Long-term follow-up of DDD pacing mode

被引:6
|
作者
Ulman, Mateusz [1 ]
Debski, Maciej [2 ]
Zabek, Andrzej [2 ]
Haberka, Kazimierz [2 ]
Lelakowski, Jacek [2 ,3 ]
Malecka, Barbara [2 ,3 ]
机构
[1] Med Ctr Niepolomice, Niepolomice, Poland
[2] John Paul II Hosp Krakow, Dept Electrocardiol, PL-31202 Krakow, Poland
[3] Jagiellonian Univ, Med Coll Krakow, Inst Cardiol, PL-31007 Krakow, Poland
关键词
sick sinus syndrome; atrioventricular block; DDD pacing; atrial fibrillation; endocardial lead failure; SICK SINUS SYNDROME; DUAL-CHAMBER; PACEMAKER; PERFORMANCE; BENEFITS; RISK;
D O I
10.5603/KP.a2014.0020
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and aim: The aim of this study was to determine the long-term survival of DDD pacing and identify the main reasons for its loss. Methods: The study group consisted of 496 patients in whom a DDD pacing system was implanted between October 1984 and March 2002 and who were followed up until July 2010. The follow-up period was 152.1 +/- 35.5 months. The patients' mean age at the time of implantation was 59.5 +/- 12.5 years, and 53.5% were male; 58% had sick sinus syndrome (SSS), 26% had atrioventricular block (AVB), 15% had both of these indications simultaneously, and 1% had other indications. The incidence of lead malfunction, progression to chronic atrial fibrillation (AF), and the rate of infective complications was analysed. Results: During the follow-up, 369 patients remained in DDD mode stimulation. DDD mode survival rate at one, five, ten and 15 years was, respectively, 96%, 86%, 77% and 72%. The most common reason for reprogramming out of DDD mode was the development of permanent AF in 65 (13.1%) patients. The occurrence of chronic AF was associated with a prior history of paroxysmal AF (p = 0.0001), SSS (p = 0.0215), and older age at time of implantation (p = 0.0068) compared to patients who remained in sinus rhythm. Lead malfunction caused loss of DDD mode pacing in 56 (11.3%) patients. Atrial leads were damaged in 37 patients, ventricular in 12 patients, and both leads in seven patients. The subclavian vein puncture was correlated with the mechanical damage of the atrial lead (p = 0.02935) compared to cephalic vein access. At the moment of complication, the patients with a dysfunctional lead were significantly younger than those who progressed to chronic AF (p = 0.0019). Infective complications which caused temporary loss of DDD pacing were observed in six patients: five had pocket infection and one had lead-dependent infective endocarditis. Conclusions: 1. Effective DDD pacing from the originally implanted system was noted in a high percentage (72%) of patients in long-term observation (15 years). 2. Progression to permanent AF is the most common reason for loss of DDD pacing; a history of paroxysmal AF and old age are the risk factors. 3. Subclavian vein puncture is associated with a higher rate of atrial lead damage.
引用
收藏
页码:519 / 526
页数:8
相关论文
共 50 条
  • [1] Prognosis of sick sinus syndrome with AAI vs DDD pacing: Long-term follow-up
    Broglia, P
    Ambrosini, F
    Arlotti, M
    Manfredini, R
    Perlini, S
    Sernesi, L
    Greco, EM
    EUROPACE '97 - THE OFFICIAL MEETING OF THE WORKING GROUPS ON CARDIAC PACING AND ARRHYTHMIAS OF THE EUROPEAN SOCIETY OF CARDIOLOGY, 1997, : 705 - 707
  • [3] LONG-TERM FOLLOW-UP OF TRANSVENOUS CARDIAC PACING IN CHILDREN
    EPSTEIN, ML
    KNAUF, DG
    ALEXANDER, JA
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 3 (02) : 485 - 485
  • [4] Long-term follow-up of pacemakers with an autocapture pacing system
    Erdinler, I
    Akyol, A
    Okmen, E
    Oguz, E
    Gurkan, K
    Ulufer, T
    JAPANESE HEART JOURNAL, 2002, 43 (06): : 631 - 641
  • [5] Long-term follow-up of pacing of the conditioned diaphragm in quadriplegia
    Elefteriades, JA
    Quin, JA
    Hogan, JF
    Holcomb, WG
    Letsou, GV
    Chlosta, WF
    Glenn, WWWL
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (06): : 897 - 906
  • [6] LONG-TERM FOLLOW-UP AFTER CARDIAC PACING IN BRADYARRHYTHMIAS
    FUKATANI, M
    HASHIBA, K
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1978, 42 (03): : 257 - 268
  • [7] LONG-TERM FOLLOW-UP OF BILATERAL PACING OF THE DIAPHRAGM IN QUADRIPLEGIA
    ELEFTERIADES, JA
    HOGAN, JF
    HANDLER, A
    LOKE, JS
    NEW ENGLAND JOURNAL OF MEDICINE, 1992, 326 (21): : 1433 - 1434
  • [8] Incidence of atrial fibrillation in patients with different mode of pacing. Long-term follow-up
    Folino, AF
    Buja, G
    Dal Corso, L
    Nava, A
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (01): : 260 - 263
  • [9] Long-term follow-up of His pacing in a single center experience
    Zanon, F.
    Marcantoni, L.
    Pastore, G.
    Baracca, E.
    Giau, G.
    Picariello, C.
    Aggio, S.
    Carraro, M.
    Roncon, L.
    Lanza, D.
    EUROPEAN HEART JOURNAL, 2017, 38 : 262 - 263
  • [10] Canadian trial of physiological pacing - Effects of physiological pacing during long-term follow-up
    Kerr, CR
    Connolly, SJ
    Abdollah, H
    Roberts, RS
    Gent, M
    Yusuf, S
    Gillis, AM
    Tang, ASL
    Talajic, M
    Klein, GJ
    Newman, DM
    CIRCULATION, 2004, 109 (03) : 357 - 362