Clinical advantages of using unflattened 6-MV and 10-MV photon beams generated by the medical accelerator Elekta Versa HD based on their dosimetric parameters in comparison to conventional beams

被引:12
|
作者
Baic, Blazej [1 ,2 ]
Kozlowska, Beata [2 ]
Kwiatkowski, Robert [1 ]
Dybek, Marcin [1 ]
机构
[1] Katowice Oncol Ctr, 26 Raciborska St, PL-41074 Katowice, Poland
[2] Univ Silesia, Inst Phys, 75 Pulku Piechoty 1A St, PL-41500 Chorzow, Poland
关键词
clinical application of FFF beams; dosimetric parameters; unflattened photon beams; FLATTENING FILTER; QUALITY;
D O I
10.2478/nuka-2019-0010
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Photon beams with wide energy ranges from 4 MV to 25 MV are commonly used in radiotherapy nowadays. In recent years, there has been a strong interest in a certain modification of a radiotherapeutic apparatus by the application of the so-called flattening filter-free (FFF) beam. Several advantages of FFF beams over standard flattening filter (FF) beams are noticed, and this technical solution has aroused great interest among radiotherapeutic facilities. The goal of the present study is to investigate the differences between the conventional FF and unflattened FFF 6-MV and 10-MV photon beams in some basic dosimetric parameters and their influence on the whole radiotherapeutic treatment. The data provided here include the detailed characteristics as follows: percent depth dose (PDD), beam profile, edge of a half-profile, total scatter correction factor (TSCF) and head scatter correction factor (HSCF) for FF and FFF 6-MV and 10-MV photon beams from the Elekta Versa HD accelerator in the Katowice Oncology Center in Poland.
引用
收藏
页码:77 / 86
页数:10
相关论文
共 20 条
  • [11] Investigation of neutron contamination of flattening filter and flattening filter-free 10-MV photon beams in Elekta InfinityTM accelerator
    Yani S.
    Budiansah I.
    Lestari F.P.
    Tursinah R.
    Rhani M.F.
    Haryanto F.
    Iranian Journal of Medical Physics, 2020, 17 (02) : 126 - 132
  • [12] A COMPARISON OF AIR-CAVITY INHOMOGENEITY EFFECTS FOR CO-60, 6-MV, AND 10-MV X-RAY-BEAMS
    BEACH, JL
    MENDIONDO, MS
    MENDIONDO, OA
    MEDICAL PHYSICS, 1987, 14 (01) : 140 - 144
  • [13] Evaluation of dosimetric properties of 6 & 10 MV photon beams from a liner accelerator with no flattening filter
    Pearson, D.
    Parsai, E.
    Feldmeier, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2099 - 2099
  • [14] Dosimetric properties of 8 and 10 MV photon beams from a flattening filter free clinical accelerator
    Vassiliev, O.
    Titt, U.
    Gillin, M.
    Mohan, R.
    MEDICAL PHYSICS, 2006, 33 (06) : 2292 - 2292
  • [15] Dosimetric Deviation of Misaligned Beams for a 6 MV Photon Linear Accelerator Using Monte Carlo Simulations
    Kim, S.
    MEDICAL PHYSICS, 2015, 42 (06) : 3405 - 3405
  • [16] Dosimetric comparison of 6 MV and 10 MV flattening filter-free beams using small diameter cone for trigeminal neuralgia radiosurgery
    Blanco, Oscar Abel Apaza
    Venencia, Carlos Daniel
    Moreno, Rogelio Manuel Diaz
    Munoz, Fabian
    Barolo, Franco
    Muriano, Oscar
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2024, 23
  • [17] Comparison of calculated and measured basic dosimetric parameters for total body irradiation with 6- and 18-MV photon beams
    Hoseinnezhad, Elham
    Geraily, Ghazale
    Esfahani, Mahbod
    Farzin, Mostafa
    Gholami, Somayeh
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2021, 20 (01) : 66 - 70
  • [18] Dosimetric Impact on the Flattening Filter and Addition of Gold Nanoparticles in Radiotherapy: A Monte Carlo Study on Depth Dose Using the 6 and 10 MV FFF Photon Beams
    Spina, Armando
    Chow, James C. L.
    MATERIALS, 2022, 15 (20)
  • [19] Validation of GEANT4 simulations for percentage depth dose calculations in heterogeneous media by using small photon beams from the 6-MV Cyberknife: Comparison with photon beam dosimetry with EBT2 film
    Chung Il Lee
    Sei-Chul Yoon
    Jae Won Shin
    Seung-Woo Hong
    Tae Suk Suh
    Kyung Joo Min
    Sang Deok Lee
    Su Mi Chung
    Jae-Yong Jung
    Journal of the Korean Physical Society, 2015, 66 : 1308 - 1316
  • [20] Validation of GEANT4 simulations for percentage depth dose calculations in heterogeneous media by using small photon beams from the 6-MV Cyberknife: Comparison with photon beam dosimetry with EBT2 film
    Lee, Chung Il
    Yoon, Sei-Chul
    Shin, Jae Won
    Hong, Seung-Woo
    Suh, Tae Suk
    Min, Kyung Joo
    Lee, Sang Deok
    Chung, Su Mi
    Jung, Jae-Yong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (08) : 1308 - 1316