Effect of conformity and contact stress on wear in fixed-bearing total knee prostheses

被引:75
|
作者
Galvin, Alison L. [1 ]
Kang, Lu [1 ]
Udofia, Itoro [1 ]
Jennings, Louise M. [1 ]
McEwen, Hannah M. J. [2 ]
Jin, Zhongmin [1 ]
Fisher, John [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Inst Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
[2] DePuy Int Ltd, Leeds, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
Total knee replacement; Polyethylene wear; MOLECULAR-WEIGHT POLYETHYLENE; TOTAL JOINT REPLACEMENTS; TIBIAL COMPONENTS; PRESSURE PREDICTION; DELAMINATION; KINEMATICS; UHMWPE; DESIGN;
D O I
10.1016/j.jbiomech.2009.05.010
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Ultra high molecular weight polyethylene (PE) remains the primary bearing surface of choice in total knee replacements (TKR). Wear is controlled by levels of cross-shear motion and contact stress. The aim of this study was to compare the wear of fixed-bearing total knee replacements with curved and flat inserts and to test the hypothesis that the flat inserts which give higher contact stresses and smaller contact areas would lead to lower levels of surface wear. A low-conforming, high contact stress knee with a low-medium level of cross shear resulted in significantly lower wear rates in comparison to a standard cruciate sacrificing fixed-bearing knee. The low wear solution found in the knee simulator was supported by fundamental studies of wear as a function of pressure and cross shear in the pin on plate system. Current designs of fixed-bearing knees do not offer this low wear solution due to their medium cross shear, moderate conformity and medium contact stress. (C) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1898 / 1902
页数:5
相关论文
共 50 条
  • [31] Comparison of Functional Performance after Total Knee Arthroplasty Using Rotating Platform and Fixed-bearing Prostheses with or without Patellar Resurfacing
    Chen, Liao-bin
    Tan, Yang
    Al-Aidaros, Mohammed
    Wang, Hua
    Wang, Xin
    Cai, Shu-han
    ORTHOPAEDIC SURGERY, 2013, 5 (02) : 112 - 117
  • [32] Comparison between mobile-bearing and fixed-bearing knees in bilateral total knee replacements
    D. S. Pandher
    K. J. Oh
    S. H. Lee
    International Orthopaedics, 2007, 31 : 131 - 132
  • [33] Rotational Kinematics of a Modern Fixed-Bearing Posterior Stabilized Total Knee Arthroplasty
    Mahoney, Ormonde M.
    Kinsey, Tracy L.
    Banks, Anne Z.
    Banks, Scott A.
    JOURNAL OF ARTHROPLASTY, 2009, 24 (04): : 641 - 645
  • [34] Comparison between mobile-bearing and fixed-bearing knees in bilateral total knee replacements
    Pandher, D. S.
    Oh, K. J.
    Lee, S. H.
    INTERNATIONAL ORTHOPAEDICS, 2007, 31 (01) : 131 - 132
  • [35] Osteolysis in failed total knee arthroplasty: A comparison of mobile-bearing and fixed-bearing knees
    Huang, CH
    Ma, HM
    Liau, JJ
    Ho, FY
    Cheng, CK
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (12): : 2224 - 2229
  • [36] Comparison between mobile-bearing and fixed-bearing knees in bilateral total knee replacements
    Watanabe, T
    Tomita, T
    Fujii, M
    Hashimoto, J
    Sugamoto, K
    Yoshikawa, H
    INTERNATIONAL ORTHOPAEDICS, 2005, 29 (03) : 179 - 181
  • [37] Comparison of mobile-bearing and fixed-bearing total knee arthroplasty - A prospective randomized study
    Aglietti, P
    Baldini, A
    Buzzi, R
    Lup, D
    De Luca, L
    JOURNAL OF ARTHROPLASTY, 2005, 20 (02): : 145 - 153
  • [38] Comparison between mobile-bearing and fixed-bearing knees in bilateral total knee replacements
    T. Watanabe
    T. Tomita
    M. Fujii
    J. Hashimoto
    K. Sugamoto
    H. Yoshikawa
    International Orthopaedics, 2005, 29 : 179 - 181
  • [39] WEAR AND CONFORMITY IN TOTAL KNEE REPLACEMENT
    TRENT, PS
    WALKER, PS
    WEAR, 1976, 36 (02) : 175 - 187
  • [40] The effect of insert conformity and material on total knee replacement wear
    Abdelgaied, Abdellatif
    Brockett, Claire L.
    Liu, Feng
    Jennings, Louise M.
    Jin, Zhongmin
    Fisher, John
    PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2014, 228 (01) : 98 - 106