A Volumetric CMUT-Based Ultrasound Imaging System Simulator With Integrated Reception and μ-Beamforming Electronics Models

被引:1
|
作者
Matrone, Giulia [1 ]
Savoia, Alessandro S. [2 ]
Terenzi, Marco [4 ]
Caliano, Giosue [3 ]
Quaglia, Fabio [4 ]
Magenes, Giovanni [1 ]
机构
[1] Univ Pavia, Dipartimento Ingn Ind & Informaz, I-27100 Pavia, Italy
[2] Univ Roma Tre, Dipartimento Ingn, Rome, Italy
[3] Univ Roma Tre, Dipartimento Ingn, ACULAB, Rome, Italy
[4] STMicroelectronics, Milan, Italy
关键词
MATRIX TRANSDUCER; PULSES; PROBE;
D O I
10.1109/TUFFC.2014.2971
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
In modern ultrasound imaging devices, two-dimensional probes and electronic scanning allow volumetric imaging of anatomical structures. When dealing with the design of such complex 3-D ultrasound (US) systems, as the number of transducers and channels dramatically increases, new challenges concerning the integration of electronics and the implementation of smart micro-beamforming strategies arise. Hence, the possibility to predict the behavior of the whole system is mandatory. In this paper, we propose and describe an advanced simulation tool for ultrasound system modeling and simulation, which conjugates the US propagation and scattering, signal transduction, electronic signal conditioning, and beamforming in a single environment. In particular, we present the architecture and model of an existing 16-channel integrated receiver, which includes an amplification and micro-beamforming stage, and validate it by comparison with circuit simulations. The developed model is then used in conjunction with the transducer and US field models to perform a system simulation, aimed at estimating the performance of an example 3-D US imaging system that uses a capacitive micromachined ultrasonic transducer (CMUT) 2-D phased-array coupled to the modeled reception front-end. Results of point spread function (PSF) calculations, as well as synthetic imaging of a virtual phantom, show that this tool is actually able to model the complete US image reconstruction process, and that it could be used to quickly provide valuable system-level feedback for an optimized tuning of electronic design parameters.
引用
收藏
页码:792 / 804
页数:13
相关论文
共 33 条
  • [1] A CMUT-based Real-time Volumetric Ultrasound Imaging System with Row-Column Addressing
    Chen, Albert I. H.
    Wong, Lawrence L.
    Logan, Andrew S.
    Yeow, John T. W.
    2011 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2011, : 1755 - 1758
  • [2] An Integrated Beamforming Driver for CMUT based Ultrasound Catheter Ablation System
    Yetik, Hasan
    Ergun, A. Sanli
    Bozkurt, Ayhan
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 1532 - 1535
  • [3] Transceiver Design for CMUT-Based Super-Resolution Ultrasound Imaging
    Behnamfar, Parisa
    Molavi, Reza
    Mirabbasi, Shahriar
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2016, 10 (02) : 383 - 393
  • [4] Receiver Design for CMUT-Based Super-Resolution Ultrasound Imaging
    Behnamfar, Parisa
    Molavi, Reza
    Mirabbasi, Shahriar
    2014 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2014, : 878 - 881
  • [5] MEMS-Based Transducers (CMUT) and Integrated Electronics for Medical Ultrasound Imaging
    Savoia, Alessandro S.
    Caliano, Giosue
    SENSORS, 2018, 431 : 421 - 429
  • [6] Integrated HIFU Drive System on a Chip for CMUT-Based Catheter Ablation System
    Farhanieh, Omid
    Sahafi, Ali
    Roy, Rupak Bardhan
    Ergun, Arif Sanli
    Bozkurt, Ayhan
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2017, 11 (03) : 534 - 546
  • [7] Fully integrated CMUT-based forward-looking intracardiac imaging for electrophysiology
    Nikoozadeh, Amin
    Wygant, Ira O.
    Linl, Der-Song
    Oralkan, Oemer
    Ergun, Arif Sanli
    Thomenius, Kai
    Dentinger, Aaron
    Wildes, Douglas
    Akopyan, Gina
    Shivkumar, Kalyanam
    Mahajan, Aman
    Stephens, Douglas N.
    Sahn, David
    Khuri-Yakub, Pierre T.
    2007 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1-6, 2007, : 900 - +
  • [8] CMUT-based Volumetric Ultrasonic Imaging Array Design for Forward Looking ICE and IVUS Applications
    Tekes, Coskun
    Zahorian, Jaime
    Xu, Toby
    Rashid, Muhammad W.
    Satir, Sarp
    Gurun, Gokce
    Karaman, Mustafa
    Hasler, Jennifer
    Degertekin, F. Levent
    MEDICAL IMAGING 2013: ULTRASONIC IMAGING, TOMOGRAPHY, AND THERAPY, 2013, 8675
  • [9] Towards a Reduced-Wire Interface for CMUT-Based Intravascular Ultrasound Imaging Systems
    Lim, Jaemyung
    Tekes, Coskun
    Degertekin, F. Levent
    Ghovanloo, Maysam
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2017, 11 (02) : 400 - 410
  • [10] A 32x32 Integrated CMUT Array for Volumetric Ultrasound Imaging
    Bhuyan, Anshuman
    Chang, Chienliu
    Choe, Jung Woo
    Lee, Byung Chul
    Nikoozadeh, Amin
    Oralkan, Oemer
    Khuri-Yakub, Butrus T.
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 541 - 544