Transcatheter occlusion of the patent ductus arteriosus with Cook detachable coils

被引:82
|
作者
Tometzki, AJP
Arnold, R
Peart, I
Sreeram, N
Abdulhamed, JM
Godman, MJ
Patel, RG
Kitchiner, DJ
BuLock, FA
Walsh, KP
机构
[1] ROYAL MANCHESTER CHILDRENS HOSP,MANCHESTER M27 1HA,LANCS,ENGLAND
[2] ROYAL LIVERPOOL CHILDRENS HOSP,LIVERPOOL L7 7DG,MERSEYSIDE,ENGLAND
[3] BIRMINGHAM CHILDRENS HOSP,BIRMINGHAM,W MIDLANDS,ENGLAND
[4] PRINCE SULTAN CARDIAC CTR,RIYADH,SAUDI ARABIA
关键词
ductus arteriosus; transcatheter occlusion; Cook spring coils; coils;
D O I
10.1136/hrt.76.6.531
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective-To report initial experience with a new occlusion device for native and residual patent ductus arteriosus. Design-Descriptive study of consecutive non-randomised patients undergoing a new method of patent ductus arteriosus closure with detachable coils. Setting-Tertiary centres for paediatric cardiology. Patients-71 consecutive patients, aged 1.2-22 years, with a patent ductus arteriosus (PDA) underwent elective transcatheter closure. 45 had native PDAs (group A) with a minimum diameter of 1.0 mm-5.0 mm (median 2.0 mm). A further 26 had undergone one or more previous occlusion attempts (group B). Interventions-A total of 133 detachable (Cook) spring coils were successfully implanted in 70 patients. The procedure was performed transvenously in 51 patients, retrograde arterially in 13, and by both routes in a further 6 patients. One 5 mm coil migrated but was successfully retrieved. Main outcome measures-In group A colour flow Doppler echocardiography showed that complete occlusion was achieved in 40/45 (89%) at 24 hours, 41/45 (91%) at 1 month, and 44/45 (98%) by 6 months post procedure. Occlusion rates in residual PDAs were 22/25 (88%) occluded at 24 hours, 23/25 (92%) at 1 month, and 24/25 (96%) at 6 months follow up. Conclusions-Transcatheter occlusion using detachable (Cook) spring coils is a safe and effective alternative to presently available devices. The delivery system allows full retrieval of the coil until a satisfactory position is obtained.
引用
收藏
页码:531 / 535
页数:5
相关论文
共 50 条
  • [1] Coil occlusion of patent ductus arteriosus with detachable coils
    Akagi, T
    Hashino, K
    Sugimura, T
    Ishii, M
    Eto, G
    Kato, H
    AMERICAN HEART JOURNAL, 1997, 134 (03) : 538 - 543
  • [2] Patent ductus arteriosus occlusion using detachable coils
    Bermúdez-Cañete, R
    Santoro, G
    Bialkowsky, J
    Herraiz, I
    Formigari, R
    Szkutnik, M
    Ballerini, L
    AMERICAN JOURNAL OF CARDIOLOGY, 1998, 82 (12): : 1547 - +
  • [3] Detachable coils for occlusion of patent ductus arteriosus in 2 dogs
    Tanaka, R
    Hoshi, K
    Nagashima, Y
    Fujii, Y
    Yamane, Y
    VETERINARY SURGERY, 2001, 30 (06) : 580 - 584
  • [4] TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH GIANTURCO COILS
    LLOYD, TR
    FEDDERLY, R
    MENDELSOHN, AM
    SANDHU, SK
    BEEKMAN, RH
    CIRCULATION, 1993, 88 (04) : 1412 - 1420
  • [5] Radiographic characteristics of Cook detachable and Gianturco coils as well as clinical results of transcatheter closure of the patent ductus arteriosus
    Galal, MO
    von Sinner, W
    Al-Fadley, F
    Solymar, L
    Bulbul, Z
    Boecker, J
    Fawzy, ME
    ZEITSCHRIFT FUR KARDIOLOGIE, 1999, 88 (12): : 1006 - 1014
  • [6] Radiographic characteristics of Cook detachable and Gianturco coils as well as clinical results of transcatheter closure of the patent ductus arteriosus
    M. O. Galal
    W. von Sinner
    F. Al-Fadley
    L. Solymar
    Z. Bulbul
    J. Boecker
    M. E. Fawzy
    Zeitschrift für Kardiologie, 1999, 88 : 1006 - 1014
  • [7] Percutaneous closure of patent ductus arteriosus: complementary use of detachable Cook patent ductus arteriosus coils and Amplatzer duct occluders
    Podnar, T
    Gavora, P
    Masura, J
    EUROPEAN JOURNAL OF PEDIATRICS, 2000, 159 (04) : 293 - 296
  • [8] Percutaneous closure of patent ductus arteriosus: complementary use of detachable Cook patent ductus arteriosus coils and Amplatzer duct occluders
    Tomaž Podnar
    Pavol Gavora
    Jozef Masura
    European Journal of Pediatrics, 2000, 159 : 293 - 296
  • [9] Transcatheter occlusion of patent ductus arteriosus using tornado platinum coils
    Suda, K
    Matsumura, M
    PEDIATRICS INTERNATIONAL, 2003, 45 (01) : 45 - 48
  • [10] Safety and efficacy of interventional occlusion of patent ductus arteriosus with detachable coils:: a multicentre experience
    Hofbeck, M
    Bartolomaeus, G
    Buheitel, G
    Esser, R
    Grävinghoff, L
    Hoffmann, W
    Kienast, W
    Michel-Behnke, I
    Scharabrine, EG
    Schranz, D
    Schmaltz, AA
    Shakhov, BE
    Singer, H
    Lindinger, A
    EUROPEAN JOURNAL OF PEDIATRICS, 2000, 159 (05) : 331 - 337