Applied Taguchi method for fatigue testing of customized hip implant

被引:2
|
作者
Dharme, Mangesh R. [1 ]
Kuthe, Abhaykumar M. [1 ]
Deshmukh, Tushar R. [2 ]
机构
[1] Visvesvaraya Natl Inst Technol, Dept Mech Engn, Nagpur, Maharashtra, India
[2] Prof Ram Meghe Inst Technol & Res, Dept Mech Engn, Amravati, India
来源
关键词
Experimental model; Fatigue; Finite element analysis; Gait cycle; Orthogonal array; Taguchi methods; FINITE-ELEMENT-ANALYSIS; 3-DIMENSIONAL SHAPE OPTIMIZATION; FEMORAL STEMS; INDIAN POPULATION; WALKING ACTIVITY; CONTACT FORCES; COBALT CHROME; ARTHROPLASTY; PROSTHESIS; JOINT;
D O I
10.5301/ijao.5000545
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose: Human activities generate stresses, which vary with time and may result in fatigue failure of the customized hip implant. This study aims to investigate fatigue testing of customized hip implants using the minimum number of experiments by the Taguchi method, for 147 patients. This study was also useful to determine the influential geometrical parameters on the fatigue safety factor of customized hip implants. Methods: Horizontal offset (HO), vertical offset (VO) and neck shaft angle (NSA) of the hip joint of 147 patients were measured on computed tomography (CT) scanned images. Stress and strain of hip implants were calculated by finite element analysis and validated by in vitro experimental tests. Fatigue safety factors were calculated by Goodman, Soderberg and Gerber's fatigue theories and maximum stresses. Results: Analysis of variance results show that the highest impact on fatigue safety factors was equal to 54.38% for HO, 16.33% for VO, and was equal to 29.16% for NSA with reference to Goodman, Soderberg and Gerber's fatigue theories. The hip implant shape of experiment no. 8 has the highest safety factor value compared to all other hip implants. Conclusions: The results show that HO has the maximum influence on fatigue safety factors. The determination of influential geometric parameters may be useful to redesign customized hip implants in order to achieve the highest fatigue safety factor. The Taguchi method is suitable for fatigue testing of custom hip implant with a minimum number of experiments.
引用
收藏
页码:611 / 618
页数:8
相关论文
共 50 条
  • [31] The fatigue design of a bone preserving hip implant with functionally graded cellular material
    Khanoki, S.A., 2013, American Society of Mechanical Engineers (ASME) (07):
  • [32] Fatigue Testing Method for Thin Wires
    O. I. Aksenov
    A. A. Fuks
    N. A. Volkov
    A. S. Aronin
    Instruments and Experimental Techniques, 2023, 66 : 488 - 491
  • [33] A METHOD OF TESTING STEEL FOR CORROSION FATIGUE
    PETROV, LN
    STEPURENKO, YV
    INDUSTRIAL LABORATORY, 1981, 47 (10): : 1082 - 1083
  • [34] AN APPRAISAL OF THE PROT METHOD OF FATIGUE TESTING
    CORTEN, HT
    DIMOFF, T
    DOLAN, TJ
    PROCEEDINGS-AMERICAN SOCIETY FOR TESTING AND MATERIALS, 1954, 54 : 875 - 902
  • [35] METHOD FOR FATIGUE TESTING OF CERAMIC MATERIALS
    ACQUAVIVA, SJ
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1971, 42 (12): : 1858 - +
  • [36] ON METHOD OF FATIGUE TESTING OF BOLT STEELS
    STEPANOV, GA
    INDUSTRIAL LABORATORY, 1964, 30 (05): : 746 - &
  • [37] A method of testing ceramics for dynamic fatigue
    Barinov, SM
    Ivanov, NV
    Orlov, SV
    Shevchenko, VY
    INDUSTRIAL LABORATORY, 1997, 63 (07): : 429 - 431
  • [38] Fatigue Testing Method for Thin Wires
    Aksenov, O. I.
    Fuks, A. A.
    Volkov, N. A.
    Aronin, A. S.
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2023, 66 (03) : 488 - 491
  • [39] Mandibular Reconstruction Using a Customized Three-Dimensional Titanium Implant Applied on the Lingual Surface of the Mandible
    Jo, You-Young
    Kim, Seong-Gon
    Kim, Min-Keun
    Shin, Sung-Ho
    Ahn, Janghoon
    Seok, Hyun
    JOURNAL OF CRANIOFACIAL SURGERY, 2018, 29 (02) : 415 - 419
  • [40] Mechanical Testing of Indirect Composite Materials Directly Applied on Implant Abutments
    Andriani, Wilson, Jr.
    Suzuki, Marcelo
    Bonfante, Estevam A.
    Carvalho, Ricardo M.
    Silva, Nelson R.
    Coelho, Paulo G.
    JOURNAL OF ADHESIVE DENTISTRY, 2010, 12 (04): : 311 - 317