Relationship between size-specific dose estimates and image quality in computed tomography depending on patient size

被引:16
|
作者
Kawashima, Hiroki [1 ]
Ichikawa, Katsuhiro [1 ]
Hanaoka, Shinsuke [2 ]
Matsubara, Kosuke [1 ]
Takata, Tadanori [2 ]
机构
[1] Kanazawa Univ, Inst Med Pharmaceut & Hlth Sci, Fac Hlth Sci, Kanazawa, Ishikawa, Japan
[2] Kanazawa Univ Hosp, Radiol Div, Kanazawa, Ishikawa, Japan
来源
基金
日本学术振兴会;
关键词
auto exposure control; computed tomography; contrast-to-noise ratio; image quality; patient size; size-specific dose estimates; AUTOMATIC EXPOSURE CONTROL; TUBE CURRENT MODULATION; PEDIATRIC CT; PHANTOM; OPTIMIZATION; MDCT; VOLTAGE; SYSTEMS; LIVER;
D O I
10.1002/acm2.12340
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study investigates the relationship between contrast-to-noise ratio (CNR) and size-specific dose estimate (SSDE) in computed tomography (CT) depending on patient size. In addition, the relationship to the auto exposure control (AEC) techniques is examined. A tissue-equivalent material having human-liver energy dependence is developed and used to evaluate these relationships. Three exposure dose levels (constant CT dose index, constant SSDE, and with AEC) are tested using four different phantom sizes (diameter: 15, 20, 25 and 30cm) in two different CT scanners (SOMATOM Definition Flash, Siemens, and LightSpeed VCT, GE). The contrast-to-noise ratios (CNRs) are measured using the developed phantom. It is found that the CNR increases with decreasing phantom size at constant SSDE, although the increase ratio is smaller than that of the constant CT dose index. This result indicates that the image characteristics differ even when the patient dose received from the CT examination is equivalent for each patient size. In the case of AEC use, the CNR results of the Siemens scanner exhibit a similar trend to those obtained for constant SSDE, for each phantom size. This suggests that the AEC technique that maintains a constant image quality (CARE Dose 4D) for each patient size corresponds well to the image quality obtained for constant SSDE. These findings facilitate further understanding of the relationship between image quality and exposure CT dose depending on patient size.
引用
收藏
页码:246 / 251
页数:6
相关论文
共 50 条
  • [31] Size-Specific Dose Estimates for Evaluation of Individual Patient Dose in CT Protocol for Renal Colic
    Waszczuk, Lukasz A.
    Guzinski, Maciej
    Czarnecka, Anna
    Sasiadek, Marek J.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 205 (01) : 100 - 105
  • [32] Novel method for calculating the effective dose using size-specific dose estimates conversion factors in abdomen-pelvis computed tomography
    Funashima, Kentaro
    Abiko, Shigeru
    Sato, Kazuhiro
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2023, 16 (04) : 506 - 515
  • [33] Relationship between radiation dose indices and patient size in adult brain computed tomography
    Mitsui, K.
    Inoue, Y.
    Itoh, H.
    Hata, H.
    Miyatake, H.
    Yamane, Takuro
    RADIOPROTECTION, 2025, 60 (01) : 37 - 43
  • [34] An Investigation of Using a Single CT Image to Assess Size-Specific Dose Estimates (SSDE)
    Abuhaimed, A.
    Martin, C.
    MEDICAL PHYSICS, 2020, 47 (06) : E514 - E514
  • [35] Simplifying Size-Specific Radiation Dose Estimates in Pediatric CT
    Khawaja, Ranish Deedar Ali
    Singh, Sarabjeet
    Vettiyil, Beth
    Lim, Ruth
    Gee, Michael
    Westra, Sjirk
    Kalra, Mannudeep K.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2015, 204 (01) : 167 - 176
  • [36] Size-specific Dose Estimates for Adult Patients at CT of the Torso
    Christner, Jodie A.
    Braun, Natalie N.
    Jacobsen, Megan C.
    Carter, Rickey E.
    Kofler, James M.
    McCollough, Cynthia H.
    RADIOLOGY, 2012, 265 (03) : 841 - 847
  • [37] Determining Size-Specific Dose Estimates for Head CT Examinations
    Hardy, A.
    Bostani, M.
    Zankl, M.
    McCollough, C.
    Cagnon, C.
    McNitt-Gray, M.
    MEDICAL PHYSICS, 2017, 44 (06) : 3314 - 3315
  • [38] Improved patient size estimates for accurate dose calculations in abdomen computed tomography
    Chang-Lae Lee
    Journal of the Korean Physical Society, 2017, 71 : 116 - 120
  • [39] Improved Patient Size Estimates for Accurate Dose Calculations in Abdomen Computed Tomography
    Lee, Chang-Lae
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 71 (02) : 116 - 120
  • [40] USEFULNESS OF SIZE-SPECIFIC DOSE ESTIMATES IN PEDIATRIC COMPUTED TOMOGRAPHY: REVALIDATION OF LARGE-SCALE PEDIATRIC CT DOSE SURVEY DATA IN JAPAN
    Yamazaki, Daisuke
    Miyazaki, Osamu
    Takei, Yasutaka
    Matsubara, Kosuke
    Shinozaki, Masafumi
    Shimada, Yoshiya
    Suzuki, Shoichi
    Muramatsu, Yoshihisa
    RADIATION PROTECTION DOSIMETRY, 2018, 179 (03) : 254 - 262