Same-Patient Reproducibility of State Entropy: A Comparison of Simultaneous Bilateral Measurements During General Anesthesia

被引:8
|
作者
Ozcan, Mehmet S. [1 ]
Thompson, David M. [2 ]
Cure, Jorge [1 ]
Hine, J. Randal [3 ]
Roberts, Pamela R. [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Dept Anesthesiol, Coll Med, Oklahoma City, OK 73104 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Dept Biostat, Coll Med, Oklahoma City, OK 73104 USA
[3] Univ Iowa Hosp & Clin, Dept Anesthesia, Iowa City, IA 52242 USA
来源
ANESTHESIA AND ANALGESIA | 2009年 / 108卷 / 06期
关键词
BISPECTRAL INDEX; SPECTRAL ENTROPY; AGREEMENT; MONITOR; VALIDATION; ARTERIAL; PROPOFOL; MODULE;
D O I
10.1213/ane.0b013e31819dc668
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: State Entropy (R) (SE) is an index of anesthetic depth similar to Bispectral Index (R) (BIS). Both indices use a single-channel electroencephalogram, recorded from a unilaterally applied electrode on the forehead, as their input. Intrapatient reproducibility of BIS was questioned in a recent study in which simultaneous measurements from two electrodes applied to the same patient showed conflicting anesthetic depths. Our purpose was to determine whether SE results are similarly reproducible, even though their computation uses a different algorithm than BIS. In this study, we investigated the reproducibility of SE measurements simultaneously obtained from bilaterally applied electrodes in the same patient. METHODS: Entropy electrodes were applied bilaterally on 21 patients under general inhaled anesthesia. Simultaneous SE measurements from both electrodes were recorded every 10 s from each patient. Data were analyzed with Bland-Altman statistics. RESULTS: We obtained 14,379 pairs of SE measurements. Four percent of the individual measurements suggested conflicting anesthetic depth along with a numeric difference more than 10 points. Bias was not clinically significant (-0.3). Ninety-five percent limits of agreement were -11.7 and +11.6. CONCLUSIONS: SE showed a clinically significant degree of disagreement when probes were applied on both sides of the forehead in the same patient. Bland-Altman statistics showed better same-patient reproducibility in SE than did a similar study on BIS. Nevertheless, 4%, of the simultaneously measured pairs of SE suggested different anesthetic depths and differed by more than 10 points. Caution is advised when using SE as a clinical index of anesthetic depth. (Anesth Analg 2009;108:1830-5)
引用
收藏
页码:1830 / 1835
页数:6
相关论文
共 19 条
  • [11] Simultaneous vs Sequential Bilateral Cataract Surgery for Infants With Congenital Cataracts: Weighing the Risks of General Anesthesia During Infancy vs Endophthalmitis
    Lambert, Scott R.
    Dave, Hreem
    Phoenix, Vidya
    Becker, Edmund R.
    ARCHIVES OF OPHTHALMOLOGY, 2011, 129 (04) : 524 - 525
  • [12] A COMPARISON OF TRANSCUTANEOUS, END-TIDAL AND ARTERIAL MEASUREMENTS OF CARBON-DIOXIDE DURING GENERAL-ANESTHESIA
    REID, CW
    MARTINEAU, RJ
    MILLER, DR
    HULL, KA
    BAINES, J
    SULLIVAN, PJ
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1992, 39 (01): : 31 - 36
  • [13] Simultaneous vs Sequential Bilateral Cataract Surgery for Infants With Congenital Cataracts: Weighing the Risks of General Anesthesia During Infancy vs Endophthalmitis reply
    Kushner, Burton J.
    ARCHIVES OF OPHTHALMOLOGY, 2011, 129 (04) : 525 - 526
  • [14] Update on the SedLine® algorithm for calculating the Patient State Index of older individuals during general anesthesia: a randomized controlled trial
    Obara, Shinju
    Oishi, Rieko
    Nakano, Yuko
    Noji, Yoshie
    Ebana, Hideaki
    Isosu, Tsuyoshi
    Kurosawa, Shin
    Murakawa, Masahiro
    MINERVA ANESTESIOLOGICA, 2021, 87 (07) : 774 - 785
  • [15] Effects of bilateral superficial cervical plexus block on sevoflurane consumption during thyroid surgery under entropy-guided general anesthesia: a prospective randomized study
    Kannan, Sudheesh
    Surhonne, Nethra S.
    Kumar, Chethan R.
    Kavitha, B.
    Rani, Devika D.
    Rao, Raghavendra R. S.
    KOREAN JOURNAL OF ANESTHESIOLOGY, 2018, 71 (02) : 141 - 148
  • [16] Comparison of Single-Use Negative-Pressure Wound Therapy (sNPWT) and Standard Dressings Applied to the Same Patient During Bilateral Tissue Expander-to-Implant Exchanges
    Molska, Maja
    Wojciech, Magdalena
    Murawa, Dawid
    CANCERS, 2025, 17 (01)
  • [17] Simultaneous comparison of depth of sedation performance between SedLine and BIS during general anesthesia using custom passive interface hardware: study protocol for a prospective, non-blinded, non-randomized trial
    Jones, James Harvey
    Nittur, Vinay Ravikumar
    Fleming, Neal
    Applegate, Richard L., II
    BMC ANESTHESIOLOGY, 2021, 21 (01)
  • [18] Simultaneous comparison of depth of sedation performance between SedLine and BIS during general anesthesia using custom passive interface hardware: study protocol for a prospective, non-blinded, non-randomized trial
    James Harvey Jones
    Vinay Ravikumar Nittur
    Neal Fleming
    Richard L. Applegate
    BMC Anesthesiology, 21
  • [19] Variations in Values of State, Response Entropy and Haemodynamic Parameters Associated with Development of Different Epileptiform Patterns during Volatile Induction of General Anaesthesia with Two Different Anaesthetic Regimens Using Sevoflurane in Comparison with Intravenous Induct: A Comparative Study
    Stasiowski, Michal
    Dulawa, Anna
    Szumera, Izabela
    Marciniak, Radoslaw
    Niewiadomska, Ewa
    Kaspera, Wojciech
    Krawczyk, Lech
    Ladzinski, Piotr
    Grabarek, Beniamin Oskar
    Jalowiecki, Przemyslaw
    BRAIN SCIENCES, 2020, 10 (06) : 1 - 27