Randomised comparison of Pentax AirwayScope and Glidescope for tracheal intubation in patients with normal airway anatomy

被引:31
|
作者
Teoh, W. H. L. [1 ]
Shah, M. K. [1 ]
Sia, A. T. H. [1 ]
机构
[1] KK Womens & Childrens Hosp, Dept Womens Anaesthesia, Singapore, Singapore
基金
英国医学研究理事会;
关键词
VIDEO LARYNGOSCOPE; MACINTOSH LARYNGOSCOPE; PERFORATION; SCOPE; VIDEOLARYNGOSCOPE;
D O I
10.1111/j.1365-2044.2009.06032.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
P>We compared intubating characteristics of the Pentax AirwayScope and Glidescope in a randomised controlled trial involving 140 patients. We found significantly shorter intubation times with mean (SD) 21.3 (12.3) vs 30.2 (13.2) s, lower intubating difficulty scores 4.4 (10.4) vs 12.8 (16.3) p < 0.001, and better grade 1 laryngeal views with the Pentax AirwayScope (95.7 vs 81.4%, p = 0.015). Significantly more optimisation manoeuvres were needed to successfully intubate with the Glidescope, with significant inability to align the tip of the tracheal tube with the glottic opening with the Glidescope in 9 (13%) vs no patients, p = 0.013. There was more mucosal bleeding (4 (5.7%) vs 1 (1.4%), p = 0.366), more lip bleeding (3 (4.3%) vs 1 (1.4%), p = 0.620) and significantly more postoperative sorethroat, 13 (18.6%) vs 0, p < 0.001 associated with the Glidescope.
引用
收藏
页码:1125 / 1129
页数:5
相关论文
共 50 条
  • [21] Cardiovascular responses to tracheal intubation with the Airway Scope (Pentax-AWS)
    Akihiro Suzuki
    Yuki Toyama
    Norifumi Katsumi
    Takayuki Kunisawa
    John J. Henderson
    Hiroshi Iwasaki
    Journal of Anesthesia, 2008, 22 : 100 - 101
  • [22] The Glidescope® for tracheal intubation in patients with ankylosing spondylitis
    Gunaydin, B.
    Gungor, I.
    Yigit, N.
    Celebi, H.
    BRITISH JOURNAL OF ANAESTHESIA, 2007, 98 (03) : 408 - 409
  • [23] Cardiovascular responses to tracheal intubation with the Airway Scope (Pentax-AWS)
    Suzuki, Akihiro
    Toyama, Yuki
    Katsumi, Norifumi
    Kunisawa, Takayuki
    Henderson, John J.
    Iwasaki, Hiroshi
    JOURNAL OF ANESTHESIA, 2008, 22 (01) : 100 - 101
  • [24] A randomised clinical trial comparing the flexible fibrescope and the Pentax Airway Scope (AWS)® for awake oral tracheal intubation
    Mendonca, C.
    Mesbah, A.
    Velayudhan, A.
    Danha, R.
    ANAESTHESIA, 2016, 71 (08) : 908 - 914
  • [25] Comparison of the Pentax, Truview, GlideScope, and the Miller laryngoscope for child intubation during resuscitation
    Szarpak, Lukasz
    Czyzewski, Lukasz
    Kurowski, Andrzej
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2015, 33 (03): : 391 - 395
  • [26] Comparison of the Pentax Airwayscope, Glidescope Video Laryngoscope, and Macintosh Laryngoscope During Chest Compression According to Bed Height
    Kim, Wonhee
    Lee, Yoonje
    Kim, Changsun
    Lim, Tae Ho
    Oh, Jaehoon
    Kang, Hyunggoo
    Lee, Sanghyun
    MEDICINE, 2016, 95 (05)
  • [27] The use of the GlideScope® for tracheal intubation in patients with ankylosing spondylitis
    Lai, H. Y.
    Chen, I. H.
    Chen, A.
    Hwang, F. Y.
    Lee, Y.
    BRITISH JOURNAL OF ANAESTHESIA, 2006, 97 (03) : 419 - 422
  • [28] Performance of GlideScope Cobalt® for tracheal intubation in pediatric patients
    Xue, F. S.
    Cheng, Y.
    Li, R. P.
    MINERVA ANESTESIOLOGICA, 2013, 79 (02) : 214 - 215
  • [29] Tracheal intubation with videolaryngoscopes in patients with cervical spine immobilization: a randomized trial of the Airway Scope® and the GlideScope®
    Liu, E. H. C.
    Goy, R. W. L.
    Tan, B. H.
    Asai, T.
    BRITISH JOURNAL OF ANAESTHESIA, 2009, 103 (03) : 446 - 451
  • [30] Reply to letter: Tracheal intubation during chest compressions performed by qualified emergency physicians unfamiliar with the Pentax-Airwayscope
    Shin, Dong Hyuk
    Han, Sang Kuk
    Choi, Pil Cho
    Sim, Min Seob
    Lee, Jeong Hun
    Park, Sang O.
    EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2013, 20 (02) : 144 - 144