A nondestructive calibration method for maximizing the range and accuracy of AFM force measurement

被引:3
|
作者
Xie, Hui [1 ]
Rong, Weibin [1 ]
Wu, Aiwen [1 ]
Sun, Lining [1 ]
机构
[1] Harbin Inst Technol, State Key Lab Robot & Syst, Harbin 150080, Peoples R China
基金
对外科技合作项目(国际科技项目);
关键词
atomic force microscope; nondestructive calibration; nonlinear compensation; dual-probe; MICROSCOPY; FRICTION; CANTILEVER; SURFACE; TIP;
D O I
10.1088/0960-1317/24/2/025005
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, a nondestructive method for the normal and lateral sensitivity calibrations of the optical lever in atomic force microscope (AFM) is presented. The practical application of this method in a dual-probe AFM is discussed in detail. To calibrate the conversion factors between photodiode responses and probe's deflection angles accurately without applying forces to the probe, a two-degrees-of-freedom flexure-hinge-based calibration device (FHCD) is developed. The device, which mainly consists of two mutually perpendicular flexure-hinge levers that share the same rotational center, serves as a switching mechanism for precise translation-to-rotation conversions both in the normal and lateral directions. During the calibration, a probe is attached to the FHCD at the meeting of the rotational axes of two levers. The FHCD is mounted on an AFM sample platform. The probe in this method acts as a mirror to be normally and laterally tilted at nanoscale angles to deflect the reflected laser beam by twisting the corresponding flexure-hinge levers, rather than the force-inducing probe deflection in traditional AFM calibration methods. With this method, the nondestructive calibration of the local and full-range sensitivities of the optical levers can be completed without destroying the probe tip or modifying the actual system setup of an AFM. Moreover, the nonlinearities of the optical levers are accurately compensated. Experimental results show that the linear ranges (with a deviation of 5% in the full range) of the force measurement are extended to 3.6 and 4.5 times in the normal and lateral directions, respectively, increasing to over 90% of the full range of the force measurement.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Calibration of Lateral Force of AFM Measurement
    Gao, Qi
    Wu, Guangfu
    Wai, King
    Lai, Chiu
    2017 IEEE 17TH INTERNATIONAL CONFERENCE ON NANOTECHNOLOGY (IEEE-NANO), 2017, : 121 - 124
  • [2] Direct force balance method for AFM lateral force calibration
    Asay, David B.
    Kim, Seong H.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2006, 231
  • [3] An Improved Method of Adhesion Force Measurement by Atomic Force Microscopy (AFM)
    Jin, Hong
    Li, Xiongyao
    Wen, Yuanyun
    Tang, Hong
    Zeng, Xiaojia
    ATOMIC SPECTROSCOPY, 2022, 43 (01) : 13 - 18
  • [4] Enhanced accuracy of force application for AFM nanomanipulation using nonlinear calibration of optical levers
    Xie, Hui
    Vitard, Julien
    Haliyo, Dogan Sinan
    Regnier, Stephane
    IEEE SENSORS JOURNAL, 2008, 8 (7-8) : 1478 - 1485
  • [5] Novel method of measuring cantilever deflection during an AFM force measurement
    Hlady, V
    Pierce, M
    Pungor, A
    LANGMUIR, 1996, 12 (22) : 5244 - 5246
  • [6] AFM lateral force calibration for an integrated probe using a calibration grating
    Wang, Huabin
    Gee, Michelle L.
    ULTRAMICROSCOPY, 2014, 136 : 193 - 200
  • [7] Maximizing the Measurement Accuracy of Digitized Signals
    Haney, Mike
    2010 IEEE AUTOTESTCON, 2010, : 255 - 260
  • [8] Quantitative electrostatic force measurement in AFM
    Jeffery, S
    Oral, A
    Pethica, JB
    APPLIED SURFACE SCIENCE, 2000, 157 (04) : 280 - 284
  • [9] Viscosity Effect On The AFM Force Measurement
    Yang, Guoliang
    Liu, Runcong
    BIOPHYSICAL JOURNAL, 2009, 96 (03) : 397A - 397A
  • [10] A METHOD OF INCREASING RANGE MEASUREMENT ACCURACY BY PARAMETRIC IDENTIFICATION OF THE MEASUREMENT CHANNELS
    PETRENKO, PB
    STRYUKOV, BA
    TELECOMMUNICATIONS AND RADIO ENGINEERING, 1990, 45 (09) : 64 - 66