High-performance liquid chromatography electrospray mass spectrometry for the analysis of modified bases in DNA: 7-(2-hydroxyethyl)guanine, the major ethylene oxide DNA adduct

被引:29
|
作者
Leclercq, L
Laurent, C
DePauw, E
机构
[1] UNIV LIEGE,INST CHIM B6,LAB SPECTROMETRIE MASSE,B-4000 LIEGE,BELGIUM
[2] CHU SART TILMAN,ORME,B-4000 LIEGE,BELGIUM
关键词
D O I
10.1021/ac9607673
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method was developed for the analysis of 7-(2-hydroxyethyl)guanine (7HEG), the major DNA adduct formed after exposure to ethylene oxide (EO). The method is based on DNA neutral thermal hydrolysis, adduct microconcentration, and final characterization and quantification by HPLC coupled to single-ion monitoring electrospray mass spectrometry (HPLC/SIR-ESMS). The method was found to be selective, sensitive, and easy to handle with no need for enzymatic digestion or previous sample derivatization. Detection limit was found to be close to 1 fmol of adduct injected (10(-10) M), thus allowing the detection of approximately three modified bases on 10(8) intact nucleotides in blood sample analysis. Quantification results are shown for 7HEG after calf thymus DNA and blood exposure to various doses of EO, in both cases obtaining clear dose-response relationships.
引用
收藏
页码:1952 / 1955
页数:4