Effect of flow rate, humidifier dome and water volume on maximising heated, humidified gas use for neonatal resuscitation

被引:7
|
作者
Farley, R. C. [1 ]
Gibbons, K. [2 ]
Jardine, L. A. [1 ,2 ,3 ]
Shearman, A. D. [1 ,2 ]
机构
[1] Mater Mothers Hosp, Dept Newborn Serv, South Brisbane, Qld, Australia
[2] Mater Med Res Inst, Brisbane, Qld, Australia
[3] Univ Queensland, Dept Paediat & Child Hlth, Brisbane, Qld 4072, Australia
关键词
Resuscitation; Neonatal; Humidification; Heated Gas; Temperature control; T-Piece resuscitator; BIRTH-WEIGHT INFANTS; CARDIOPULMONARY-RESUSCITATION; TEMPERATURE; DELIVERY; HYPOTHERMIA;
D O I
10.1016/j.resuscitation.2013.04.004
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Aim: Dry, cold gas is used for neonatal resuscitation, contributing to low admission temperatures and exacerbation of lung injury. Recently, a method of heating and humidifying neonatal resuscitation gases has become available. We aimed to determine the optimal flow rate, humidifier chamber and water volume needed to reach 36 degrees C, and near 100% humidity at the patient T-piece in the shortest possible time. Method: AT-piece resuscitator was connected via a heated patient circuit to a humidifier chamber. Trials were performed using different gas flow rates (6, 8 and 10 L/min), humidification chambers (MR290, MR225) and water volumes (30g. 108g). Temperature was recorded at the humidifier chamber (T1), distal temperature probe (T2) and the T-piece (T3) over a 20 min period at 30s intervals. A test lung was added during one trial. Results: No significant difference existed between flow rates 8L/min and 10 L/min (p = 0.091, p = 0.631). T3 reached 36 C and remained stable at 360s (8 L/min, MR225, 30 mL); near 100% RH was reached at 107 s(10 L/min, MR225, 30 mL). T3 and humidity reached and remained stable at 480 s(10 L/min, MR290, 30 mL). Target temperature and humidity was not reached with the test lung. Conclusions: It is possible to deliver heated, humidified gases in neonatal resuscitation in a clinically acceptable timeframe. We suggest the set-up to achieve optimal temperature and humidity for resuscitation purposes is 10 L/min of gas flow, a MR290 humidification chamber, and 30 mL of water. Crown Copyright (C) 2013 Published by Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:1428 / 1432
页数:5
相关论文
共 29 条
  • [21] THE EFFECT OF GAS FLOW RATE, EXPOSURE TIMES AND AGEING ON THE PHYSICOCHEMICAL PROPERTIES OF WATER ACTIVATED BY GLOW DISCHARGE PLASMA JET&DAG;
    Naeim, Farah A.
    Humud, Hammad R.
    EAST EUROPEAN JOURNAL OF PHYSICS, 2023, (02): : 370 - 377
  • [22] Effect of Gas Flow Rate, Bubble Size and Inclusion Size on Inclusion Removal Under High Throughput Conditions Using Water Model Experiment
    Furumai, Kohei
    Murai, Takeshi
    Aramaki, Norichika
    Miki, Yuji
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 2017, 103 (09): : 19 - 25
  • [23] INVERSE GAS-CHROMATOGRAPHY OF POLY(N-BUTYL METHACRYLATE) - EFFECT OF FLOW-RATE ON SPECIFIC RETENTION VOLUME AND DETECTION OF GLASS-TRANSITION TEMPERATURE
    TYAGI, OS
    DESHPANDE, DD
    JOURNAL OF APPLIED POLYMER SCIENCE, 1987, 34 (07) : 2377 - 2388
  • [24] The Effect of Water-Vapor Content and Gas Flow Rate on the Oxidation Mechanism of a 10%Cr-Ferritic Steel in Ar-H2O Mixtures
    J. Żurek
    M. Michalik
    F. Schmitz
    T. -U. Kern
    L. Singheiser
    W. J. Quadakkers
    Oxidation of Metals, 2005, 63 : 401 - 422
  • [25] The effect of water-vapor content and gas flow rate on the oxidation mechanism of a 10%Cr-Ferritic steel in Ar-H2O mixtures
    Zurek, J
    Michalik, M
    Schmitz, F
    Kern, TU
    Singheiser, L
    Quadakkers, WJ
    OXIDATION OF METALS, 2005, 63 (5-6): : 401 - 422
  • [26] ESTIMATION OF EXTRAVASCULAR LUNG WATER IN HUMANS WITH USE OF (H2O)-H-2 - EFFECT OF BLOOD-FLOW AND CENTRAL BLOOD-VOLUME
    WALLIN, CJB
    LEKSELL, LG
    JOURNAL OF APPLIED PHYSIOLOGY, 1994, 76 (05) : 1868 - 1875
  • [27] EFFECT OF CENTRAL GAS-FLOW RATE AND WATER ON AN ARGON INDUCTIVELY COUPLED PLASMA - SPATIALLY RESOLVED EMISSION, ION-ATOM INTENSITY RATIOS AND ELECTRON NUMBER DENSITIES
    OLESIK, JW
    DEN, SJ
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (07) : 731 - 752
  • [28] Effect of liquid volume fraction and shear rate on rheological properties and microstructure formation in ternary particle/oil/water dispersion systems under shear flow: two-dimensional direct numerical simulation
    Ishigami, Toru
    Karasudani, Taisei
    Onitake, Shu
    Shirzadi, Mohammadreza
    Fukasawa, Tomonori
    Fukui, Kunihiro
    Mino, Yasushi
    SOFT MATTER, 2022, 18 (22) : 4338 - 4350
  • [29] Use of Reduced Irrigation Operating Pressure in Irrigation Scheduling. II. Effect of Reduced Irrigation System Operating Pressure on Drip-tape Flow Rate, Water Application Uniformity, and Soil Wetting Pattern on a Sandy Soil
    Poh, Bee Ling
    Gazula, Aparna
    Simonne, Eric H.
    Hochmuth, Robert C.
    Alligood, Michael R.
    HORTTECHNOLOGY, 2011, 21 (01) : 22 - 29