摘要
Microwave frequency power measurement is one of the main types of the measurement for measuring instruments and systems in the radio frequency range. Therefore, improving the accuracy of measuring the microwave frequency power requires the establishment of more precise standards, and the development of calibration methods for meters of microwave frequency power is an urgent task. Microwave frequency power standards that are used to calibrate the relevant measuring instruments must ensure high accuracy of the unit size reproduction over a wide measurement and frequency range.The study allowed determining typical calibration schemes for meters of microwave frequency power. For measurements, the calibration scheme for meters of microwave frequency power by the method of a direct comparison with the help of a calibrator when measuring the absorbed power of microwave frequencies is substantiated and suggested.The proposed methodology for evaluating the uncertainty of absorbed power measurements can be used when calibrating power meters in the frequency range from 30 MHz to 18 GHz. It allows determining the most significant components of the combined standard uncertainty of the absorbed power measurements of ultrahigh frequencies, as well as to receive the result of the corresponding calibration. This methodology can also be used to evaluate the uncertainty of microwave frequency directional measurements.
引用
收藏
页码:9 / 14
页数:6
相关论文
共 50 条
  • [31] Electronic power supply used for calibration of electric energy meters
    Elektronski vir za umerjanje števcev električne energije
    Zajec, P. (peter.zajec@fe.uni-lj.si), 2001, Electrotechnical Society of Slovenia (68):
  • [32] Calibration of current transducers at high frequencies
    Ruddle, AR
    Pomeroy, SC
    Ward, DD
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2001, 43 (01) : 100 - 104
  • [33] Navigating the features and limitations of laser power and energy meters
    Lara, H
    LASER FOCUS WORLD, 2003, 39 (01): : 137 - +
  • [34] RELATIVISTIC ELECTRON BEAM INTERACTIONS FOR GENERATION OF HIGH POWER AT MICROWAVE FREQUENCIES.
    Godlove, T.F.
    Granatstein, V.L.
    1600, IEEE (77CH1219-5 MTT), New York, NY
  • [35] High PRF Microwave Power Module with Electronic Power Conditioner of enhanced features
    Martorana, Rosario
    Bisconti, David
    Mistretta, Antonino
    2017 EIGHTEENTH INTERNATIONAL VACUUM ELECTRONICS CONFERENCE (IVEC), 2017,
  • [36] HIGH-POWER INTEGRATED OPTIC IR MODULATOR AT MICROWAVE-FREQUENCIES
    CHEO, PK
    GILDEN, M
    APPLIED PHYSICS LETTERS, 1976, 28 (10) : 626 - 627
  • [37] About calibration features of microwave noise meter
    Belskiy, AY
    Gusinskiy, AV
    Dzisiak, AB
    Kostrikin, AM
    12TH INTERNATIONAL CONFERENCE - MICROWAVE & TELECOMMUNICATION TECHNOLOGY, CONFERENCE PROCEEDINGS, 2002, : 532 - 533
  • [38] ON THE CALIBRATION OF CONDUCTIVITY METERS
    SMITH, LG
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1953, 24 (10): : 998 - 998
  • [39] On calibration of pH meters
    Cheng, KL
    Zhu, DM
    SENSORS, 2005, 5 (4-5) : 209 - 219
  • [40] CRITICAL NOTE ON THE USE OF INCIDENT POWER METERS OF MICROWAVE GENERATORS
    JUTTE, BAHG
    AGTERDENBOS, J
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1979, 34 (2-3) : 133 - 135