Advanced Virtual Monochromatic Images from Dual Source Dual-Energy CT for a Reduction in the Contrast Medium Dose (389 mgI/kg) in Multiphase Liver CT: Evaluation of Image Quality and Focal Lesion Conspicuity

被引:1
|
作者
Kim, Ji Eun [1 ]
Kim, Hyun Ok [2 ,3 ]
Bae, Kyungsoo [1 ]
Cho, Jae Min [1 ]
Choi, Ho Cheol [1 ]
Choi, Dae Seob [1 ]
Na, Jaebeom [1 ]
机构
[1] Gyeongsang Natl Univ, Gyeongsang Natl Univ Hosp, Dept Radiol, Coll Med, Jinju, South Korea
[2] Gyeongsang Natl Univ, Coll Med, Dept Internal Med, 79 Gangnam Ro, Jinju 52727, South Korea
[3] Gyeongsang Natl Univ Hosp, Jinju, South Korea
关键词
Contrast Medium; Computed Tomography; Dual-Energy; Virtual Monochromatic Image; LOW TUBE VOLTAGE; ITERATIVE RECONSTRUCTION ALGORITHM; ABDOMINAL CT; DYNAMIC CT; AGENT; NOISE; SELECTION; PROTOCOL; SIZE; MDCT;
D O I
10.5812/iranjradiol.83014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Low-energy advanced virtual monochromatic images (VMIs) from dual source dual-energy CT(DSCT) could be applied to reduce contrast medium dose due to superior iodine contrast-to-noise ratios (CNRs). Objectives: To investigate feasibility of advanced VMIs from DSCT to reduce contrast medium dose in multiphase liver CT without impairing image quality and conspicuity of focal hepatic lesions (FHLs). Patients and Methods: Ninety-four patients with 110 FHL5 underwent follow-up CT twice prospectively with different protocols: protocol A, conventional 100 kVp and 555 mgI/kg; protocol B, dual-energy mode and 389 mgI/kg. The signal-to-noise ratio (SNR) and lesion-to-liver CNR were compared between the VMIs (40 -120 keV,10-keV interval) reconstructed using advanced image-based algorithm and conventional images. Two radiologists qualitatively evaluated VMIs, compared with the 100 kVp images, using a six-point scale. Results: The SNRs of pancreas, aorta, portal and hepatic vein were similar to those at 100 kVp images (P > 0.05) at 40-50 key images during late arterial phase (LAP) and at 40-60 key images during portal venous phase (PVP) except for the SNRs of aorta and veins at 40 key images which were significantly higher (P < 0.0109) during both phases and PVP, respectively. The CNR of 56 hypervascular FHLs was significantly higher at 40 - 50 keV images (P < 0.0002) and was similar at 60 keV images (P > 0.05) during LAP. The CNR of 58 hypovascular FHLs was similar at 40 -120 key images (P > 0.05) during PVP. Among the VMIs, 50 keV and 50 - 60 keV images received the highest scores (P < 0.0315 and P < 0.0041) by both reviewers during LAP and PVP, respectively. Conclusion: Advanced 50 - 60 keV images from DSCT allowed contrast dose reduction by 30% in multiphase liver CT without impairing image quality and conspicuity of FHL5, compared with the conventional 100 kVp images with 555 mgI/kg.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Virtual monochromatic imaging in dual-source dual-energy CT: Radiation dose and image quality
    Yu, Lifeng
    Christner, Jodie A.
    Leng, Shuai
    Wang, Jia
    Fletcher, Joel G.
    McCollough, Cynthia H.
    MEDICAL PHYSICS, 2011, 38 (12) : 6371 - 6379
  • [2] Radiation dose reduction in dual-energy CT using Prior Image Constrained Compressed Sensing: Image quality evaluation in virtual monochromatic imaging
    Brunner, Stephen
    Garrett, John
    Li, Yinsheng
    Tang, Tie
    Pulfer, Kari
    Hsieh, Jiang
    Rowley, Howard
    Chen, Guang-Hong
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [3] Image quality optimization and evaluation of linearly mixed images in dual-source, dual-energy CT
    Yu, Lifeng
    Primak, Andrew N.
    Liu, Xin
    McCollough, Cynthia H.
    MEDICAL PHYSICS, 2009, 36 (03) : 1019 - 1024
  • [4] Evaluation of image quality and radiation dose of abdominal dual-energy CT
    Schmidt, David
    Soderberg, Marcus
    Nilsson, Mats
    Lindvall, Hakan
    Christoffersen, Christina
    Leander, Peter
    ACTA RADIOLOGICA, 2018, 59 (07) : 845 - 852
  • [5] The potential of iodinated contrast reduction in dual-energy CT thoracic angiography; an evaluation of image quality
    Johansen, C. B.
    Martinsen, A. C. T.
    Enden, T. R.
    Svanteson, M.
    RADIOGRAPHY, 2022, 28 (01) : 2 - 7
  • [6] Advanced virtual monoenergetic images: improving the contrast of dual-energy CT pulmonary angiography
    Meier, A.
    Wurnig, M.
    Desbiolles, L.
    Leschka, S.
    Frauenfelder, T.
    Alkadhi, H.
    CLINICAL RADIOLOGY, 2015, 70 (11) : 1244 - 1251
  • [7] Virtual Monochromatic Images from Dual-Energy Multidetector CT: Variance in CT Numbers from the Same Lesion between Single-Source Projection-based and Dual-Source Image-based Implementations
    Mileto, Achille
    Barina, Andrew
    Marin, Daniele
    Stinnett, Sandra S.
    Choudhury, Kingshuk Roy
    Wilson, Joshua M.
    Nelson, Rendon C.
    RADIOLOGY, 2016, 279 (01) : 269 - 277
  • [8] Dual-energy CT enterography in evaluation of Crohn's disease: the role of virtual monochromatic images
    Guler, Ezgi
    Unal, Nalan Gulsen
    Hekimsoy, Ilhan
    Kose, Timur
    Harman, Mustafa
    Ozutemiz, Ahmet Omer
    Elmas, Nevra Zehra
    JAPANESE JOURNAL OF RADIOLOGY, 2021, 39 (04) : 341 - 348
  • [9] Dual-energy CT enterography in evaluation of Crohn’s disease: the role of virtual monochromatic images
    Ezgi Guler
    Nalan Gulsen Unal
    Ilhan Hekimsoy
    Timur Kose
    Mustafa Harman
    Ahmet Omer Ozutemiz
    Nevra Zehra Elmas
    Japanese Journal of Radiology, 2021, 39 : 341 - 348
  • [10] Dual-Energy Liver CT: Effect of Monochromatic Imaging on Lesion Detection, Conspicuity, and Contrast-to-Noise Ratio of Hypervascular Lesions on Late Arterial Phase
    Shuman, William P.
    Green, Douglas E.
    Busey, Janet M.
    Mitsumori, Lee M.
    Choi, Eunice
    Koprowicz, Kent M.
    Kanal, Kalpana M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 203 (03) : 601 - 606