Accuracy of target-controlled infusion (TCI) with 2 different propofol formulations

被引:0
|
作者
Ihmsen, H [1 ]
Jeleazcov, C [1 ]
Schüttler, J [1 ]
Schwilden, H [1 ]
Bremer, F [1 ]
机构
[1] Univ Erlangen Nurnberg, Anasthesiol Klin, D-91054 Erlangen, Germany
来源
ANAESTHESIST | 2004年 / 53卷 / 10期
关键词
propofol; formulation; target-controlled infusion (TCI); pharmacokinetics; cross-over study;
D O I
10.1007/s00101-004-0753-6
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. Target-controlled infusion (TCI) of propofol was initially realized as a device for prefilled syringes (Diprifusor(R)). New TCI systems can be used with any propofol formulation. We compared two different propofol formulations with respect to accuracy of TCI and pharmacokinetics. Materials and met hods. A total of 10 volunteers received Diprivan(R) 1% and Propofol 1% MCT Fresenius(R) as TO using the pharmacokinetic model of the Diprifusor(R). The prediction error was determined from measured arterial concentrations. A three-compartment model was fitted to the concentration data. Results. The median prediction error and the median absolute prediction error were -1.4% and 23.3% for Diprivan(R), and -5.9% and 17.8% for Propofol Fresenius(R). The drugs did not differ in pharmacokinetics but showed a smaller central volume of distribution than used for infusion control. Conclusions. The pharmacokinetic model of Diprifusor(R) can also be used for TCI of Propofol Fresenius(R). The large volume of distribution in this model may cause an overshoot in concentration.
引用
收藏
页码:937 / 943
页数:7
相关论文
共 50 条
  • [41] The effects of two levels of midazolam (M) and propofol (P), using target-controlled infusion (TCI), on memory and sedation.
    de Roode, A
    van Gerven, JMA
    Cohen, AF
    Schoemaker, RC
    Bovill, JG
    CLINICAL PHARMACOLOGY & THERAPEUTICS, 1998, 63 (02) : 187 - 187
  • [42] What target-controlled infusion of propofol for COPA and LMA placement?
    Fanelli, G
    Casati, A
    Casaletti, E
    Aldegheri, G
    Tarantino, F
    Torri, G
    ANESTHESIOLOGY, 1998, 89 (3A) : U423 - U423
  • [43] Concentrations of propofol in cerebral spinal fluid: target-controlled infusion
    罗爱伦
    易杰
    郭向阳
    任洪智
    黄宇光
    叶铁虎
    ChineseMedicalJournal, 2004, (02)
  • [44] Laryngeal mask insertion during target-controlled infusion of propofol
    Baik, HJ
    Kim, JH
    Lee, CH
    JOURNAL OF CLINICAL ANESTHESIA, 2001, 13 (03) : 175 - 181
  • [45] Influence of Bayesian optimization on the performance of propofol target-controlled infusion
    van den Berg, J. P.
    Eleveld, D. J.
    De Smet, T.
    van den Heerik, A. V. M.
    van Amsterdam, K.
    Lichtenbelt, B. J.
    Scheeren, T. W. L.
    Absalom, A. R.
    Struys, M. M. R. F.
    BRITISH JOURNAL OF ANAESTHESIA, 2017, 119 (05) : 918 - 927
  • [46] Pharmacokinetics of propofol administered by target-controlled infusion to alcoholic patients
    Servin, FS
    Bougeois, B
    Gomeni, R
    Mentré, F
    Farinotti, R
    Desmonts, JM
    ANESTHESIOLOGY, 2003, 99 (03) : 576 - 585
  • [47] Modelling the effect site compartment in a target-controlled infusion of propofol
    Croall, A.
    Hawthorne, C.
    Shaw, M.
    BRITISH JOURNAL OF ANAESTHESIA, 2017, 119 (02) : E3 - E4
  • [48] Science vs supposition: the case of target-controlled propofol infusion
    Kempen, P. M.
    BRITISH JOURNAL OF ANAESTHESIA, 2011, 106 (04) : 603 - 603
  • [49] INDUCTION OF ANESTHESIA WITH PROPOFOL USING A TARGET-CONTROLLED INFUSION SYSTEM
    CHAUDHRI, S
    WHITE, M
    KENNY, GNC
    ANAESTHESIA, 1992, 47 (07) : 551 - 553
  • [50] Concentrations of propofol in cerebral spinal fluid: target-controlled infusion
    Luo, AL
    Yi, J
    Guo, XY
    Ren, HZ
    Huang, YG
    Ye, TH