SCREEN-FILM VERSUS FULL-FIELD DIGITAL MAMMOGRAPHY: RADIATION DOSE AND IMAGE QUALITY IN A LARGE TEACHING HOSPITAL

被引:2
|
作者
Stantic, Tomislav J. [1 ]
Ciraj-Bjelac, Olivera E. [2 ]
Stojanovic, Saida S. [3 ]
Basta-Nikolic, Marijana D. [3 ]
Arandjic, Danijela D. [2 ]
Stoiljkovic, Dragana S. [4 ]
机构
[1] Ctr Diagnost, Subotica, Serbia
[2] Univ Belgrade, Vinca Inst Nucl Sci, Radiat Protect Dept, Belgrade, Serbia
[3] Clin Ctr Vojvodina, Novi Sad, Serbia
[4] St Elizabeth Krankenhaus, Bad Kissingen, Germany
来源
关键词
mammography; image quality; radiation dose; mean glandular dose; REFERENCE LEVEL; BREAST; PERFORMANCE; RADIOGRAPHY; ACCURACY; PROGRAM;
D O I
10.2298/NTRP1304398S
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The objective of this paper is to measure the radiation dose and image quality in conventional screen-film mammography and full-field digital mammography in women referred to mammography examination. Participants underwent bilateral, two-view screen-film mammography or full-field digital mammography. The visibility of anatomical regions and overall clinical image quality was rated by experienced radiologists. Total of 387 women and 1548 mammograms were enrolled in the study. Image quality was assessed in terms of image quality score, whereas patient dose assessment was performed in terms of mean glandular dose. Average mean glandular dose for cranio-caudal projection was 1.5 mGy and 2.1 mGy in full-field digital mammography and screen-film mammography, respectively. For medio-lateral oblique projection, corresponding values were 2.3 and 2.1 mGy. Overall image quality criteria scoring was 0.82 and 0.99 for screen-film and digital systems, respectively. The scores were in the range from 0.11 to 1.0 for different anatomical structures. Overall, full-field digital mammography was superior both in terms of image quality and dose over the screen-film mammography. The results have indicated that phantom dose values can assist in setting the optimisation activities in mammography and for comparison between mammography units. To obtain accurate diagnostic information with an acceptable radiation dose to breast, it is necessary to periodically perform patient dose and image quality surveys in all mammography units.
引用
收藏
页码:398 / 405
页数:8
相关论文
共 50 条
  • [41] Digital Mammography Versus Full-Field Digital Mammography
    Hall, Ferris M.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (01) : 240 - 240
  • [42] Optimization of technique factors for a silicon diode array full-field digital mammography system and comparison to screen-film mammography with matched average glandular dose
    Berns, EA
    Hendrick, RE
    Cutter, GR
    MEDICAL PHYSICS, 2003, 30 (03) : 334 - 340
  • [43] Comparison of full-field digital mammography with screen-film mammography for cancer detection: Results of 4,945 paired examinations
    Lewin, JM
    Hendrick, RE
    D'Orsi, CJ
    Isaacs, PK
    Moss, LJ
    Karellas, A
    Sisney, GA
    Kuni, CC
    Cutter, GR
    RADIOLOGY, 2001, 218 (03) : 873 - 880
  • [44] Comparison of full-field digital mammography to screen-film mammography with respect to contrast and spatial resolution in tissue equivalent breast phantoms
    Kuzmiak, CM
    Pisano, ED
    Cole, EB
    Zeng, DL
    Burns, CB
    Roberto, C
    Pavic, D
    Lee, Y
    Seo, BK
    Koomen, M
    Washburn, D
    MEDICAL PHYSICS, 2005, 32 (10) : 3144 - 3150
  • [45] Computer-aided detection of masses in full-field digital mammography using screen-film mammograms for training
    Kallenberg, Michiel
    Karssemeijer, Nico
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23): : 6879 - 6891
  • [46] PATIENT DOSES FROM SCREEN-FILM AND FULL-FIELD DIGITAL MAMMOGRAPHY IN A POPULATION-BASED SCREENING PROGRAMME
    Hauge, I. H. R.
    Pedersen, K.
    Sanderud, A.
    Hofvind, S.
    Olerud, H. M.
    RADIATION PROTECTION DOSIMETRY, 2012, 148 (01) : 65 - 73
  • [47] Repeat Rates and Their Effect on Dose in Transitioning from Screen/Film to Full-Field Digital Mammography
    Ripley, E.
    Otto, P.
    Hatab, M.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [48] Full-Field Digital versus Screen-Film Mammography: Comparison within the UK Breast Screening Program and Systematic Review of Published Data
    Vinnicombe, Sarah
    Pinto Pereira, Snehal M.
    McCormack, Valerie A.
    Shiel, Susan
    Perry, Nick
    Silva, Isabel M. dos Santos
    RADIOLOGY, 2009, 251 (02) : 347 - 358
  • [49] Optimization of radiation dose and image quality in diagnostic radiology: Digital and screen-film imaging
    Gray, J
    Strauss, K
    MEDICAL PHYSICS, 2004, 31 (06) : 1805 - 1805
  • [50] The readout thickness versus the measured thickness for a range of screen film mammography and full-field digital mammography units
    Hauge, Ingrid H. R.
    Hogg, Peter
    Szczepura, Katy
    Connolly, Paul
    McGill, George
    Mercer, Claire
    MEDICAL PHYSICS, 2012, 39 (01) : 263 - 271