Peak-fitting of high resolution ToF-SIMS spectra: a preliminary study

被引:6
|
作者
Abel, M. L. [1 ]
Shimizu, K. [1 ]
Holliman, M. [1 ]
Watts, J. F. [1 ]
机构
[1] Univ Surrey, Fac Engn & Phys Sci, Surface Anal Lab, Guildford GU2 7XH, Surrey, England
关键词
peak-fitting; ToF-SIMS; roughness; charging; quantification; XPS;
D O I
10.1002/sia.3036
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Peak-fitting has been performed on a series of peaks obtained by time-of-flight secondary ion mass spectrometry (ToF-SIMS) analysis in order to assess whether information may be obtained from this procedure on the sa m pies' characteristics. A variety of samples were examined including a range of treatments for aluminium leading to different surface roughnesses, polymer films with a range of polydispersities, molecular weight (MW) and thicknesses as well as aluminium samples with adsorbed adhesion promoters on the surface. Variation of peak-fitting was assessed by varying the peak intensity, full width at half maximum (FWHM) and peak asymmetry. Although further studies are needed it is possible to say that the peak width increases with roughness whereas peak asymmetry seems to be related to oxide thickness. Polymer characteristics do not seem to influence the width whereas the peak asymmetry increases either versus MW or polydispersity. A possible assumption is that the peak asymmetry relates to the ion formation processes. Additional work with varying polymer films thickness indicates that both FWHM and peak asymmetry may be related to sample charging and this could be used for assessment of film thicknesses. Finally, peak-fitting was used to obtain a more reliable peak area when peaks are too close in mass to use current methods. Copyright (C) 2009 John Wiley & Sons, Ltd.
引用
收藏
页码:265 / 268
页数:4
相关论文
共 50 条
  • [31] Curve resolution for multivariate images with applications to TOF-SIMS and Raman
    Gallagher, NB
    Shaver, JM
    Martin, EB
    Morris, J
    Wise, BM
    Windig, W
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2004, 73 (01) : 105 - 117
  • [32] High-resolution TOF-SIMS imaging of eukaryotic cells preserved in a trehalose matrix
    Parry, S
    Winograd, N
    ANALYTICAL CHEMISTRY, 2005, 77 (24) : 7950 - 7957
  • [33] High resolution surface imaging of Co/ZrO2 catalyst by TOF-SIMS
    Grams, J
    Góralski, J
    Paryjczak, T
    SURFACE SCIENCE, 2004, 549 (01) : L21 - L26
  • [34] Bioimaging TOF-SIMS: High resolution 3D Imaging of single cells
    Nygren, Hakan
    Hagenhoff, Birgit
    Malmberg, Per
    Nilsson, Mikael
    Richter, Katrin
    MICROSCOPY RESEARCH AND TECHNIQUE, 2007, 70 (11) : 969 - 974
  • [35] Ga+ TOF-SIMS lineshape analysis for resolution enhancement of MALDI MS spectra of a peptide mixture
    Malyarenko, DI
    Chen, H
    Wilkerson, AL
    Tracy, ER
    Cooke, WE
    Manos, DM
    Sasinowski, M
    Semmes, OJ
    APPLIED SURFACE SCIENCE, 2004, 231 : 357 - 361
  • [36] SIMS/TOF-SIMS study of microparticles: surface analysis, imaging and quantification
    Gijbels, R
    Verlinden, G
    Geuens, I
    MICROBEAM ANALYSIS 2000, PROCEEDINGS, 2000, (165): : 331 - 332
  • [37] Characterization of high-k dielectrics with ToF-SIMS
    Ferrari, S
    APPLIED SURFACE SCIENCE, 2004, 231 : 609 - 613
  • [38] High resolution TOF-SIMS study of varying chain length self-assembled monolayer surfaces.
    Wolf, KV
    Cole, DA
    Bernasek, SL
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2002, 224 : U453 - U453
  • [39] High-resolution TOF-SIMS study of varying chain length self-assembled monolayer surfaces
    Wolf, KV
    Cole, DA
    Bernasek, SL
    ANALYTICAL CHEMISTRY, 2002, 74 (19) : 5009 - 5016
  • [40] Positive secondary ion mass spectrum of poly(methyl methacrylate):: a high mass resolution ToF-SIMS study
    Briggs, D
    Fletcher, IW
    Gonçalves, NM
    SURFACE AND INTERFACE ANALYSIS, 2000, 29 (05) : 303 - 309