A New Capsule Platinum Resistance Thermometer for Cryogenic Use

被引:1
|
作者
Courts, S. S. [1 ]
Krause, J. K. [1 ]
机构
[1] Lake Shore Cryotron Inc, Westerville, OH 43082 USA
关键词
thermometer; cryogenic thermometer; platinum resistance thermometer;
D O I
10.1063/1.4819534
中图分类号
O414.1 [热力学];
学科分类号
摘要
Standards grade platinum resistance thermometers (SPRTs) obtain their high stability in part due to the strain-free mounting of the sensing wire. The space required for this strain-free mounting normally results in thermometers on the order of 6 mm diameter by 40 mm length in size. While these SPRTs are acceptable in many applications, it is desirable to reduce the size as much as possible for cryogenic use where space is of major concern. For over 40 years Minco Products, Inc. provided a smaller alternative with their model S1059, a high-stability cryogenic capsule platinum resistance thermometer (PRT) packaged in a copper canister sized only 3.2 mm diameter by 9.7 mm length. The packaging was compatible for use over the 13 K to 533 K temperature range. Unfortunately, this product was discontinued in 2009. In its absence, Lake Shore Cryotronics, Inc., has worked with Advanced Sensing Products to develop a similarly sized replacement sensor for cryogenic use. The replacement capsule PRT is manufactured using the model S1059 design, but with modifications to reduce the chance of lead breakage at the epoxy-lead interface. Test devices have been fabricated and tested for temperature response and stability upon repeated calibration from 13 K to 330 K. The new sensor design features and performance data are presented in this work.
引用
收藏
页码:168 / 173
页数:6
相关论文
共 50 条
  • [31] Thin film resistance thermometer with simple package for cryogenic application
    Li, Yun
    Wang, Fangfang
    You, Minmin
    Lin, Zude
    Liu, Jingquan
    CRYOGENICS, 2020, 105
  • [32] A new high-temperature platinum resistance thermometer made of a sapphire single crystal
    Gam, KS
    Kim, YG
    Park, JC
    Kang, KH
    Jang, CG
    TEMPMEKO 2001: 8TH INTERNATIONAL SYMPOSIUM ON TEMPERATURE AND THERMAL MEASUREMENT IN INDUSTRY AND SCIENCE, VOL 1 & 2, PROCEEDINGS, 2002, : 121 - 125
  • [33] A DIRECT-READING BRIDGE FOR A PLATINUM RESISTANCE THERMOMETER
    VANDERWAL, CW
    STRUIK, LCE
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1969, 2 (02): : 143 - +
  • [34] MINIATURE LOW-INERTIA PLATINUM RESISTANCE THERMOMETER
    ZAITSEV, AM
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1964, (01) : 250 - &
  • [35] Comparison of Methods for Nonlinearity Correction of Platinum Resistance Thermometer
    Hocenski, Zeljko
    Cvitas, Ljubivoj
    Lasinger, Zeljko
    2008 PROCEEDINGS OF SICE ANNUAL CONFERENCE, VOLS 1-7, 2008, : 3035 - +
  • [36] DIGITAL LINEARIZATION METHOD FOR PLATINUM RESISTANCE THERMOMETER RESPONSE
    MAZURIK, BI
    KOSHELEV, VA
    MEASUREMENT TECHNIQUES, 1979, 22 (11) : 1371 - 1373
  • [37] High temperature standard platinum resistance thermometer calibration
    Teixeira, RN
    TEMPMEKO 2001: 8TH INTERNATIONAL SYMPOSIUM ON TEMPERATURE AND THERMAL MEASUREMENT IN INDUSTRY AND SCIENCE, VOL 1 & 2, PROCEEDINGS, 2002, : 669 - 674
  • [38] ACCURACY OF TEMPERATURE MEASUREMENT BY A STANDARD PLATINUM RESISTANCE THERMOMETER
    SYCHEV, VV
    GORBUNOV.NI
    HIGH TEMPERATURE, 1965, 3 (04) : 579 - &
  • [39] LINEARIZATION OF THE RESPONSE FROM A PLATINUM RESISTANCE THERMOMETER - COMMENT
    BOLK, WT
    JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1983, 16 (02): : 177 - 177
  • [40] Evaluation of a horizontal furnace for precise comparison between a radiation thermometer and a platinum resistance thermometer
    Yamazawa, Kazuaki
    Anso, Kiyoshi
    Arai, Masaru
    PROCEEDINGS OF SICE ANNUAL CONFERENCE, VOLS 1-8, 2007, : 1392 - 1395