differed significantly from point-of-care values for blood from a fingerstick (t(1,61) = 5.01; P < .001) and from a catheter (t(1,61) = 3.91; P < .001). Bias and precision for the point-of-care device was large (fingerstick, 8.7 and 13.7; catheter, 7.0 and 14.0); 20% of the values differed from the laboratory glucose values by more than 20 mg/dL. Point-of-care glucose values for fingerstick and catheter samples did not differ (P = .41). Conclusions Glucose values obtained with a point-of-care device differ significantly from those obtained by laboratory analysis. The magnitude of these differences calls into question the widespread practice of using point-of-care glucose testing to guide insulin titration for tight glucose control. Errors in dosing could easily be made because of the large bias and precision associated with a point-of-care device. (American Journal of Critical Care. 2009; 18: 224-230)
引用
收藏
页码:224 / 230
页数:7
相关论文
共 50 条
  • [31] Evaluation of agreement between point-of-care and laboratory automated immunoassay in procalcitonin measurement among critically ill patients
    Ismail, Muhamad Aizuddin
    Mazlan, Mohd Zulfakar
    Hassan, Shamsul Kamalrujan
    Azman, Wan Norlina Wan
    Koon, Tan Say
    Yaacob, Najib Majdi
    Omar, Mahamarowi
    Mohamad, Nik Abdullah Nik
    ANAESTHESIA PAIN & INTENSIVE CARE, 2023, 27 (04) : 470 - 477
  • [32] Comparison of point-of-care hemostatic assays, routine coagulation tests, and outcome scores in critically ill patients
    Larsson, A.
    Tynngard, N.
    Kander, T.
    Bonnevier, J.
    Schott, U.
    JOURNAL OF CRITICAL CARE, 2015, 30 (05) : 1032 - 1038
  • [33] ECONOMIC EVALUATION OF A STRICT GLUCOSE CONTROL-GUIDELINE IN CRITICALLY ILL PATIENTS, FOCUSING ON COSTS OF POINT-OF-CARE TESTING OF GLUCOSE
    van Hooijdonk, R. T. M.
    Steuten, L. M. G.
    Monteban, H.
    Mulder, M. R.
    Houckgeest, F. van Braam
    Sluijs, J. P.
    Abu-Hanna, A.
    Spronk, P. E.
    Schultz, M. J.
    INTENSIVE CARE MEDICINE, 2013, 39 : S361 - S361
  • [34] Point-of-care ultrasound in critically ill patients in developing countries of Latin America
    Abdo-Cuza, Anselmo
    Blanco, Pablo
    MEDWAVE, 2019, 19 (09): : e7709
  • [35] Point-of-care glucose measurement systems should be used with great caution in critically ill intensive care unit patients Reply
    Meynaar, Iwan A.
    van Spreuwel, Margot
    Tahgkau, Peter L.
    Dawson, Lilian
    Visser, Steven Sleeswijk
    Rijks, Lode
    Vlieland, Thea Vliet
    CRITICAL CARE MEDICINE, 2010, 38 (01) : 339 - 340
  • [36] Concordance between point-of-care blood gas analysis and laboratory autoanalyzer in measurement of hemoglobin and electrolytes in critically ill patients
    Prakash, Shivesh
    Bihari, Shailesh
    Lim, Zhan Y.
    Verghese, Santosh
    Kulkarni, Hemant
    Bersten, Andrew D.
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (06)
  • [37] The effects of precision, haematocrit, pH and oxygen tension on point-of-care glucose measurement in critically ill patients: a prospective study
    Watkinson, Peter James
    Barber, Vicki Suzanne
    Amira, Esther
    James, Timothy
    Taylor, Richard
    Young, John Duncan
    ANNALS OF CLINICAL BIOCHEMISTRY, 2012, 49 : 144 - 151
  • [38] Are point-of-care measurements of glycated haemoglobin accurate in the critically ill?
    Weinel, Luke M.
    Summers, Matthew J.
    Finnis, Mark E.
    Poole, Alexis
    Kar, Palash
    Chapman, Marianne J.
    Deane, Adam M.
    Abdelhamid, Yasmine Ali
    AUSTRALIAN CRITICAL CARE, 2019, 32 (06) : 465 - 470
  • [39] SPURIOUS LABORATORY VALUES IN A CRITICALLY ILL PATIENT: THE MYSTERY OF LEUKOCYTE LARCENY
    Clements, Kevin
    Koubeissy, Assile
    Donenfeld, Thai
    Seneviratne, Chanaka
    Kupfer, Yizhak
    CRITICAL CARE MEDICINE, 2025, 53 (01)
  • [40] Multiorgan Point-of-Care Ultrasound Assessment in Critically Ill Adults
    Hobbs, Hailey
    Millington, Scott
    Wiskar, Katie
    JOURNAL OF INTENSIVE CARE MEDICINE, 2024, 39 (03) : 187 - 195