Comparison of Dual- and Single-Source Dual-Energy CT for Diagnosis of Acute Pulmonary Artery Embolism

被引:9
|
作者
Petritsch, Bernhard [1 ]
Pannenbecker, Pauline [1 ]
Weng, Andreas Max [1 ]
Veldhoen, Simon [1 ]
Grunz, Jan-Peter [1 ]
Bley, Thorsten Alexander [1 ]
Kosmala, Aleksander [1 ]
机构
[1] Univ Hosp Wurzburg, Inst Diagnost & Intervent Radiol, Wurzburg, Germany
关键词
CT-angiography; dual-energy; pulmonary embolism; image quality; radiation dose;
D O I
10.1055/a-1245-0035
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose Comparison of dual-source dual-energy CT (DS-DECT) and split-filter dual-energy CT (SF-DECT) regarding image quality and radiation dose in patients with suspected pulmonary embolism. Materials and Methods We retrospectively analyzed pulmonary dual-energy CT angiography (CTPA) scans performed on two different CT scanners in 135 patients with suspected pulmonary embolism (PE). Scan parameters for DS-DECT were 90/Sn150 kV (n = 68 patients), and Au/Sn120 kV for SF-DECT (n = 67 patients). The iodine delivery rate was 1400 mg/s in the DS-DECT group vs. 1750 mg/s in the SF-DECT group. Color-coded iodine distribution maps were generated for both protocols. Objective (CT attenuation of pulmonary trunk [HU], signal-to-noise ratio [SNR], contrast-to-noise ratio [CNR]) and subjective image quality parameters (two readers [R], five-point Likert scale), as well as radiation dose parameters (effective radiation dose, size-specific dose estimations [SSDE]) were compared. Results All CTPA scans in both groups were of diagnostic image quality. Subjective CTPA image quality was rated as good or excellent in 80.9 %/82.4 % (R1 / R2) of DS-DECT scans, and in 77.6 %/76.1 % of SF-DECT scans. For both readers, the image quality of split-filter iodine distribution maps was significantly lower (p < 0.05) with good or excellent ratings in only 43.3 %/46.3 % (R1 / R2) vs. 83.8 %/88.2 % for maps from DS-DECT. The HU values of the pulmonary trunk did not differ between the two techniques (p = n. s.), while both the SNR and CNR were significantly higher in the split-filter group (p < 0.001; p = 0.003). Both effective radiation dose (2.70 +/- 1.32 mSv vs. 2.89 +/- 0.94 mSv) and SSDE (4.71 +/- 1.63 mGy vs. 5.84 +/- 1.11 mGy) were significantly higher in the split-filter group (p < 0.05). Conclusion The split-filter allows for dual-energy imaging of suspected pulmonary embolism but is associated with lower iodine distribution map quality and higher radiation dose. Citation Format Petritsch B, Pannenbecker P, Weng AM et al. Comparison of Dual- and Single-Source Dual-Energy CT for Diagnosis of Acute Pulmonary Artery Embolism. Fortschr Rontgenstr 2020; DOI: 10.1055/a-1245-0035
引用
收藏
页码:427 / 436
页数:10
相关论文
共 50 条
  • [31] Dual-energy computed tomography in pulmonary embolism
    Lu, G-M
    Wu, S-Y
    Yeh, B. M.
    Zhang, L-J
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (992): : 707 - 718
  • [32] Dual-energy CT in the obese: a preliminary retrospective review to evaluate quality and feasibility of the single-source dual-detector implementation
    Atwi, Noah E.
    Smith, David L.
    Flores, Carson D.
    Dharaiya, Ekta
    Danrad, Raman
    Kambadakone, Avinash
    Toshav, Aran M.
    ABDOMINAL RADIOLOGY, 2019, 44 (02) : 783 - 789
  • [33] Dual-layer dual-energy CT-derived pulmonary perfusion for the differentiation of acute pulmonary embolism and chronic thromboembolic pulmonary hypertension
    Gertz, Roman Johannes
    Gerhardt, Felix
    Pienn, Michael
    Lennartz, Simon
    Kroeger, Jan Robert
    Caldeira, Liliana
    Pennig, Lenhard
    Schoemig, Thomas Henning
    Hokamp, Nils Grosse
    Maintz, David
    Rosenkranz, Stephan
    Bunck, Alexander Christian
    EUROPEAN RADIOLOGY, 2024, 34 (05) : 2944 - 2956
  • [34] The Value of Nonenhanced Single-Source Dual-Energy CT for Differentiating Metastases From Adenoma in Adrenal Glands
    Ju, Ye
    Liu, Ailian
    Dong, Yue
    Liu, Yijun
    Wang, Heqing
    Sun, Meiyu
    Pu, Renwang
    Chen, Anliang
    ACADEMIC RADIOLOGY, 2015, 22 (07) : 834 - 839
  • [35] Split-filter dual-energy CT pulmonary angiography for the diagnosis of acute pulmonary embolism: a study on image quality and radiation dose
    Petritsch, Bernhard
    Pannenbecker, Pauline
    Weng, Andreas M.
    Grunz, Jan-Peter
    Veldhoen, Simon
    Bley, Thorsten A.
    Kosmala, Aleksander
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2021, 11 (05) : 1817 - 1827
  • [36] Osteitis: a retrospective feasibility study comparing single-source dual-energy CT to MRI in selected patients with suspected acute gout
    Diekhoff, Torsten
    Scheel, Michael
    Hermann, Sandra
    Mews, Jurgen
    Hamm, Bernd
    Hermann, Kay-Geert A.
    SKELETAL RADIOLOGY, 2017, 46 (02) : 185 - 190
  • [37] Osteitis: a retrospective feasibility study comparing single-source dual-energy CT to MRI in selected patients with suspected acute gout
    Torsten Diekhoff
    Michael Scheel
    Sandra Hermann
    Jürgen Mews
    Bernd Hamm
    Kay-Geert A. Hermann
    Skeletal Radiology, 2017, 46 : 185 - 190
  • [38] Material differentiation in forensic radiology with single-source dual-energy computed tomography
    Thomas D. Ruder
    Yannick Thali
    Stephan A. Bolliger
    Sandra Somaini-Mathier
    Michael J. Thali
    Gary M. Hatch
    Sebastian T. Schindera
    Forensic Science, Medicine, and Pathology, 2013, 9 : 163 - 169
  • [39] Pulmonary Embolism Detection with Dual-Energy CT: Experimental Study of Dual-Source CT in Rabbits (vol 252, pg 61, 2009)
    Zhang, L. J.
    Zhao, Y. E.
    Wu, S. Y.
    Yeh, B. M.
    Zhou, C. S.
    Hu, X. B.
    Hu, Q. J.
    Lu, G. M.
    RADIOLOGY, 2009, 253 (03) : 900 - 900
  • [40] Dose Comparison for Pulmonary Embolism CT Studies: Single Energy Vs. Dual Energy
    Mahmood, U.