New approach for the QCM sensors characterization

被引:36
|
作者
Casteleiro-Roca, J. L. [1 ]
Calvo-Rolle, J. L. [1 ]
Meizoso-Lopez, M. C. [1 ]
Pinon-Pazos, A. [1 ]
Rodriguez-Gomez, B. A. [1 ]
机构
[1] Univ A Coruna, Dept Ingn Ind, La Coruna, Spain
关键词
Bio-sensor; BVD; Electronics interface; Impedance measurement; Characterization; QCM sensor; QUARTZ-CRYSTAL MICROBALANCE; IMPEDANCE; LIQUIDS; MASS; DEGRADATION; INTERFACE; FREQUENCY; MHZ;
D O I
10.1016/j.sna.2013.12.002
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The present work shows a novel approach for the QCM characterization. The method is based on the resonance principle of passive components. With the proposal, the QCM equivalent circuit parameters based on BVD (Butterworth-Van Dyke) model are obtained. The results achieved with this new technique are compared with specific commercial equipment usually employed for this purpose. The best advantages of the proposed method are its intuitivity and the possibility to obtain it with basic electronic equipment, usually present in any electronic laboratory. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 50 条
  • [41] An FPGA-based system on chip for the measurement of QCM sensors resolution
    Moure, Maria Jose
    Valdes, Maria Dolores
    Rodiz, Pablo
    Rodriguez-Pardo, Loreto
    Farina, Jose
    2006 INTERNATIONAL CONFERENCE ON FIELD PROGRAMMABLE LOGIC AND APPLICATIONS, PROCEEDINGS, 2006, : 693 - 696
  • [42] Preprocessing of matrix QCM sensors data for the classification by means of neural network
    Reznik, AM
    Galinskaya, AA
    Dekhtyarenko, OK
    Nowicki, DW
    SENSORS AND ACTUATORS B-CHEMICAL, 2005, 106 (01) : 158 - 163
  • [43] Chemisorbed PEGylated lipopolymers as sensing film supports for QCM odor sensors
    Wyszynski, Bartosz
    Somboon, Pakpum
    Nakamoto, Takamichi
    SENSORS AND ACTUATORS B-CHEMICAL, 2008, 130 (02) : 857 - 863
  • [44] Air monitoring system in elders' apartment with QCM type gas sensors
    Kikuchi, Masashi
    Ito, Tsukasa
    Shiratori, Seimei
    IEEJ Transactions on Sensors and Micromachines, 2006, 126 (12) : 631 - 636
  • [45] Use of the dynamic response of QCM sensors to detect CO in presence of interferents
    Di Nucci, C
    Fort, A
    Rocchi, S
    Tondi, L
    Vignoli, V
    Serrano, B
    Di Francesco, F
    Pioggia, G
    Sensors for Environmental Control, 2003, : 173 - 177
  • [46] Development of QCM sensors modified by AlPO4-18 films
    Vilaseca, M.
    Yague, C.
    Coronas, J.
    Santamaria, J.
    SENSORS AND ACTUATORS B-CHEMICAL, 2006, 117 (01) : 143 - 150
  • [47] QCM sensors based on PEI films for CO2 detection
    Xie, Guang-Zhong
    Kang, Ting
    Zhou, Yong
    Xie, Tao
    Tai, Hui-Ling
    Jiang, Ya-Dong
    Journal of Electronic Science and Technology, 2015, 13 (02) : 181 - 187
  • [48] Apparent negative mass in QCM sensors due to punctual rigid loading
    Castro, P.
    Resa, P.
    Elvira, L.
    INTERNATIONAL SYMPOSIUM ON ULTRASOUND IN THE CONTROL OF INDUSTRIAL PROCESSES (UCIP 2012), 2012, 42
  • [49] QCM Sensors Based on PEI Films for CO2 Detection
    Guang-Zhong Xie
    Ting Kang
    Yong Zhou
    Tao Xie
    Hui-Ling Tai
    Ya-Dong Jiang
    Journal of Electronic Science and Technology, 2015, (02) : 181 - 187
  • [50] Phthalocyanine as Sensitive Coatings for QCM Sensors-Experimental and Computational Approaches
    Erbahar, D. D.
    Harbeck, M.
    Gurol, I.
    Musluoglu, E.
    Ozturk, Z. Z.
    Berber, S.
    OLFACTION AND ELECTRONIC NOSE: PROCEEDINGS OF THE 14TH INTERNATIONAL SYMPOSIUM ON OLFACTION AND ELECTRONIC NOSE, 2011, 1362 : 58 - +