In vitro flow analysis of artificial heart valve

被引:0
|
作者
Owida, AA [1 ]
Morsi, Y [1 ]
Sakhaeimanesh, A [1 ]
机构
[1] Swinburne Univ Technol, Sch Sci & Engn, Ctr Modelling & Proc Simulat, Bio Fluid Dynam Grp, Hawthorn, Vic 3122, Australia
关键词
prosthetic valve; aortic valves; steady flow; pulsatile flow; jellyfish heart valve;
D O I
暂无
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
Before clinical use of any new valve design, it is necessary to carry out in vitro analysis of the flow characteristics through the valve. This helps in assessing the two major problems related to the blood flow namely thromboembolism and hemolysis. In vitro experiments are carried out under either steady or pulsatile flow conditions. Steady flow experiments are simple in setting up and analysis of data; however, their validity and accuracy have been questioned. Pulsatile experiments on the other hand are complex and can be time consuming. Moreover, due to the recent development of Computational Fluid Dynamics the need to carry out the in vitro experiment under pulsatile flow conditions is becoming questionable. This paper presents and discusses the fluid flow characteristics of a Jellyfish valve obtained experimentally under steady and pulsatile flow conditions. It also demonstrates the accuracy of data obtained using Computational Fluid Dynamics. The findings indicate that the main features of the flow are equally observed under both steady and pulsatile conditions. Moreover, the preliminary results of the computational flow using Fluent agreed well with the experimental findings. From the argument presented in this paper it is concluded that, at least in the initial optimization of the valve haemodynamic performance, the steady hydrodynamic evaluation of the valve could be an effective tool for analyzing the flow characteristics and provide sufficient data for the construction of a complete three-dimensional analysis of the flow.
引用
收藏
页码:893 / 898
页数:6
相关论文
共 50 条
  • [31] Operating point control system for a continuous flow artificial heart:: In vitro study
    Kosaka, R
    Yanagi, K
    Sato, T
    Ishitoya, H
    Ichikawa, S
    Motomura, T
    Kawahito, S
    Mikami, M
    Linneweber, J
    Nonaka, K
    Takano, T
    Glueck, J
    Sankai, Y
    Nosè, Y
    ASAIO JOURNAL, 2003, 49 (03) : 259 - 264
  • [32] Initial in vitro testing of a paediatric continuous-flow total artificial heart
    Fukamachi, Kiyotaka
    Karimov, Jamshid H.
    Horvath, David J.
    Sunagawa, Gengo
    Byram, Nicole A.
    Kuban, Barry D.
    Moazami, Nader
    INTERACTIVE CARDIOVASCULAR AND THORACIC SURGERY, 2018, 26 (06) : 897 - 901
  • [33] ARTIFICIAL HEART - NEW DESIGN OF A BUILT IN INLET VALVE IN-VITRO STUDY - FIRST RESULTS
    BELENGER, J
    BUCLIN, M
    EUROPEAN SURGICAL RESEARCH, 1973, 5 : 99 - 99
  • [34] In vitro Analysis of the Thrombogenic Potential of Heart Valve Prostheses in Blood Testing
    Linde, Torsten
    Mager, Ilona
    Schmitz-Rode, Thomas
    Steinseifer, Ulrich
    CARDIOLOGY, 2016, 134 (02) : 236 - 236
  • [35] A trileaflet heart valve crown in vitro
    Hoerstrup, SP
    Sodian, R
    Sperling, JS
    Martin, DP
    Schoen, FJ
    Vacanti, JP
    Mayer, JE
    CIRCULATION, 1999, 100 (18) : 527 - 528
  • [36] IN VITRO STUDY OF THE INFLUENCE OF THE AORTIC ROOT GEOMETRY ON FLOW CHARACTERISTICS OF A PROSTHETIC HEART VALVE
    Barannyk, Oleksandr
    Karri, Satya
    Oshkai, Peter
    PROCEEDINGS OF THE ASME PRESSURE VESSELS AND PIPING CONFERENCE - 2013, VOL 4: FLUID-STRUCTURE INTERACTION, 2014,
  • [37] Design and Analysis of a Magnetic Levitation Axial Flow Artificial Heart
    Sun, Chuanyu
    Zhuang, Peng
    Li, Jingkai
    Han, Shangke
    Yang, Hang
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2020, 126 : 87 - 87
  • [38] Closing behavior of the mechanical heart valve in a total artificial heart
    H. S. Lee
    T. Tsukiya
    A. Homma
    Y. Taenaka
    E. Tatsumi
    H. Takano
    Journal of Artificial Organs, 2003, 6 (1) : 0037 - 0041
  • [39] Observation of cavitation in a mechanical heart valve in a total artificial heart
    Lee, H
    Tsukiya, T
    Homma, A
    Kamimura, T
    Takewa, Y
    Nishinaka, T
    Tatsumi, E
    Taenaka, Y
    Takano, H
    Kitamura, S
    ASAIO JOURNAL, 2004, 50 (03) : 205 - 210
  • [40] A Novel Material for the Coating of Artificial Heart Valve
    Xi Tingfei
    Huang Nan
    Tian Wenhua
    Yang Ping
    Leng Yongxiang
    Cheng Xiang
    Liu Xianghuai(1 ) Center of Medical Device
    Chinese Journal of Biomedical Engineering(English Edition), 1998, (04) : 200 - 205