Surface dose measurements with GafChromic EBT film for 6 and 18 MV photon beams

被引:44
|
作者
Bilge, Hatice [1 ]
Cakir, Aydin [1 ]
Okutan, Murat [1 ]
Acar, Hilal [1 ]
机构
[1] Istanbul Univ, Inst Oncol, Dept Med Phys, TR-34390 Istanbul, Turkey
关键词
Surface dose; High energy photons; GafChromic EBT films; PARALLEL IONIZATION CHAMBERS; RADIOCHROMIC FILM; WATER;
D O I
10.1016/j.ejmp.2008.05.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to determine the surface doses using GafChromic EBT films and compare them with plane-parallel. ionization chamber measurements for 6 and 18 MV high energy photon beams. The measurements were made in a water equivalent solid phantom in the build-up region of the 6 and 18 MV photon beams at 100 cm SSD for various field sizes. Markus type plane-parallel ion chamber with fixed-separation between collecting electrodes was used to measure the percent depth doses. GafChromic EBT film measurements were performed both on the phantom surface and maximum dose depth at the same geometry with ion chamber measurements. The surface doses found using GafChromic EBT film were 15%, 20%, 29% and 39% +/- 2% (1SD) for 6 MV photons, 6%, 11%, 23% and 32% +/- 2% (1SD) for 18 MV photons at 5, 10, 20 and 30 cm(2) field sizes, respectively. GafChromic EBT film provides precise measurements for surface dose in the high energy photons. Agreement between film and plane-parallel chamber measurements was found to be within +/- 3% for 18 MV photon beams. There was 5% overestimate on the surface doses when compared with the plane-parallel chamber measurements for all field sizes in the 6 MV photon beams. (C) 2008 Published by Elsevier Ltd on behalf of Associazione Italiana di Fisica Medica.
引用
收藏
页码:101 / 104
页数:4
相关论文
共 50 条
  • [31] Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film
    Sonier, Marcus
    Wronski, Matt
    Yeboah, Collins
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 304 - 314
  • [32] Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film
    Sonier, M.
    Wronski, M.
    Yeboah, C.
    MEDICAL PHYSICS, 2014, 41 (08) : 12 - 12
  • [33] Dose verification of proton beam therapy using the Gafchromic EBT film
    Park, S. A.
    Kwak, J. W.
    Yoon, M. G.
    Shin, D. H.
    Lee, S. B.
    Cho, K. H.
    Kang, S. K.
    Kim, K. J.
    Bae, H. S.
    Park, S. Y.
    RADIATION MEASUREMENTS, 2011, 46 (08) : 717 - 721
  • [34] Characterization and dosimetry of photon multileaf collimated electron beams using Gafchromic film measurements
    Mutsakanyi, Stalyn
    du Plessis, Freek
    RADIATION PHYSICS AND CHEMISTRY, 2021, 182
  • [35] A Study of Dose Response Characteristic for Gafchromic EBT2 Film
    Chen, W.
    Di, X.
    Wang, S.
    Wang, B.
    Bai, X.
    Yang, J.
    Guo, B.
    Huang, K.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [36] Study of Gafchromic® EBT film response over a large dose range
    Martisikova, Maria
    Jaekel, Oliver
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (10): : N281 - N290
  • [37] Verification of CyberKnife Dose Distributions with Gafchromic EBT2 Film
    Koksal, Canan
    Akbas, Ugur
    Kesen, Nazmiye Donmez
    Bilge, Hatice
    Kemikler, Gonul
    TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2018, 33 (04): : 147 - 153
  • [38] Investigation of the gafchromic film-EBT2: Features for UVR measurements
    Abukassem, I.
    Bero, M. A.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 97 : 360 - 365
  • [39] Characterization of a Three-Dimensional Radiochromic Film Stack Dosimeter for Measurements of 6 MV Photon Beams
    McCaw, T.
    Micka, J.
    DeWerd, L.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [40] Evaluation of the Energy Dependence of Gafchromic EBT3 Film for Electron Beams
    Inoue, T.
    Sugimoto, S.
    Fukata, K.
    Kurokawa, C.
    Ozawa, S.
    Sasai, K.
    MEDICAL PHYSICS, 2012, 39 (06) : 3735 - 3736