Development of a SiPM-based PET imaging system for small animals

被引:15
|
作者
Lu, Yanye [1 ]
Yang, Kun [2 ]
Zhou, Kedi [1 ]
Zhang, Qiushi [1 ]
Pang, Bo [1 ]
Ren, Qiushi [1 ]
机构
[1] Peking Univ, Dept Biomed & Engn, Coll Engn, Beijing 100871, Peoples R China
[2] Hebei Univ, Coll Qual & Tech Supervis, Dept Control Technol & Instrumentat, Baoding 071000, Peoples R China
基金
中国国家自然科学基金;
关键词
Geiger-mode APD; LYSO; Instrumentation; Animal imaging; SILICON PHOTOMULTIPLIER; PERFORMANCE EVALUATION; DESIGN;
D O I
10.1016/j.nima.2014.01.010
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Advances in small animal positron emission tomography (PET) imaging have been accelerated by many new technologies such as the successful incorporation of silicon photomultiplier (SiPM). In this paper, we have developed a compact, lightweight PET imaging system that is based on SiPM detectors for small animals imaging, which could be integrated into a multi-modality imaging system. This PET imaging system consists of a stationary detector gantry, a motor-controlled animal bed module, electronics modules, and power supply modules: The PET detector, which was designed as a multi-slice circular ring geometry of 27 discrete block detectors, is composed of a cerium doped lutetium-yttrium oxyorthosilicate (LYSO) scintillation crystal and SiPM arrays. The system has a 60 mm transaxial field of view (FOV) and a 26 mm axial FOV. Performance tests (e.g. spatial resolution, energy resolution, and sensitivity) and phantom and animal imaging studies were performed to evaluate the imaging performance of the PET imaging system. The performance tests and animal imaging results demonstrate the feasibility of an animal PET system based on SiPM detectors and indicate that SiPM detectors can be promising photodetectors in animal PET instrumentation development. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:30 / 38
页数:9
相关论文
共 50 条
  • [41] Evaluation of SiPM-based readout system with gain stabilization for TL dosimetry
    Hwang, Jisung
    Kim, Hyoungtaek
    Lim, Kyung Taek
    Cho, Gyuseong
    RADIATION PHYSICS AND CHEMISTRY, 2023, 212
  • [42] A simultaneous FMT-PET imaging system for small animals
    Wang, Xin
    Zhang, Bin
    Liu, Shuangquan
    Liu, Xin
    Shan, Baoci
    Bai, Jing
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS IV, 2010, 7845
  • [43] A SiPM-based gamma spectrometer for nuclear science
    Buonanno, L.
    Di Vita, D.
    Canclini, F.
    Carminati, M.
    Camera, F.
    Fiorini, C.
    NUOVO CIMENTO C-COLLOQUIA AND COMMUNICATIONS IN PHYSICS, 2021, 44 (2-3):
  • [44] Application of the VMM ASIC for SiPM-based calorimetry
    Bearden, I.
    Buchakchiev, V.
    Buhl, A.
    Dufke, L.
    Isidori, T.
    Jia, S.
    Kozhuharov, V.
    Loizides, C.
    Muller, H.
    Pfeiffer, D.
    Rauch, M.
    Rusu, A.
    Simeonov, R.
    JOURNAL OF INSTRUMENTATION, 2024, 19 (10):
  • [45] Hybrid charge division multiplexing method for SiPM-based high-resolution PET detectors
    Park, Haewook
    Ko, Guen Bae
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [46] Initial experience with a SiPM-based PET/CT scanner: influence of acquisition time on image quality
    Ida Sonni
    Lucia Baratto
    Sonya Park
    Negin Hatami
    Shyam Srinivas
    Guido Davidzon
    Sanjiv Sam Gambhir
    Andrei Iagaru
    EJNMMI Physics, 5
  • [47] Initial Experience with a SiPM-based PET/CT Scanner: Influence of Acquisition Time on Image Quality
    Sonni, Ida
    Park, Sonya
    Baratto, Lucia
    Hatami, Negin
    Davidzon, Guido
    Srinivas, Shyam
    Gambhir, Sanjiv
    Iagaru, Andrei
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [48] Design of a Monolithic Multi-Channel Front-End Readout ASIC for LYSO/SiPM-based Small-Animal Flat-Panel PET Imaging
    Gao, W.
    Gao, D.
    Wei, T.
    Zeng, H.
    Duan, Y.
    Lu, S.
    Shen, L.
    Xu, W.
    Xie, Q.
    Hu, Y.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2447 - 2450
  • [49] Development of a passive quenching SiPM-based detection system for quantification of real time PCR products based on fluorescence lifetime
    Siminfar, A. R.
    Shoaei, O.
    JOURNAL OF INSTRUMENTATION, 2019, 14 (05):
  • [50] NEMA NU 2-2012 performance studies for the SiPM-based ToF-PET component of the GE SIGNA PET/MR system
    Grant, Alexander M.
    Deller, Timothy W.
    Khalighi, Mohammad Mehdi
    Maramraju, Sri Harsha
    Delso, Gaspar
    Levin, Craig S.
    MEDICAL PHYSICS, 2016, 43 (05) : 2334 - 2343