摘要
Photon-counting x-ray detectors (PCXDs) are being investigated as a replacement for conventional x-ray detectors because they promise several advantages, including better dose efficiency, higher resolution and spectral imaging. However, many of these advantages disappear when the x-ray flux incident on the detector is too high. We recently proposed a dynamic, piecewise-linear attenuator (or beam shaping filter) that can control the flux incident on the detector. This can restrict the operating range of the PCXD to keep the incident count rate below a given limit We simulated a system with the piecewise-linear attenuator and a PCXD using raw data generated from forward projected DICOM files. We investigated the classic paralyzable and nonparalyzable PCXD as well as a weighted average of the two, with the weights chosen to mimic an existing PCXD (Taguchi et al, Med Phys 2011). The dynamic attenuator has small synergistic benefits with the nonparalyzable detector and large synergistic benefits with the paralyzable detector. Real PCXDs operate somewhere between these models, and the weighted average model still shows large benefits from the dynamic attenuator. We conclude that dynamic attenuators can reduce the count rate performance necessary for adopting PCXDs.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] SIS Detectors for Terahertz Photon Counting System
    Ezawa, Hajime
    Matsuo, Hiroshi
    Ukibe, Masahiro
    Fujii, Go
    Shiki, Shigetomo
    JOURNAL OF LOW TEMPERATURE PHYSICS, 2016, 184 (1-2) : 244 - 249
  • [22] Cascaded systems analysis of photon counting detectors
    Xu, J.
    Zbijewski, W.
    Gang, G.
    Stayman, J. W.
    Taguchi, K.
    Lundqvist, M.
    Fredenberg, E.
    Carrino, J. A.
    Siewerdsen, J. H.
    MEDICAL PHYSICS, 2014, 41 (10)
  • [23] Photon Counting Detectors and Their Applications in Medical Imaging
    McCollough, Cynthia
    Schmidt, T.
    Leng, S.
    Molloi, S.
    Li, K.
    MEDICAL PHYSICS, 2017, 44 (06) : 3161 - 3162
  • [24] Estimation of photon counting statistics with imperfect detectors
    Han, Xiao-Chuan
    Zhuang, Dong-Wei
    Li, Yu-Xuan
    Song, Jun-Feng
    Zhang, Yong-Sheng
    CHINESE PHYSICS B, 2018, 27 (07)
  • [25] THEORY OF OPTIMUM PHOTON-COUNTING DETECTORS
    KAMAL, AK
    MALAVIYA, N
    INTERNATIONAL JOURNAL OF ELECTRONICS, 1974, 37 (06) : 817 - 820
  • [26] Requirements on Photon Counting Detectors for Terahertz Interferometry
    Hiroshi Matsuo
    Journal of Low Temperature Physics, 2012, 167 : 840 - 845
  • [27] Photon-counting chip for avalanche detectors
    Zappa, F
    Lotito, A
    Tisa, S
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2005, 17 (01) : 184 - 186
  • [28] Linear arrays of single-photon detectors for photon counting and timing
    Guerrieri, F.
    Tisa, S.
    Tosi, A.
    Bellisai, S.
    Markovic, B.
    Zappa, F.
    SENSORS, CAMERAS, AND SYSTEMS FOR INDUSTRIAL, SCIENTIFIC, AND CONSUMER APPLICATIONS XII, 2011, 7875
  • [29] Infrared photon counting with superconducting nanowire single-photon detectors
    Hadfield, Robert H.
    Natarajan, Chandra M.
    Tanner, Michael G.
    2013 IEEE PHOTONICS CONFERENCE (IPC), 2013, : 368 - 369
  • [30] Photon-counting distribution for arrays of single-photon detectors
    Jonsson, Mattias
    Bjork, Gunnar
    PHYSICAL REVIEW A, 2020, 101 (01)