An eight-channel transmit/receive array of TE01 mode high permittivity ceramic resonators for human imaging at 7 T

被引:33
|
作者
Aussenhofer, S. A. [1 ]
Webb, A. G. [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Radiol, CJ Garter Ctr High Field MRI, NL-2333 ZA Leiden, Netherlands
关键词
Transverse electromagnetic modes; High field imaging; RF coils; Dielectric resonators; Cardiac imaging; CORONARY MR-ANGIOGRAPHY; DIELECTRIC RESONATORS; CARDIAC MRI; COIL ARRAY; DESIGN;
D O I
10.1016/j.jmr.2014.04.001
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This study describes the design, construction and operation of a new type of transmit/receive array using ceramic resonators operating in a transverse electromagnetic (TE) mode. Single element function and performance at 298.1 MHz (7 T) are analyzed and compared to a lumped element design loop coil with comparable geometry. The results show that ceramic resonators working in the TE018 mode configuration produce similar efficiency, defined as the transmit magnetic field (131) per square root of the specific absorption rate (SAR), to conventional surface coils. An array consisting of eight ceramic elements was then designed to operate in transmit/receive mode. This array was driven via power/phase splitters by two independent transmit channels and functional cardiac images were produced from a number of healthy volunteers. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:122 / 129
页数:8
相关论文
共 24 条
  • [21] Numerical Evaluation of Image Homogeneity, Signal-to-Noise Ratio, and Specific Absorption Rate for Human Brain Imaging at 1.5, 3, 7, 10.5, and 14T in an 8-Channel Transmit/Receive Array
    Cao, Zhipeng
    Park, Joshua
    Cho, Zang-Hee
    Collins, Christopher M.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2015, 41 (05) : 1432 - 1439
  • [22] Using a whole-body 31P birdcage transmit coil and 16-element receive array for human cardiac metabolic imaging at 7T
    Valkovic, Ladislav
    Dragonu, Iulius
    Almujayyaz, Salam
    Batzakis, Alex
    Young, Liam A. J.
    Purvis, Lucian A. B.
    Clarke, William T.
    Wichmann, Tobias
    Lanz, Titus
    Neubauer, Stefan
    Robson, Matthew D.
    Klomp, Dennis W. J.
    Rodgers, Christopher T.
    PLOS ONE, 2017, 12 (10):
  • [23] Indirect 1H-[13C] MRS of the human brain at 7 T using a 13C-birdcage coil and eight transmit-receive 1H-dipole antennas with a 32-channel 1H-receive array
    Jacobs, Sarah M.
    Prompers, Jeanine J.
    van Der Kemp, Wybe J. M.
    van Der Velden, Tijl A.
    Gosselink, Mark
    Meliado, Ettore Flavio
    Hoogduin, Hans M.
    Mason, Graeme F.
    de Graaf, Robin A.
    Miller, Corin O.
    Bredael, Gerard M.
    van Der Kolk, Anja G.
    Alborahal, Cezar
    Klomp, Dennis W. J.
    Wiegers, Evita C.
    NMR IN BIOMEDICINE, 2024, 37 (10)
  • [24] In vivo phosphorus magnetic resonance spectroscopic imaging of the whole human liver at 7 T using a phosphorus whole-body transmit coil and 16-channel receive array: Repeatability and effects of principal component analysis-based denoising
    van den Wildenberg, Lieke
    Gursan, Ayhan
    Seelen, Leonard W. F.
    van der Velden, Tijl A.
    Gosselink, Mark W. J. M.
    Froeling, Martijn
    van der Kemp, Wybe J. M.
    Klomp, Dennis W. J.
    Prompers, Jeanine J.
    NMR IN BIOMEDICINE, 2023, 36 (05)