Unexpected high failure rate of a specific MicroPort/Livallova/Sorin pacing lead

被引:12
|
作者
Haeberlin, Andreas [1 ,2 ,3 ]
Anwander, Marie-Theres [1 ]
Kueffer, Thomas [1 ]
Tholl, Maximilien [2 ]
Baldinger, Samuel [1 ]
Servatius, Helge [1 ]
Lam, Anna [1 ]
Franzeck, Florian [1 ]
Asatryan, Babken [1 ]
Zurbuchen, Adrian [2 ]
Tanner, Hildegard [1 ]
Reichlin, Tobias [1 ]
Roten, Laurent [1 ]
Noti, Fabian [1 ]
机构
[1] Univ Bern, Bern Univ Hosp, Dept Cardiol, Inselspital, Bern, Switzerland
[2] Univ Bern, Sitem Ctr Translat Med & Biomed Entrepreneurship, Bern, Switzerland
[3] Bordeaux Univ Hosp CHU, CardioThorac Unit, Pessac, France
关键词
Beflex; CapSureFix; Complication; Lead failure; Lead survival; Medtronic; MicroPort; Oversensing; Pacemaker; Vega; COMPLICATIONS; MULTICENTER; THERAPY; RIATA;
D O I
10.1016/j.hrthm.2020.08.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BACKGROUND Pacing leads are the Achilles heel of pacemakers. Most manufacturers report a 3-year survival rate of >99% of their leads. We observed several failures of the Beflex/Vega leads (MicroPort, Shanghai, China; formerly Sorin/Livallova). OBJECTIVE The purpose of this study was to investigate failure rates of Beflex/Vega leads. METHODS We analyzed the performance of Beflex/Vega leads implanted at our tertiary referral center. All-cause lead failures (any issues requiring reinterventions such as lead dislocations, cardiac perforations, and electrical abnormalities) were identified during follow-up. The Beflex/Vega lead was compared with a reference lead (CapSureFix Novus 5076, Medtronic, Minneapolis, MN) implanted within the same period and by the same operators. RESULTS A total of 585 leads were analyzed (382 Beflex/Vega and 203 CapSureFix Novus 5076 leads). Cumulative failure rate estimates were 5.2%, 6.3%, and 12.4% after 1, 2, and 3 years for the Beflex/Vega lead. This was worse compared to the reference lead (1.5%, 1.5%, 3.7% after 1, 2, and 3 years; P = .001). Early failure manifestations up to 3 months occurred at a similar rate (Beflex/Vega vs CapSureFix Novus 5076 lead: 1.3% vs 0.5% for dislocations; 1.3% vs 1.0% for perforations). During follow-up, electrical abnormalities such as noise oversensing (P = .013) and increased pacing thresholds (P = .003) became more frequent in the Beflex/Vega group. Electrical abnormalities were the most common failure manifestation 3 years after implantation in this group (9.4% vs 2.2% for the CapSureFix Novus 5076). CONCLUSION The failure rate of the Beflex/Vega lead of >10% after 3 years was higher than that of a competitor lead. This gives rise to concern since >135,000 such leads are active worldwide.
引用
收藏
页码:41 / 49
页数:9
相关论文
共 31 条
  • [21] Cesarean section rate during the past three years: too early diagnosis of CPD/failure to progress may lead to high section rate?
    Imai, K
    Yagi, H
    Yorifuji, H
    Koda, H
    FIRST WORLD CONGRESS ON CONTROVERSIES IN OBSTETRICS, GYNECOLOGY & INFERTILITY (FREE PAPERS), 1999, : 101 - 103
  • [22] Chronic Vagus Nerve Stimulation Improves Autonomic Control and Attenuates Systemic Inflammation and Heart Failure Progression in a Canine High-Rate Pacing Model
    Zhang, Youhua
    Popovic, Zoran B.
    Bibevski, Steve
    Fakhry, Itaf
    Sica, Domenic A.
    Van Wagoner, David R.
    Mazgalev, Todor N.
    CIRCULATION-HEART FAILURE, 2009, 2 (06) : 692 - 699
  • [23] High Failure Rate of Modular Exchange With a Specific Design of a Constrained Liner in High-Risk Patients Undergoing Revision Total Hip Arthroplasty
    Chalmers, Brian P.
    Arsoy, Diren
    Sierra, Rafael J.
    Lewallen, David G.
    Trousdale, Robert T.
    JOURNAL OF ARTHROPLASTY, 2016, 31 (09): : 1963 - 1969
  • [24] DEVELOPMENT OF A DILATED CARDIOMYOPATHY HEART FAILURE MODEL EMPLOYING HIGH-RATE PACING VIA AN IMPLANTABLE LEADLESS PACEMAKER: UTILIZED FOR IN SITU AND EX VIVO INVESTIGATIONS
    Schinstock, Emma
    Bielecki, Michael
    Kudlik, D'Anne
    Eggen, Michael
    Laizzo, Paul A.
    PROCEEDINGS OF THE 2022 DESIGN OF MEDICAL DEVICES CONFERENCE (DMD2022), 2022,
  • [25] Failure mode of valve-regulated lead-acid batteries under high-rate partial-state-of-charge operation
    Lam, LT
    Haigh, NP
    Phyland, CG
    Urban, AJ
    JOURNAL OF POWER SOURCES, 2004, 133 (01) : 126 - 134
  • [26] High failure rate during oral food challenge to baked milk in children with high serum cow's milk (CM)-Specific IgE antibody level
    Nowak-Wegrzyn, A. H.
    Bloom, K. A.
    Sicherer, S. H.
    Shreffler, W. G.
    Sampson, H. A.
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 2008, 121 (02) : S247 - S247
  • [27] High failure rate for an epicardial implantable cardioverter-defibrillator lead: Implications for long-term follow-up of patients with an implantable cardioverter-defibrillator
    Brady, PA
    Friedman, PA
    Trusty, JM
    Grice, S
    Hammill, SC
    Stanton, MS
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (03) : 616 - 622
  • [28] High error rate in a specific region of the CFTR gene may lead to false variant findings when using massively parallel sequencing in molecular diagnostics of cystic fibrosis
    Kubiritova, Z.
    Budis, J.
    Kucharik, M.
    Szemes, T.
    Radvanszky, J.
    EUROPEAN JOURNAL OF HUMAN GENETICS, 2019, 27 : 1621 - 1621
  • [29] High failure rate for an epicardial cardioverter-defibrillator lead: implications for long term follow-up of patients with an implantable cardioverter-defibrillator (vol 31, pg 616, 1998)
    Brady, PA
    Friedman, PA
    Trusty, JM
    Grice, S
    Hammill, SC
    Stanton, MS
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 32 (02) : 555 - 555
  • [30] Remarkably high and accelerating failure rate of a widely used implantable cardioverter-defibrillator fi brillator lead: A large-scale manufacturer-independent multicenter study with long accurate follow-up
    Oosterwerff, Erik F. J.
    Theuns, Dominic A. M. J.
    Maass, Alexander H.
    van Erven, Lieselot
    HEART RHYTHM, 2024, 5 (09) : 614 - 622