Liquid ionization chambers for radiation therapy

被引:0
|
作者
Berghöfer, T [1 ]
Eberle, K [1 ]
Engler, J [1 ]
Hartmann, G [1 ]
Hofmann, R [1 ]
Hörandel, JR [1 ]
机构
[1] Univ Karlsruhe, Inst Expt Kernphys, D-76021 Karlsruhe, Germany
关键词
on-tine monitor; liquid ionization chamber; IMRT; ion therapy;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Liquid ionization chambers are described to be used as monitors in radiation therapy. The development focuses on two areas of application, an on-line monitor for intensity modulated radiation therapy and the measurement of the energy deposit of heavy ions in tissue. This requires a thin detector with two-dimensional read-out and an absorbing chamber with three-dimensional read-out, respectively. Both detectors are realized using similar techniques. First measurements with prototype chambers are described. The liquids isooctane, isononane (TMP), and tetramethylsilane (TMS) are used in a high purity grade in order to realize high current signals for electronic read-out in parallel at frequencies exceeding 10 Hz. Signals of more than a factor 4 with respect to isooctane, analysis grade, are obtained. With an electrode structure of 400 pads in a two-dimensional geometry an uniformity in efficiency within 1.2% has been measured.
引用
收藏
页码:2128 / 2132
页数:5
相关论文
共 50 条
  • [41] Ionization chambers
    Schrader, H.
    METROLOGIA, 2007, 44 (04) : S53 - S66
  • [42] Onset of space charge effects in liquid argon ionization chambers
    Toggerson, B.
    Newcomer, A.
    Rutherfoord, J.
    Walker, R. B.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (02): : 238 - 245
  • [43] SPECTROMETRIC PROPERTIES OF CYLINDRICAL LIQUID-XENON IONIZATION CHAMBERS
    KIRSANOV, MA
    KRUGLOV, AA
    OBODOVSKII, IM
    POKACHALOV, SG
    SHILOV, VA
    VORONOVA, TY
    RAZINOVA, IG
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1991, 34 (01) : 66 - 69
  • [44] The use of recombination chambers at radiation therapy facilities
    Zielczynski, Mieczyslaw
    Golnik, Natalia
    Gryzinski, Michal A.
    Tulik, Piotr
    RADIATION MEASUREMENTS, 2010, 45 (10) : 1472 - 1475
  • [45] PERTURBATION CORRECTION FACTOR OF IONIZATION CHAMBERS IN BETA-RADIATION FIELDS
    BOHM, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (01): : 65 - 75
  • [46] THE MEASUREMENT OF EXPOSURE DOSE FOR HIGH ENERGY RADIATION WITH CAVITY IONIZATION CHAMBERS
    BURLIN, TE
    PHYSICS IN MEDICINE AND BIOLOGY, 1959, 3 (03): : 197 - 206
  • [47] REM RATEMETER FOR MIXED RADIATION USING 2 SPHERICAL IONIZATION CHAMBERS
    KEMMOCHI, M
    HEALTH PHYSICS, 1978, 35 (02): : 333 - 340
  • [48] PROPERTIES OF LIQUID IONIZATION CHAMBERS AT LDR BRACHYTHERAPY DOSE-RATES
    JOHANSSON, B
    WICKMAN, G
    HOLMSTROM, T
    PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04): : 575 - 587
  • [49] LIQUID-ARGON IONIZATION CHAMBERS AS TOTAL-ABSORPTION DETECTORS
    WILLIS, WJ
    RADEKA, V
    NUCLEAR INSTRUMENTS & METHODS, 1974, 120 (02): : 221 - 236
  • [50] LIQUID IONIZATION CHAMBERS WITH ELECTRON EXTRACTION AND MULTIPLICATION IN THE GASEOUS PHASE.
    Anderson, D.F.
    Charpak, G.
    Holroyd, R.A.
    Lamb, D.C.
    Nuclear instruments and methods in physics research, 1987, A261 (03): : 445 - 448