An Accurate Noninvasive Blood Glucose Measurement System Using Portable Near-Infrared Spectrometer and Transfer Learning Framework

被引:40
|
作者
Yu, Yan [1 ]
Huang, Jipeng [1 ]
Zhu, Juan [2 ]
Liang, Shili [1 ]
机构
[1] Northeast Normal Univ, Sch Phys, Changchun 130024, Peoples R China
[2] Changchun Univ Technol, Sch Mechatron Engn, Changchun 130012, Peoples R China
关键词
Sugar; Blood; Computational modeling; Sensors; Spectroscopy; Predictive models; Data models; Blood glucose level; chemometric algorithms; extreme learning machine; near-infrared spectroscopy; noninvasive; portable device; TrAdaBoost;
D O I
10.1109/JSEN.2020.3025826
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Diabetes is considered one of the life-threatening diseases in the world, which needs regular monitoring of blood glucose levels. In this article, we developed a portable system that makes near-infrared spectroscopy (NIRS) technology available to non-professionals through a mobile application and a specially-made enclosure. It overcomes the shortcomings of traditional spectroscopy systems, such as large volume, high cost, complicated operation, and difficulty in online detection. To verify the feasibility of NIRS in noninvasive blood glucose concentration detection, after the pretreatment of the acquired original spectra, we compared two different feature extraction algorithms of synergy interval (Si) and genetic algorithm (GA). On this basis, two quantitative prediction models of partial least squares (PLS) and extreme learning machine (ELM) were established. The experimental results showed the model based on the combination of Si and GA and ELM (i.e., Si-GA-ELM model) as the most accurate among the selected models. At the same time, the prediction accuracy of the spectral waveband was higher than that of the full. To further overcome the difficulty of establishing a finite sample data model and reduce the influence of individual differences, the model transfer method TrAdaBoost was used to enhance the accuracy and stability of our model. The final experimental results show that the NIR spectrometer used is portable and light and can be encased as a handheld device form. Computation models combining machine learning and chemometric methods make the estimated blood glucose more feasible, which is an innovative work in noninvasive blood glucose measurement fields.
引用
收藏
页码:3506 / 3519
页数:14
相关论文
共 50 条
  • [21] Noninvasive Measurement of Glucose in Artificial Plasma with Near-Infrared and Raman Spectroscopy
    Xue, Jintao
    Chen, Han
    Xiong, Dongmei
    Huang, Guo
    Ai, Hong
    Liang, Yan
    Yan, Xinyu
    Gan, Yuan
    Chen, Cong
    Chao, Ruobing
    Ye, Liming
    APPLIED SPECTROSCOPY, 2014, 68 (04) : 428 - 433
  • [22] NONINVASIVE BLOOD-GLUCOSE SENSORS BASED ON NEAR-INFRARED SPECTROSCOPY
    HEISE, HM
    MARBACH, R
    KOSCHINSKY, T
    GRIES, FA
    ARTIFICIAL ORGANS, 1994, 18 (06) : 439 - 447
  • [23] Noninvasive Blood Glucose Analysis Based on Near-Infrared Reflectance Spectroscopy
    Lu Xiao-feng
    Zhang Ting-lin
    Xiao Feng
    Li Guang
    Wang You
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36 (07) : 2312 - 2317
  • [24] A nonlinear model for calibration of blood glucose noninvasive measurement using near infrared spectroscopy
    Li, Qing-Bo
    Li, Li-Na
    Zhang, Guang-Jun
    INFRARED PHYSICS & TECHNOLOGY, 2010, 53 (05) : 410 - 417
  • [25] Measurement and analysis of the glucose concentration of the whole blood using near-infrared spectrum
    Ding, Dong
    Zhang, Hong-Yan
    Wang, Li-Qiu
    Zhou, Wei-Dong
    Wang, Xi-Ming
    Shen, Xuan-Guo
    Jiguang Yu Hongwai/Laser and Infrared, 2003, 33 (05):
  • [26] Preliminary Clinical Validation of a Differential Correction Method for Improving Measurement Accuracy in Noninvasive Measurement of Blood Glucose Using Near-Infrared Spectroscopy
    Han, Guang
    Yu, Xuyao
    Xia, Dandan
    Liu, Rong
    Liu, Jin
    Xu, Kexin
    APPLIED SPECTROSCOPY, 2017, 71 (09) : 2177 - 2186
  • [27] COMPLETELY NONINVASIVE MEASUREMENTS OF HUMAN BLOOD-GLUCOSE INVIVO USING NEAR-INFRARED WAVES
    NODA, M
    TANIGUCHI, N
    KIMURA, M
    KUZUYA, N
    KANAZAWA, Y
    DIABETOLOGIA, 1992, 35 : A204 - A204
  • [28] Near-Infrared Spectroscopic Method for Noninvasive Blood Glucose Prediction without Using Multivariate Analyses
    Maruo, Katsuhiko
    Yamada, Yukio
    DIABETES, 2018, 67
  • [29] Development of Human Blood Glucose Noninvasive Measurement System Based on Near Infrared Spectral Technology
    Li Qing-bo
    Liu Jie-qiang
    Li Xiang
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32 (03) : 642 - 646
  • [30] Noninvasive Blood Glucose Detection Using Near Infrared Sensor
    Hotmartua, Rolamjaya
    Pangestu, Pujianto Wira
    Zakaria, Hasballah
    Irawan, Yoke Saadia
    5TH INTERNATIONAL CONFERENCE ON ELECTRICAL ENGINEERING AND INFORMATICS 2015, 2015, : 687 - 692