Quantum efficiency of a digital x-ray imaging system for small animal studies

被引:2
|
作者
Wang, Q [1 ]
Wu, XZ [1 ]
Chen, WR [1 ]
Cheung, J [1 ]
Liu, H [1 ]
机构
[1] Univ Oklahoma, Ctr Bioengn, Norman, OK 73019 USA
关键词
noise power spectrum (NPS); detective quantum efficiency (DQE); X-ray imaging; small animal studies;
D O I
10.1117/12.543429
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study is to characterize the quantum efficiency of a digital x-ray imaging system. This system is designed for small animal experiments. It is equipped with an x-ray tube with a Be-filtered tungsten target, 0.02 mm focal spot and 0.3 mA operating current. The fiber optically coupled CCD module, through a scintillating screen, covers a field of 5 cm X 10 cm, at 1024 X 2048 pixel format. To analyze the impact of x-ray photon flux and CCD electronic noise to the quantum efficiency of the system, the noise power spectrum (NPS) and detective quantum efficiency (DQE) were measured as a function of x-ray exposure and detector integration time. A BR- 12 phantom was placed between the x-ray tube and the detector during measurements. The results showed consistent DQE when exposure/integration is in the range of 2 to 7 seconds (for a 0.5cm thick phantom), and 3 to 15 seconds (for a 2cm thick phantom). With a 0.5cm phantom, DQE are approximately 36.7%, 25% and 5.4% at frequencies of 0 lp/mm, 3 lp/mm and 8 lp/mm respectively. With a 2 cm BR-12 slab, hardened x-ray beam at 26 KVp doesn't have much impact on DQE, approximately 36.2%, 27.3% and 6% for 01p/mm, 31p/nim. and 81p/mm frequencies. In summary, the CCD based digital x-ray imaging system investigated in this study is an efficient, x-ray quantum noise limited system for small animal experiments.
引用
收藏
页码:94 / 98
页数:5
相关论文
共 50 条
  • [31] Low Dose X-Ray Sources and High Quantum Efficiency Sensors: The Next Challenge in Dental Digital Imaging?
    Mistry, Arnav R.
    Feldman, Daniel Uzbelger
    Yang, Jie
    Ryterski, Eric
    RADIOLOGY RESEARCH AND PRACTICE, 2014, 2014
  • [32] An x-ray image guidance system for small animal stereotactic irradiation
    Song, K. H.
    Pidikiti, R.
    Stojadinovic, S.
    Speiser, M.
    Seliounine, S.
    Saha, D.
    Solberg, T. D.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23): : 7345 - 7362
  • [33] Evaluation of Detective Quantum Efficiency of Digital X-Ray Detectors in Clinical Environment
    Zelikman M.I.
    Kruchinin S.A.
    Biomedical Engineering, 2015, 48 (5) : 273 - 276
  • [34] Monitoring of quantum efficiency of digital x-ray detectors based on photostimulated luminophores
    Zelikman M.I.
    Kabanov S.P.
    Kruchinin S.A.
    Lobov D.P.
    Biomedical Engineering, 2008, 42 (5) : 249 - 252
  • [35] Application of Computed Radiography in Small Animal Imaging: X-Ray Contrast Agents
    Maghsoodpour, A.
    Keyes, G.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [36] Impact of additive noise on system performance of a digital X-ray imaging system
    Zhang, Da
    Rong, John
    Chen, Wei R.
    Gao, Feng
    Xu, Kexin
    Wu, Xizeng
    Liu, Hong
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (01) : 69 - 73
  • [37] Detective quantum efficiency and Swank noise for spectroscopic X-ray imaging detectors
    Tanguay, Jesse
    Richtsmeier, Devon
    Dydula, Christopher
    Iniewski, Kris
    Bazalova-Carter, Magdalena
    MEDICAL PHYSICS, 2021, 48 (08) : 4668 - 4669
  • [38] The X-Ray Digital Imaging System Designing Based on EMCCD Detector
    Liu Xia
    Wu Baoping
    Liu Chongguang
    ISTM/2009: 8TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1-6, 2009, : 671 - 673
  • [39] Development of a digital panoramic X-ray imaging system for dental applications
    Lee, S. Y.
    Choi, S. I.
    Cho, H. S.
    Oh, J. E.
    Kim, K. Y.
    Koo, Y. S.
    Kim, T. W.
    Lee, B. S.
    Kirn, S.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 2987 - +
  • [40] Calibration of a soft X-ray digital imaging system for biological materials
    Kotwaliwale, N.
    Subbiah, J.
    Weckler, P. R.
    Brusewitz, G. H.
    Kranzler, G. A.
    TRANSACTIONS OF THE ASABE, 2007, 50 (02): : 661 - 666