A well-type NaI(Tl)/BC404 phoswich detector for atmospheric radioxenon monitoring

被引:5
|
作者
Luo Fei [1 ]
Xiang Yongchun [1 ]
Gong Jian [1 ]
Ze Rende [1 ]
Yan Zhaotong [1 ]
Wang Qian [1 ]
Weng Kuiping [1 ]
Hao Fanhua [1 ]
Zeng Jun [1 ]
机构
[1] China Acad Engn Phys, Inst Nucl Phys & Chem, 64 MianShan Rd, Mianyang, Sichuan, Peoples R China
来源
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT | 2021年 / 986卷
关键词
Radioxenon; Nuclear explosion monitoring; Beta-gamma coincidence detection; Phoswich detector;
D O I
10.1016/j.nima.2020.164715
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The monitoring of radioxenon in the atmosphere is one of several methods to detect nuclear weapon tests as part of the Comprehensive Nuclear Test Ban Treaty. For ultra-low radioxenon detection, a phoswich detector with two scintillators coupled to a single photomultiplier was explored to simplify the time-based coincidence detection systems such as SAUNA and ARSA. We present a well-type NaI(Tl)/BC404 phoswich detector, newly developed by the Institute of Nuclear Physics and Chemistry, which consist of a thin BC404 cell to detect beta and conversion electrons, and a NaI(Tl) crystal to detect X-rays and gamma-rays. Moreover, the pulses of the detector were sampled and processed at 1GS/s with 10-bit ADC by a high-speed digitizer. In the preliminary experiments with various sources, the detector was test to have comparable sensitivity with the existing detectors in energy resolution, detection efficiency, background, memory effect and MDCs. The gamma resolution achieved 6.9% at 662keV. The coincidence efficiency of Xe-133 at 81keV region is about 71% while gas loss in the stainless tube and the valve was accounted. In addition, the MDC of Xe-133 was calculated by 0.2mBq/m(3) under current gas process capability.
引用
收藏
页数:6
相关论文
共 27 条
  • [1] A phoswich well detector for radioxenon monitoring
    Hennig, Wolfgang
    Tan, Hui
    Fallu-Labruyere, Anthony
    Warburton, William K.
    McIntyre, Justin I.
    Gleyzer, Anshel
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 579 (01): : 431 - 436
  • [2] A well-type phoswich detector for nuclear explosion monitoring
    B. Alemayehu
    A. T. Farsoni
    L. Ranjbar
    E. M. Becker
    Journal of Radioanalytical and Nuclear Chemistry, 2014, 301 : 323 - 332
  • [3] A well-type phoswich detector for nuclear explosion monitoring
    Alemayehu, B.
    Farsoni, A. T.
    Ranjbar, L.
    Becker, E. M.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2014, 301 (02) : 323 - 332
  • [4] SURVEY OF STANDARDIZATION POSSIBILITIES WITH A NAI(TL) WELL-TYPE DETECTOR
    PAVLIK, A
    WINKLER, G
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1983, 34 (08): : 1167 - 1174
  • [5] CALIBRATION OF WELL-TYPE NAI (TL) DETECTORS
    BRINKMAN, GA
    VEENBOER, JT
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1979, 30 (03): : 171 - 176
  • [6] DESIGN AND CONSTRUCTION OF A SHIELD FOR THE 9" x 9" NaI(Tl) WELL-TYPE DETECTOR
    Hansman, Jan
    NUCLEAR TECHNOLOGY & RADIATION PROTECTION, 2014, 29 (02): : 165 - 169
  • [7] Calibration of well-type NaI(Tl) detector using a point sources measured out the detector well at different axial distances
    Gouda, M. M.
    Badawi, M. S.
    El-Khatib, A. M.
    Mohamed, M. M.
    Thabet, A. A.
    Abbas, M. I.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [8] Characterisation of a well-type NaI(Tl) detector by means of a Monte Carlo simulation for radionuclide metrology application
    Heranudin
    Smith, M. L.
    van Wyngaardt, W. M.
    Guatelli, S.
    Li, E.
    Rosenfeld, A.
    APPLIED RADIATION AND ISOTOPES, 2021, 176
  • [9] Some Aspects of Activity Measurements with NaI(Tl) Well-Type Detectors
    WINKLER, G
    PAVLIK, A
    INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1983, 34 (02): : 547 - 553
  • [10] Analytical formulae for well-type NaI (Tl) and HPGe detectors efficiency computation
    Abbas, MI
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (02) : 245 - 252