Safety of Internal Electronic Fetal Heart Rate Monitoring During Labor

被引:1
|
作者
Liang, Yiheng [1 ,2 ]
Li, Yun [1 ]
Huang, Chunhua [1 ]
Li, Xiaohong [1 ]
Cai, Qiaoqing [1 ]
Peng, Jiying [2 ]
Fan, Shangrong [1 ,3 ]
机构
[1] Peking Univ Shenzhen Hosp, Dept Obstet & Gynecol, Shenzhen 518036, Peoples R China
[2] Peking Univ Shenzhen Hosp, Dept Pathol, Shenzhen 518036, Peoples R China
[3] Shenzhen Key Lab Technol Early Diag Major Gynecol, Shenzhen 518036, Peoples R China
基金
中国国家自然科学基金;
关键词
Fetal monitoring; Neonatal academia; Cesarean section; Chorioamnionitis; Puerperal morbidity; Puerperal infection; SCALP ELECTRODE; ABSCESS;
D O I
10.1097/FM9.0000000000000145
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective: To investigate the safety and efficacy of internal electronic fetal heart rate (FHR) monitoring during labor. Methods: This was a retrospective case-control study, which was an analysis of monitoring FHR with a fetal scalp electrode or a Doppler probe (94 pregnant women per group), In the internal monitoring group, when the opening of the uterine orifice was >= 3 cm, the fetal scalp electrode was placed after natural or artificial rupture of the membrane. FHR was simultaneously monitored using a Doppler probe. In the external monitoring group, continuous FHR monitoring was performed using an ultrasound Doppler transducer fixed on the maternal abdomen. The toco transducer was used to record uterine contractions. Pathological examination of the placenta was performed prospectively in 49 and 48 cases in the internal electronic FHR monitoring group and control group, respectively. Maternal-infant outcomes (e.g. fever, puerperal infection, puerperal morbidity, delivery mode, Apgar score, and scalp injury) were recorded. Umbilical artery blood was extracted for blood gas analysis. Differences between the two groups were compared using the paired t-test, chi(2) test, Yates corrected chi(2) test or Fisher exact test. Results: Non-statistically significant differences between the internal and external monitoring groups were observed in the incidence of neonatal acidosis (1/94 (1.06%) vs. 3/94 (3.19%), respectively; chi(2)= 0 .255, P=0.613), cesarean section/operative vaginal delivery (8/94 (8.51%) vs. 15/94 (15.96%), respectively; chi(2) = 2.427, P=0.181), fever during labor (18/94 (19.15%) vs. 15/94 (15.96%), respectively; chi(2) = 0. 331, P= 0.565), puerperal morbidity (2/94 (2.13%) vs. 3/94 (3.19%), respectively; chi(2) = 0.000, P = 1.000), chorioamnionitis (9/49 (18.37%) vs. 7/48(14.58%), respectively; chi(2 )=0.252, P= 0.616), and neonatal asphyxia (0/94 (0.00%) vs. 1/94 (1.06%), respectively; chi(2 )= 0 .000, p=1.000). There were no puerperal infections, neonatal scalp injuries, or scalp abscesses found in either group. Using the internal monitoring value as reference, the incidence of FHR false deceleration in external FHR monitoring was 20.21% (19/94 women). Conclusion: Internal FHR monitoring during labor does not increase the incidence of adverse perinatal outcomes. External monitoring was associated with FHR false decelerations.
引用
收藏
页码:121 / 126
页数:6
相关论文
共 50 条
  • [1] Safety of Internal Electronic Fetal Heart Rate Monitoring During Labor
    Liang Yiheng
    Li Yun
    Huang Chunhua
    Li Xiaohong
    Cai Qiaoqing
    Peng Jiying
    Fan Shangrong
    母胎医学杂志(英文), 2022, 04 (02) : 121 - 126
  • [2] CONTINUOUS ELECTRONIC MONITORING OF FETAL HEART-RATE DURING LABOR
    GABERT, HA
    STENCHEVER, MA
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1973, 115 (07) : 919 - 923
  • [3] Interpretation of the electronic fetal heart rate during labor
    Sweha, A
    Hacker, TW
    Nuovo, J
    AMERICAN FAMILY PHYSICIAN, 1999, 59 (09) : 2487 - 2500
  • [4] FETAL HEART-RATE MONITORING DURING LABOR
    SCHIFRIN, BS
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1972, 222 (02): : 196 - &
  • [5] Monitoring the Fetal Heart Rate Variability during Labor
    Moslem, B.
    Mohydeen, A.
    Bazzi, O.
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 5846 - 5850
  • [6] ROUTINE ELECTRONIC MONITORING OF FETAL HEART-RATE AND UTERINE ACTIVITY DURING LABOR
    SHENKER, L
    POST, RC
    SEILER, JS
    OBSTETRICS AND GYNECOLOGY, 1975, 46 (02): : 185 - 189
  • [7] The Efficacy of In-Phase and Quadrature Demodulation in Electronic Fetal Heart Rate Monitoring During Labor
    Liang Yiheng
    Liu Ping
    Yan Shaomei
    Li Yun
    Chen Duijin
    Fan Shangrong
    母胎医学杂志(英文), 2022, 04 (02) : 113 - 120
  • [8] The Efficacy of In-Phase and Quadrature Demodulation in Electronic Fetal Heart Rate Monitoring During Labor
    Liang, Yiheng
    Liu, Ping
    Yan, Shaomei
    Li, Yun
    Chen, Duijin
    Fan, Shangrong
    MATERNAL-FETAL MEDICINE, 2022, 4 (02) : 113 - 120
  • [9] ELECTRONIC FETAL HEART-RATE MONITORING DURING LABOR - INFORMATION FROM RANDOMIZED TRIALS
    NEILSON, JP
    BIRTH-ISSUES IN PERINATAL CARE, 1994, 21 (02): : 101 - 104
  • [10] Fetal Heart Rate Monitoring in Labor
    不详
    JOURNAL OF MIDWIFERY & WOMENS HEALTH, 2021, 66 (02) : 285 - 286