Magnetically suspended centrifugal blood pump with a self bearing motor

被引:22
|
作者
Masuzawa, T [1 ]
Onuma, H [1 ]
Kim, SJ [1 ]
Okada, Y [1 ]
机构
[1] Ibaraki Univ, Fac Engn, Dept Mech Engn, Hitachi, Ibaraki 3168511, Japan
关键词
D O I
10.1097/00002480-200207000-00019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A magnetically suspended centrifugal blood pump with a self bearing motor has been developed for long-term ventricular assistance. A rotor of the self bearing motor is actively suspended and rotated by an electromagnetic field without mechanical bearings. Radial position of the rotor is controlled actively, and axial position of the rotor is passively stable within the thin rotor structure. An open impeller and a semiopened impeller were examined to determine the best impeller structure. The outer diameter and height of the impeller are 63 and 34 mm, respectively. Both the impellers indicated similar pump performance. Single volute and double volute structures were also tested to confirm the performance of the double volute. Power consumption for levitation and radial displacement of the impeller with a rotational speed of 1,500 rpm were 0.7 W and 0.04 mm in the double volute, while those in the single volute were 1.3 W and 0.07 mm, respectively. The stator of the self bearing motor was redesigned to avoid magnetic saturation and improve motor performance. Maximum flow rate and pressure head were 9 L/min and 250 mm Hg, respectively. The developed magnetically suspended centrifugal blood pump is a candidate for an implantable left ventricular assist device.
引用
收藏
页码:437 / 442
页数:6
相关论文
共 50 条
  • [1] Magnetically suspended centrifugal blood pump with an axially levitated motor
    Masuzawa, T
    Ezoe, S
    Kato, T
    Okada, Y
    ARTIFICIAL ORGANS, 2003, 27 (07) : 631 - 638
  • [2] CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    NAKAZEKI, T
    ITOH, H
    ARTIFICIAL ORGANS, 1992, 16 (03) : 305 - 308
  • [3] Modeling and control of a magnetic bearing system for the magnetically suspended centrifugal blood pump
    Kim, H
    Kim, HC
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2000, 23 (10): : 689 - 696
  • [4] A magnetically suspended and hydrostatically stabilized centrifugal blood pump
    Hart, RM
    Filipenco, VG
    Kung, RTV
    ARTIFICIAL ORGANS, 1996, 20 (06) : 591 - 596
  • [5] Use of motor current in flow rate measurement for the magnetically suspended centrifugal blood pump
    Tsukiya, T
    Akamatsu, T
    Nishimura, K
    Yamada, T
    Nakazeki, T
    ARTIFICIAL ORGANS, 1997, 21 (05) : 396 - 401
  • [6] RECENT STUDIES OF THE CENTRIFUGAL BLOOD PUMP WITH A MAGNETICALLY SUSPENDED IMPELLER
    AKAMATSU, T
    TSUKIYA, T
    NISHIMURA, K
    PARK, CH
    NAKAZEKI, T
    ARTIFICIAL ORGANS, 1995, 19 (07) : 631 - 634
  • [7] Performance and unsteady characteristics of magnetically suspended centrifugal blood pump
    Yoshino, Y
    Akamatsu, T
    JSME INTERNATIONAL JOURNAL SERIES B-FLUIDS AND THERMAL ENGINEERING, 1997, 40 (01) : 114 - 120
  • [8] Magnetically suspended rotary blood pump with radial type combined motor-bearing
    Masuzawa, T
    Kita, T
    Matsuda, K
    Okada, Y
    ARTIFICIAL ORGANS, 2000, 24 (06) : 468 - 474
  • [9] Magnetically suspended centrifugal blood pump with a radial magnetic driver
    Hoshi, H
    Katakoa, K
    Ohuchi, K
    Asama, J
    Shinshi, T
    Shimokohbe, A
    Takatani, S
    ASAIO JOURNAL, 2005, 51 (01) : 60 - 64
  • [10] Design of a self-bearing slice motor for a centrifugal blood pump
    Ueno, S
    Chen, C
    Ohishi, T
    Matsuda, K
    Okada, Y
    Taenaka, Y
    Masuzawa, T
    PROCEEDINGS OF THE SIXTH INTERNATIONAL SYMPOSIUM ON MAGNETIC BEARINGS, 1998, : 143 - 151