Displacement of the double-lumen endobronchial tube can be detected by bronchial cuff pressure change

被引:16
|
作者
Araki, K [1 ]
Nomura, R [1 ]
Urushibara, R [1 ]
Yoshikawa, Y [1 ]
Hatano, Y [1 ]
机构
[1] WAKAYAMA MED COLL, DEPT ANESTHESIOL, WAKAYAMA 640, JAPAN
来源
ANESTHESIA AND ANALGESIA | 1997年 / 84卷 / 06期
关键词
D O I
10.1097/00000539-199706000-00032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We measured the bronchial cuff pressure of left-sided double-lumen endobronchial tubes (DLTs) in 54 patients to confirm the effect of DLT displacement on cuff pressure. After positioning the cephalad surface of the bronchial cuff of the DLT 2.5 cm distal to the carina (23 patients in the first part of the study) or just below the carina (23 patients in the second part), the cuff was withdrawn in 0.5-cm steps during right-sided, one-lung ventilation. The bronchial cuff pressure was measured, and the capnogram and pressure-volume loop, displayed by a side-stream spirometer, was evaluated. After positioning the cephalad surface of the bronchial cuff just below the carina (eight patients in the third part), bronchial cuff pressure was measured while confirming the position of the bronchial cuff during the 30 min of the operative procedure. The bronchial cuff pressure decreased significantly by 28.4 cm H2O (P < 0.01) and 21.3 cm H2O (P < 0.01) in the first and second parts, respectively, before the pressure-volume loop or the capnogram changed. The bronchial cuff pressure in the third part showed no significant change. We conclude that bronchial cuff pressure monitoring was very helpful in detecting displacement of the DLT during right-sided, one-lung ventilation.
引用
收藏
页码:1349 / 1353
页数:5
相关论文
共 50 条
  • [41] Changes in the Bronchial Cuff Pressure of Left-Sided Double-Lumen Endotracheal Tube by Lateral Positioning: A Prospective Observational Study
    Kim, Jong-Hae
    Kim, Eugene
    Kim, In-Young
    Choi, Eun-Joo
    Byun, Sung-Hye
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (08)
  • [42] Rapid deflation of the bronchial cuff of the double-lumen tube after decreasing the concentration of inspired nitrous oxide
    Karasawa, F
    Takita, A
    Takamatsu, I
    Mori, T
    Oshima, T
    Kawatani, Y
    ANESTHESIA AND ANALGESIA, 2002, 95 (01): : 238 - 242
  • [43] UNUSUAL DIFFICULTY WITH DOUBLE-LUMEN ENDO-BRONCHIAL TUBE
    JENKINS, V
    ANAESTHESIA, 1963, 18 (02) : 236 - &
  • [44] BRONCHIAL TRAUMA SECONDARY TO INTUBATION WITH A PVC DOUBLE-LUMEN TUBE
    PEDEN, CJ
    SMITH, RB
    GALIZIA, EJ
    JOURNAL OF THE ROYAL SOCIETY OF MEDICINE, 1992, 85 (11) : 705 - 706
  • [45] BRONCHIAL RUPTURE, A COMPLICATION OF USE OF THE ROBERTSHAW DOUBLE-LUMEN TUBE
    HEISER, M
    STEINBERG, JJ
    MACVAUGH, H
    KLINEBERG, PL
    ANESTHESIOLOGY, 1979, 51 (01) : 88 - 88
  • [46] Ventilation failure resulting from defective double-lumen endobronchial tube
    Irie, T
    Kurahashi, K
    Ogawa, K
    Furuya, R
    Yamada, Y
    ANESTHESIA AND ANALGESIA, 2005, 100 (06): : 1866 - 1866
  • [47] Airway troubles related to the double-lumen endobronchial tube in thoracic surgery
    Taguchi H.
    Yamada K.
    Matsumoto H.
    Kato A.
    Imanishi T.
    Shingu K.
    Journal of Anesthesia, 1997, 11 (3) : 173 - 178
  • [48] 180° rotation of double-lumen endobronchial tube during intubation Reply
    Bahk, J. -H.
    Sea, J. -H.
    BRITISH JOURNAL OF ANAESTHESIA, 2014, 112 (05) : 938 - 938
  • [49] Predicting the size of a double-lumen endobronchial tube based on tracheal diameter
    Chow, MYH
    Liam, BL
    Lew, TWK
    Chelliah, RY
    Ong, BC
    ANESTHESIA AND ANALGESIA, 1998, 87 (01): : 158 - 160
  • [50] DOUBLE-LUMEN ENDOBRONCHIAL TUBE INTUBATION USING A RETROGRADE WIRE TECHNIQUE
    ALFERY, DD
    ANESTHESIA AND ANALGESIA, 1993, 76 (06): : 1374 - 1375