Validation of a point of care lipid analyser using a hospital based reference laboratory

被引:71
|
作者
Carey, M.
Markham, C.
Gaffney, P.
Boran, G.
Maher, V. [1 ]
机构
[1] Adelaide Meath Hosp, Dept Cardiol, Dublin 24, Ireland
[2] Adelaide Meath Hosp, Dept Chem Pathol, Dublin 24, Ireland
[3] Natl Childrens Hosp, Dublin, Ireland
关键词
High Density Lipoprotein Cholesterol; IRISH Journal; Venous Sample; Lipid Measurement; Finger Prick;
D O I
10.1007/BF03167964
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Lipid measurements are essential in cardiovascular disease management in primary care. However, utilizing hospital laboratories may result in excess patient travel, sample loss and repeat clinic visits. Point of care (POC) lipid measurement would overcome these difficulties but has not been validated in an Irish setting. Aim To compare POC lipid profile measurements using a cholestech LDX analyser with a hospital reference laboratory (Lab). Method One-hundred subjects (30 men,70 women) participated. Finger prick and venous samples were analysed directly by Cholestech LDX and the Lab. Results A broad range of lipid values were measured. Absolute differences between POC and Lab measurements were insignificant except for a small over-estimation by the POC method of triglyceride 0.25 mmol/l (95% CI 0.17 to 0.24), and an underestimation of HDLc - 0.11 mmol/l (95% CI -0.143 to -0.078). There were significant correlations between POC and Lab. levels; total cholesterol r = 0.92, triglyceride r = 0.93, HDLc r = 0.92 and LDLc r = 0.86 (all p < 0.0001). Conclusion These results validate the use of the Cholestech LDX(R) analyser for point of care lipid measurements in clinical practice, provided well trained operators are supported by a hospital laboratory delivering quality assurance support.
引用
收藏
页码:30 / 35
页数:6
相关论文
共 50 条
  • [11] Haemoglobin A1c: Comparing performance of two point of care devices with laboratory analyser
    Wan Mohd Zin R.M.
    Ahmad Kamil Z.I.
    Tuan Soh T.R.
    Embong M.
    Wan Mohamud W.N.
    BMC Research Notes, 6 (1)
  • [12] REFERENCE VALUES BASED ON HOSPITAL ADMISSION LABORATORY DATA
    HARWOOD, SJ
    COLE, GW
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1978, 240 (03): : 270 - 274
  • [13] Lipid Profile in Ambulatory Patients Using 3 Point-of-Care Devices and Comparison With Reference Methods
    Mendez-Gonzalez, Jesus
    Bonet-Marques, Rosa
    Ordonez-Llanos, Jordi
    POINT OF CARE, 2009, 8 (03): : 110 - 116
  • [14] Lipid Profile in Ambulatory Subjects Using 3 Point-of-Care Devices and Comparison With Reference Methods
    Mendez-Gonzalez, J.
    Bonet-Marques, R.
    Ordonez-Llanos, Jordi
    POINT OF CARE, 2010, 9 (02): : 102 - 107
  • [16] Point of care monitoring of coagulation during liver transplantation using the TAS coagulation analyser
    Gibbs, NM
    Weightman, WM
    Thackray, NM
    Michalopoulos, N
    ANESTHESIOLOGY, 1999, 91 (3A) : U263 - U263
  • [17] Symptom based approaches in point of care laboratory testing
    Risch, Lorenz
    Senn, Oliver
    THERAPEUTISCHE UMSCHAU, 2015, 72 (02) : 105 - 112
  • [18] Reproducibility of thrombelastometry (ROTEM®): Point-of-care versus hospital laboratory performance
    Haas, Thorsten
    Spielmann, Nelly
    Mauch, Jacqueline
    Speer, Oliver
    Schmugge, Markus
    Weiss, Markus
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2012, 72 (04): : 313 - 317
  • [19] Safe exclusion of pulmonary embolism using the wells rule and a point-of-care or laboratory based D-dimer test in primary care: a diagnostic validation study
    Geersing, G-J
    Erkens, P.
    Lucassen, W.
    Buller, H.
    Hoes, A.
    Oudega, R.
    Prins, M.
    Stoffers, H.
    ten Cate, H.
    van Weert, H.
    Moons, K.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 : 605 - 605
  • [20] Staffing to Workload in a Hospital-Based Immunohematology Reference Laboratory (IRL)
    Moldenhauer, S.
    Bendix, B.
    Stubbs, J. R.
    Jacob, E. K.
    TRANSFUSION, 2013, 53 : 260A - 260A