Peak-fitting of high resolution ToF-SIMS spectra: a preliminary study

被引:6
|
作者
Abel, M. L. [1 ]
Shimizu, K. [1 ]
Holliman, M. [1 ]
Watts, J. F. [1 ]
机构
[1] Univ Surrey, Fac Engn & Phys Sci, Surface Anal Lab, Guildford GU2 7XH, Surrey, England
关键词
peak-fitting; ToF-SIMS; roughness; charging; quantification; XPS;
D O I
10.1002/sia.3036
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Peak-fitting has been performed on a series of peaks obtained by time-of-flight secondary ion mass spectrometry (ToF-SIMS) analysis in order to assess whether information may be obtained from this procedure on the sa m pies' characteristics. A variety of samples were examined including a range of treatments for aluminium leading to different surface roughnesses, polymer films with a range of polydispersities, molecular weight (MW) and thicknesses as well as aluminium samples with adsorbed adhesion promoters on the surface. Variation of peak-fitting was assessed by varying the peak intensity, full width at half maximum (FWHM) and peak asymmetry. Although further studies are needed it is possible to say that the peak width increases with roughness whereas peak asymmetry seems to be related to oxide thickness. Polymer characteristics do not seem to influence the width whereas the peak asymmetry increases either versus MW or polydispersity. A possible assumption is that the peak asymmetry relates to the ion formation processes. Additional work with varying polymer films thickness indicates that both FWHM and peak asymmetry may be related to sample charging and this could be used for assessment of film thicknesses. Finally, peak-fitting was used to obtain a more reliable peak area when peaks are too close in mass to use current methods. Copyright (C) 2009 John Wiley & Sons, Ltd.
引用
收藏
页码:265 / 268
页数:4
相关论文
共 50 条
  • [1] High resolution surface analysis by TOF-SIMS
    Hagenhoff, B
    MIKROCHIMICA ACTA, 2000, 132 (2-4) : 259 - 271
  • [2] High Resolution Surface Analysis by TOF-SIMS
    Birgit Hagenhoff
    Microchimica Acta, 2000, 132 (2-4) : 259 - 271
  • [3] Peak detection of TOF-SIMS using continuous wavelet transform and curve fitting
    Zheng, Ying
    Tian, Di
    Liu, Ke
    Bao, Zemin
    Wang, Peizhi
    Qiu, Chunling
    Liu, Dunyi
    Fan, Runlong
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2018, 428 : 43 - 48
  • [4] HIGH-RESOLUTION TOF-SIMS STUDIES OF SUBSTITUTED POLYSTYRENES
    CHIARELLI, MP
    PROCTOR, A
    BLETSOS, IV
    HERCULES, DM
    FELD, H
    LEUTE, A
    BENNINGHOVEN, A
    MACROMOLECULES, 1992, 25 (25) : 6970 - 6976
  • [5] Polyolefin Homopolymer and Copolymer Spectra by ToF-SIMS
    Abmayr, David W., Jr.
    SURFACE SCIENCE SPECTRA, 2006, 13 (01): : 117 - 165
  • [6] EFFECT OF STEREOREGULARITY ON TOF-SIMS SPECTRA OF POLYMERS
    ZIMMERMAN, PA
    HERCULES, DM
    APPLIED SPECTROSCOPY, 1994, 48 (05) : 620 - 622
  • [7] High-mass-resolution ToF-SIMS study of chlorine-containing polymers
    Briggs, D
    Fletcher, IW
    Gonçalves, NM
    SURFACE AND INTERFACE ANALYSIS, 2002, 33 (03) : 178 - 184
  • [8] High-resolution imaging and proteomics of peptide fragments by TOF-SIMS
    Nygren, Hakan
    Malmberg, Per
    PROTEOMICS, 2010, 10 (08) : 1694 - 1698
  • [9] High-resolution peak analysis in TOF SIMS data
    Gelb, Lev D.
    Shahrokh Esfahani, Negar
    Walker, Amy V.
    SURFACE AND INTERFACE ANALYSIS, 2021, 53 (01) : 53 - 67
  • [10] Peak fitting to resolve CN- isotope ratios in biological and environmental samples using TOF-SIMS
    Cliff, JB
    Gaspar, DJ
    Bottomley, PJ
    Myrold, DD
    APPLIED SURFACE SCIENCE, 2004, 231 : 912 - 916