Radiation dose and image quality of computed tomography of the supra-aortic arteries: A comparison between single-source and dual-source CT Scanners

被引:11
|
作者
Saba, Luca [1 ]
di Martino, Michele [3 ]
Siotto, Paolo [2 ]
Anzidei, Michele [3 ]
Argiolas, Giovanni Maria [2 ]
Porcu, Michele [2 ]
Suri, Jasjit S. [4 ]
Wintermark, Max [5 ]
机构
[1] AOU, Dept Radiol, Cagliari Polo Monserrato Ss 554, I-09045 Cagliari, Italy
[2] Azienda Osped Brotzu, Dept Radiol, I-09100 Cagliari, Italy
[3] Univ Roma La Sapienza, Dept Radiol, Rome, Italy
[4] AtheroPoint TM, Monitoring & Diagnost Div, Roseville, CA USA
[5] Stanford Sch Med, Dept Neuroradiol, 291 Campus Dr, Stanford, CA 94305 USA
关键词
CTA; Radiation dose; CAROTID PLAQUE; ENERGY CT; ANGIOGRAPHY; STENOSIS; RISK; VOLUME; SIZE;
D O I
10.1016/j.neurad.2017.09.007
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Purpose. - The purpose of this work was to compare the image quality and radiation dose delivered to patients during computed tomography (CT) angiography (CTA) of the supra-aortic arteries using two single-source (SS) and two dual-energy (DE) CT scanners. Material and methods. - In this retrospective study, 120 patients who underwent CTA of supra-aortic arteries were studied using four different types of CT scanners: a sixteen and forty-detector-row SS and two DE CT scanners. Seventy milliters of contrast medium were injected at a flow rate of 4 mL/s using a power injector. For each patient the dose-length product (DLP), the volume computed tomography dose index (CDTlvol), the length of the scan and the effective dose (ED) were calculated. Qualitative and quantitative [image noise, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR)] image quality assessment was performed. Results. - A statistically significant lower value of the DE compared to the SS technology (P<0.0001) for the CDTI, DLP and ED was found, whereas we did not find any statistically significant difference between the four scanners for the measurements of the image noise, SNR and CNR. Conclusion. - DS CT scanners allow performing CTAs with a reduced dose compared to SS CT scanner with comparable image quality. (C) 2017 Published by Elsevier Masson SAS.
引用
收藏
页码:136 / 141
页数:6
相关论文
共 50 条
  • [1] High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose
    Korn, A.
    Fenchel, M.
    Bender, B.
    Danz, S.
    Thomas, C.
    Ketelsen, D.
    Claussen, C. D.
    Moonis, G.
    Krauss, B.
    Heuschmid, M.
    Ernemann, U.
    Brodoefel, H.
    NEURORADIOLOGY, 2013, 55 (04) : 423 - 430
  • [2] High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose
    A. Korn
    M. Fenchel
    B. Bender
    S. Danz
    C. Thomas
    D. Ketelsen
    C. D. Claussen
    G. Moonis
    B. Krauss
    M. Heuschmid
    U. Ernemann
    H. Brodoefel
    Neuroradiology, 2013, 55 : 423 - 430
  • [3] Image Quality Performance of Virtual Single-Source CT Using Dual-Source Computed Tomography
    Alikhani, Babak
    Werner, Martin
    Jamali, Leila
    Wacker, Frank
    Werncke, Thomas
    ACADEMIC RADIOLOGY, 2019, 26 (08) : 1095 - 1101
  • [4] Dual-Source Versus Single-Source Cardiac CT Angiography: Comparison of Diagnostic Image Quality
    Donnino, Robert
    Jacobs, Jill E.
    Doshi, Jay V.
    Hecht, Elizabeth M.
    Kim, Danny C.
    Babb, James S.
    Srichai, Monvadi B.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 192 (04) : 1051 - 1056
  • [5] Impact on Image Quality and Radiation Dose of Third-Generation Dual-Source Computed Tomography of the Coronary Arteries
    Apfaltrer, Georg
    Szolar, Dieter H.
    Wurzinger, Eric
    Takx, Richard A. P.
    Nance, John W.
    Dutschke, Anja
    Tschauner, Sebastian
    Loewe, Christian
    Ringl, Helmut
    Sorantin, Erich
    Apfaltrer, Paul
    AMERICAN JOURNAL OF CARDIOLOGY, 2017, 119 (08): : 1156 - 1161
  • [6] Radiation Dose and Image Quality of Prospective Triggering With Dual-Source Cardiac Computed Tomography
    Blankstein, Ron
    Shah, Amar
    Pale, Rodrigo
    Abbara, Suhny
    Bezerra, Hiram
    Bolen, Michael
    Mamuya, Wilfred S.
    Hoffmann, Udo
    Brady, Thomas J.
    Cury, Ricardo C.
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 103 (08): : 1168 - 1173
  • [7] Dual-source cardiac computed tomography: image quality and dose considerations
    Achenbach, Stephan
    Anders, Katharina
    Kalender, Willi A.
    EUROPEAN RADIOLOGY, 2008, 18 (06) : 1188 - 1198
  • [8] Dual-source cardiac computed tomography: image quality and dose considerations
    Stephan Achenbach
    Katharina Anders
    Willi A. Kalender
    European Radiology, 2008, 18 : 1188 - 1198
  • [9] Estimated radiation dose and image quality comparison of the scan protocols in dual-source computed tomography coronary angiography
    Karaarslan, Ercan
    Ulus, Sila
    Bavbek, Cengiz
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2013, 57 (04) : 407 - 414
  • [10] Coronary CT angiography with single-source and dual-source CT: Comparison of image quality and radiation dose between prospective ECG-triggered and retrospective ECG-gated protocols
    Sabarudin, Akmal
    Sun, Zhonghua
    Yusof, Ahmad Khairuddin Md
    INTERNATIONAL JOURNAL OF CARDIOLOGY, 2013, 168 (02) : 746 - 753