strontium-90;
inductively coupled plasma mass spectrometry;
electrothermal vaporization;
D O I:
暂无
中图分类号:
O61 [无机化学];
学科分类号:
070301 ;
081704 ;
摘要:
A two-step temperature program has been developed to determine Sr-90 in highly active samples in the presence of Zr. In a first step, Sr is volatilized at 2400 degrees C using argon as a carrier gas, while Zr is completely retained in the graphite tube. It is removed in a second step by adding 0.1% CHF3 to the argon carrier gas and increasing the temperature to 2650 degrees C. Even a 100fold excess of Zr has no detrimental effects on the determination of Sr-90. The detection limit is 2 pg/ml (10 Bq/ml) without using a modifier.
机构:
Natl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, CanadaNatl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, Canada
Grinberg, Patricia
Willie, Scott
论文数: 0引用数: 0
h-index: 0
机构:
Natl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, CanadaNatl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, Canada
Willie, Scott
Sturgeon, Ralph E.
论文数: 0引用数: 0
h-index: 0
机构:
Natl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, CanadaNatl Res Council Canada, Inst Natl Measurement Stand, Ottawa, ON K1A 0R6, Canada