Displacement of the double-lumen endobronchial tube can be detected by bronchial cuff pressure change

被引:16
|
作者
Araki, K [1 ]
Nomura, R [1 ]
Urushibara, R [1 ]
Yoshikawa, Y [1 ]
Hatano, Y [1 ]
机构
[1] WAKAYAMA MED COLL, DEPT ANESTHESIOL, WAKAYAMA 640, JAPAN
来源
ANESTHESIA AND ANALGESIA | 1997年 / 84卷 / 06期
关键词
D O I
10.1097/00000539-199706000-00032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
We measured the bronchial cuff pressure of left-sided double-lumen endobronchial tubes (DLTs) in 54 patients to confirm the effect of DLT displacement on cuff pressure. After positioning the cephalad surface of the bronchial cuff of the DLT 2.5 cm distal to the carina (23 patients in the first part of the study) or just below the carina (23 patients in the second part), the cuff was withdrawn in 0.5-cm steps during right-sided, one-lung ventilation. The bronchial cuff pressure was measured, and the capnogram and pressure-volume loop, displayed by a side-stream spirometer, was evaluated. After positioning the cephalad surface of the bronchial cuff just below the carina (eight patients in the third part), bronchial cuff pressure was measured while confirming the position of the bronchial cuff during the 30 min of the operative procedure. The bronchial cuff pressure decreased significantly by 28.4 cm H2O (P < 0.01) and 21.3 cm H2O (P < 0.01) in the first and second parts, respectively, before the pressure-volume loop or the capnogram changed. The bronchial cuff pressure in the third part showed no significant change. We conclude that bronchial cuff pressure monitoring was very helpful in detecting displacement of the DLT during right-sided, one-lung ventilation.
引用
收藏
页码:1349 / 1353
页数:5
相关论文
共 50 条
  • [1] Bronchial cuff pressure change caused by left-sided double-lumen endobronchial tube displacement
    Kazukuni Araki
    Ryoichi Nomura
    Reiko Urushibara
    Yukiko Yoshikawa
    Yoshio Hatano
    Canadian Journal of Anesthesia, 2000, 47 : 775 - 779
  • [2] Bronchial cuff pressure change caused by left-sided double-lumen endobronchial tube displacement
    Araki, K
    Nomura, R
    Urushibara, R
    Yoshikawa, Y
    Hatano, Y
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2000, 47 (08): : 775 - 779
  • [3] BRONCHIAL CUFF PRESSURE OF ENDOBRONCHIAL DOUBLE-LUMEN TUBES
    NETO, PPR
    ANESTHESIA AND ANALGESIA, 1990, 71 (02): : 209 - 209
  • [4] Double-lumen endobronchial tube cuff pressures: a need for monitoring?
    Daunt, M.
    Latter, J.
    ANAESTHESIA, 2013, 68 (10) : 1089 - 1089
  • [5] A DOUBLE-LUMEN ENDOBRONCHIAL TUBE
    BRYCESMITH, R
    BRITISH JOURNAL OF ANAESTHESIA, 1959, 31 (06) : 274 - 275
  • [6] Con: A Bronchial Blocker Is Not a Substitute for a Double-Lumen Endobronchial Tube
    Brodsky, Jay B.
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2015, 29 (01) : 237 - 239
  • [7] CONTINUOUS SPIROMETRY FOR DETECTION OF DOUBLE-LUMEN ENDOBRONCHIAL TUBE DISPLACEMENT
    BARDOCZKY, GI
    LEVARLET, M
    ENGELMAN, E
    DEFRANCQUEN, P
    BRITISH JOURNAL OF ANAESTHESIA, 1993, 70 (05) : 499 - 502
  • [8] Bronchial rupture by a double-lumen endobronchial tube during staging thoracoscopy
    Gilbert, TB
    Goodsell, CW
    Krasna, MJ
    ANESTHESIA AND ANALGESIA, 1999, 88 (06): : 1252 - 1253
  • [9] Positioning the double-lumen endobronchial tube
    Jae-Hyon Bahk
    Ho-Geol Ryu
    Byun-Moon Ham
    Canadian Journal of Anesthesia, 2002, 49 : 526 - 527
  • [10] Positioning the double-lumen endobronchial tube
    Bahk, JH
    Ryu, HG
    Ham, BM
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2002, 49 (05): : 526 - 527