摘要
Purpose: We report on a series of patients with erosion of 1 cuff of an AMS800 dagger double cuff artificial urinary sphincter and determine the success rate of removing the eroded cuff, leaving the other components behind and converting the device to a single cuff system. Materials and Methods: In 9 patients with a double cuff artificial sphincter 1 eroded cuff was removed and the system was converted to a single cuff sphincter. Results: Followup was based on an incontinence scoring system. In 1 patient the remaining cuff was removed because it also eroded. In another patient a wound infection developed 5 months after placement and the entire device was removed. The remaining 7 patients had better continence scores than before artificial sphincter placement but each remarked that continence was better with 2 cuffs. Conclusions: An eroded cuff of a double cuff artificial urinary sphincter system can be successfully removed and the device can be converted to a single cuff system.
引用
收藏
页码:85 / 86
页数:2
相关论文
共 50 条
  • [21] ANDROGEN DEPRIVATION THERAPY IS ASSOCIATED WITH URETHRAL ATROPHY AND ARTIFICIAL URINARY SPHINCTER CUFF EROSION
    Davenport, Michael
    Akhtar, Abdulhadi
    Shakir, Nabeel
    Yi, Yooni
    Bergeson, Rachel
    Morey, Allen
    JOURNAL OF UROLOGY, 2019, 201 (04): : E573 - E573
  • [22] IS REMOVAL OF THE PRESSURE REGULATING BALLOON NECESSARY AFTER ARTIFICIAL URINARY SPHINCTER CUFF EROSION?
    Rozanski, A. T.
    Viers, B. R.
    Shakir, N. A.
    McKibben, M. J.
    Pagliara, T. J.
    Scott, J. M.
    Morey, A. F.
    JOURNAL OF SEXUAL MEDICINE, 2018, 15 (02): : S64 - S64
  • [23] WHEN TO SUSPECT ARTIFICIAL URINARY SPHINCTER CUFF EROSION: AN ANALYSIS OF PRESENTING SIGNS AND SYMPTOMS
    Diao, L.
    Nealon, S.
    Wolfe, A.
    Dropkin, B.
    Joice, G.
    Sanders, S.
    Hudak, S.
    Morey, A.
    JOURNAL OF SEXUAL MEDICINE, 2022, 19 (04): : S78 - S78
  • [24] 3.5 cm Artificial Urinary Sphincter Cuff Erosion Occurs Predominantly in Irradiated Patients
    Simhan, Jay
    Morey, Allen F.
    Singla, Nirmish
    Tausch, Timothy J.
    Scott, J. Francis
    Lemack, Gary E.
    Roehrborn, Claus G.
    JOURNAL OF UROLOGY, 2015, 193 (02): : 593 - 597
  • [25] Conformation and mechanics of the polymeric cuff of artificial urinary sphincter
    Natali, Arturo Nicola
    Fontanella, Chiara Giulia
    Todros, Silvia
    Pavan, Piero G.
    Carmignato, Simone
    Zanini, Filippo
    Carniel, Emanuele Luigi
    MATHEMATICAL BIOSCIENCES AND ENGINEERING, 2020, 17 (04) : 3894 - 3908
  • [26] Most men with artificial urinary sphincter cuff erosion have low serum testosterone levels
    Wolfe, Avery R.
    Ortiz, Nicolas M.
    Baumgarten, Adam S.
    VanDyke, Maia E.
    West, Mary L.
    Dropkin, Benjamin M.
    Joice, Gregory A.
    Sanders, Sarah C.
    Hudak, Steven J.
    Morey, Allen F.
    NEUROUROLOGY AND URODYNAMICS, 2021, 40 (04) : 1035 - 1041
  • [27] NO CLEAR INDICATION FOR PENILE IMPLANT REMOVAL AT THE TIME OF ARTIFICIAL URINARY SPHINCTER EXPLANTATION FOR CUFF EROSION
    Gaines, Tyler M.
    Loh-Doyle, Jeffrey C.
    Maas, Marisa
    Markarian, Emily
    Michel, Joaquin
    Matluck, Nicole
    Cordon, Billy
    Klein, Roger
    Myrga, John
    Rusilko, Paul
    Livingston, Austin
    Grimaud, Logan
    Foreman, Jordan
    Lentz, Aaron
    Peterson, Andrew
    Mikhail, Mark
    Hwang, Alex
    Shridharani, Anand
    Tryfonyuk, Liliya
    Martins, Francisco
    Gonzalez, Rebecca
    Van Der Voort, Grant
    Bhalla, Rohan
    Flynn, Brian
    Nikolavsky, Dmitriy
    Simhan, Jay
    JOURNAL OF UROLOGY, 2024, 211 (05): : E1247 - E1247
  • [28] Transcorporal artificial urinary sphincter cuff placement in cases requiring revision for erosion and urethral atrophy
    Guralnick, ML
    Miller, E
    Toh, KL
    Webster, GD
    JOURNAL OF UROLOGY, 2002, 167 (05): : 2075 - 2078
  • [29] Is Removal of the Pressure-regulating Balloon Necessary After Artificial Urinary Sphincter Cuff Erosion?
    Rozanski, Alexander T.
    Viers, Boyd R.
    Shakir, Nabeel A.
    Pagliara, Travis J.
    Scott, Jeremy M.
    Morey, Allen F.
    UROLOGY, 2018, 113 : 225 - 229
  • [30] Comparison of outcomes after single or double-cuff artificial urinary sphincter insertion
    O'Connor, RC
    Gerber, GS
    Avila, D
    Chen, AA
    Bales, GT
    UROLOGY, 2003, 62 (04) : 723 - 726