Hepatitis B virus (HBV) genotype determination by the COBAS® AmpliPrep/COBAS® TaqMan® HBV Test, v2.0 in serum and plasma matrices

被引:17
|
作者
Goedel, Sebastian [1 ]
Rullkoetter, Matthias [1 ]
Weisshaar, Sabrina [1 ]
Mietag, Carola [1 ]
Leying, Hermann [1 ]
Boehl, Florian [1 ]
机构
[1] Roche Diagnost Ltd, CH-6343 Rotkreuz, Switzerland
关键词
HBV genotypes; Quantification; COBAS; AmpliPrep; TaqMan; Fully automated; ANTIVIRAL THERAPY; QUANTITATIVE PCR; INFECTION; DNA; STANDARD; ASSAY;
D O I
10.1016/j.jcv.2009.05.021
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: Viral load quantification is established in the clinical management of chronic Hepatitis B virus (HBV) infection for assessing efficacies and resistance developments in anti-viral drug treatment. Objectives: The fully automated COBAS (R) AmpliPrep/COBAS (R) TaqMan (R) HBV Test, v2.0 was evaluated for the linear measuring range and the inclusivity of HBV genotype determination in EDTA plasma and serum samples. Study design: Two kit lots of the test were used to determine the linear measuring range as well as linearity and limit of detection applying different concentration levels of specimens representing HBV genotypes A to H along with a pre-core mutant and the WHO Standard. Results: The COBAS (R) AmpliPrep/COBAS (R) TaqMan (R) HBV Test, v2.0 displayed a linear measuring range of seven log(10) steps from 20 IU/mL(lower limit of quantification) to 2.3E+08 IU/mL(upper limit of quantification) yielding similar results for EDTA plasma and serum. Inclusivity was shown by reliable quantification of HBV genotypes A to H at different concentration levels. The >= 95% hit rate LOD was 15 IU/ml for genotypes C, D, F, C, the pre-core mutant and the WHO Standard and 20 IU/ml. for genotypes A, B, E and H matching the test's lower limit of quantification. 95% PROBIT analysis yielded concentrations of 8.9 IU/mL for the WHO Standard and of 6.0-16.4 IU/mL for the genotypes. Conclusions: The COBAS (R) AmpliPrep/COBAS (R) TaqMan (R) HBV Test, v2.0 provides genotype inclusivity for accurate viral load monitoring in serum and EDTA plasma samples and supports clinical routine in the management of HBV infection. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:232 / 236
页数:5
相关论文
共 50 条
  • [31] Hepatitis C Virus Genotype 2 may not be detected by the Cobas AmpliPrep/Cobas TaqMan HCV Test, version 1.0
    Watanabe, Tsunamasa
    Inoue, Takako
    Tanoue, Yasushi
    Maekawa, Hisato
    Iio, Etsuko
    Matsunami, Kayoko
    Shinkai, Noboru
    Yoshiba, Makoto
    Tanaka, Yasuhito
    HEPATOLOGY, 2013, 58 : 924A - 924A
  • [32] Hepatitis C Virus Genotype 2 May Not Be Detected by the Cobas AmpliPrep/Cobas TaqMan HCV Test, Version 1.0
    Watanabe, Tsunamasa
    Inoue, Takako
    Tanoue, Yasushi
    Maekawa, Hisato
    Hamada-Tsutsumi, Susumu
    Yoshiba, Shinsho
    Tanaka, Yasuhito
    JOURNAL OF CLINICAL MICROBIOLOGY, 2013, 51 (12) : 4275 - 4276
  • [33] COMPARISON OF THE APTIMA HIV-1 QUANT ASSAY TO THE COBAS AMPLIPREP/COBAS TAQMAN HIV-1 TEST, V2.0
    Hatzakis, A.
    Papachristou, H.
    Kantzanou, M.
    SEXUALLY TRANSMITTED INFECTIONS, 2013, 89 : A353 - A353
  • [34] Performance characteristics of the COBAS Ampliprep/COBAS TaqMan v2.0 and the Abbott RealTime hepatitis C assays - implications for response-guided therapy in genotype 1 infections
    Taylor, Ninon
    Haschke-Becher, Elisabeth
    Greil, Richard
    Strasser, Michael
    Oberkofler, Hannes
    ANTIVIRAL THERAPY, 2014, 19 (05) : 449 - 454
  • [35] Underestimation of Hepatitis C Virus Genotype 4 RNA Levels by the Cobas AmpliPrep/Cobas TaqMan Assay
    Chevaliez, Stephane
    Fix, Jenna
    Soulier, Alexandre
    Pawlotsky, Jean-Michel
    HEPATOLOGY, 2009, 50 (05) : 1681 - 1681
  • [36] Evaluation of the COBAS® AmpliPrep/COBAS® TaqMan® HCV Test v2.0 for HCV viral load monitoring using dried blood spot specimens
    Marins, Ed. G.
    Bodinaidu, Keerthi
    Lin, Matthew
    Deforest, Alison
    JOURNAL OF VIROLOGICAL METHODS, 2017, 247 : 77 - 80
  • [37] Multicenter Evaluation of Cobas HBV Real-Time PCR Assay on the Roche Cobas 4800 System in Comparison with COBAS AmpliPrep/COBAS TaqMan HBV Test: Leading Circle for Cobas 4800 Virology (LCCV) Project
    Kim, H.
    Hurt, M.
    Bae, E.
    Lee, K.
    Lee, W.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2017, 19 (06): : 979 - 979
  • [38] Comparison of Versant HBV DNA 3.0 and COBAS AmpliPrep-COBAS TaqMan assays for hepatitis B DNA quantitation: Possible clinical implications
    Garbuglia, A. R.
    Angeletti, C.
    Lauria, F. N.
    Zaccaro, P.
    Cocca, A. M.
    Pisciotta, M.
    Solmone, M.
    Capobianchi, M. R.
    JOURNAL OF VIROLOGICAL METHODS, 2007, 146 (1-2) : 274 - 280
  • [39] Dynamic range of hepatitis C virus RNA quantification with the cobas ampliprep-cobas amplicor HCV monitor v2.0 assay
    Gourlain, K
    Soulier, A
    Pellegrin, B
    Bouvier-Alias, M
    Hézode, C
    Darthuy, F
    Rémiré, J
    Pawlotsky, JM
    JOURNAL OF CLINICAL MICROBIOLOGY, 2005, 43 (04) : 1669 - 1673
  • [40] Performance of Version 2.0 of the Cobas AmpliPrep/Cobas TaqMan Real-Time PCR Assay for Hepatitis B Virus DNA Quantification
    Chevaliez, Stephane
    Bouvier-Alias, Magali
    Laperche, Syria
    Hezode, Christophe
    Pawlotsky, Jean-Michel
    JOURNAL OF CLINICAL MICROBIOLOGY, 2010, 48 (10) : 3641 - 3647