An Analysis of CMRR Degradation in Multi-Channel Biosignal Recording Systems

被引:6
|
作者
Malekzadeh-Arasteh, Omid [1 ]
Danesh, Ahmad Reza [1 ]
Do, An H. [2 ]
Nenadic, Zoran [3 ]
Heydari, Payam [1 ]
机构
[1] Univ Calif Irvine, Dept Elect Engn & Comp Sci, Irvine, CA 92697 USA
[2] Univ Calif Irvine, Dept Neurol, Irvine, CA 92697 USA
[3] Univ Calif Irvine, Dept Biomed Engn, Irvine, CA 92697 USA
关键词
Impedance; Electrodes; Biological system modeling; Degradation; Loading; Integrated circuit modeling; Common mode rejection ratio; multi-channel; biosignal acquisition; shared reference; impedance mismatch; CMRR degradation; MODE REJECTION RATIO; INTERFERENCE; AMPLIFIER;
D O I
10.1109/TCSII.2020.3011180
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Multi-channel biosignal recording systems that employ a shared-reference scheme and achieve high common-mode rejection ratio (i.e., CMRR > 80dB) have been reported in recent works. While it is well-understood that a shared-reference scheme causes impedance mismatch at the input terminals of bioamplifier, and thus limits the maximum achievable CMRR, a theoretical study that can provide quantitative assessment of this source of degradation is still lacking. This brief provides an equivalent electrical circuit model of the input interface consisting of an electrode array and bioamplifiers, followed by a complete analysis to formulate the CMRR degradation. Simulation results based on a previously designed and fabricated 32-channel neural recording front-end in a 180nm CMOS process are presented, exhibiting close agreement with the theory.
引用
收藏
页码:151 / 155
页数:5
相关论文
共 50 条
  • [21] A 220nW Neural Amplifier for Multi-Channel Neural Recording Systems
    Kim, Jungsuk
    Chae, Moo Sung
    Liu, Wentai
    ISCAS: 2009 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS, VOLS 1-5, 2009, : 1257 - 1260
  • [22] QUEUING ANALYSIS OF MULTI-CHANNEL TACTICAL RADIO SYSTEMS
    MARKS, BI
    JANC, RV
    THOMOPOULOS, NT
    OPERATIONS RESEARCH, 1975, 23 : B300 - B300
  • [23] A multi-channel analog IC for in vitro neural recording
    袁丰
    王志功
    吕晓迎
    Journal of Semiconductors, 2016, 37 (02) : 146 - 151
  • [24] DEVELOPMENT OF MULTI-CHANNEL INTRAMUSCULAR EMG RECORDING ELECTRODES
    Poppendieck, Wigand
    Muceli, Silvia
    Welsch, Christine
    Krob, Marc-Oliver
    Sossalla, Adam
    Yoshida, Ken
    Farina, Dario
    Hoffmann, Klaus-Peter
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2013, 58
  • [25] A multi-channel analog IC for in vitro neural recording
    Yuan Feng
    Wang Zhigong
    Lu Xiaoying
    JOURNAL OF SEMICONDUCTORS, 2016, 37 (02)
  • [26] SIMPLE MULTI-CHANNEL DEVICE FOR RECORDING DEWATERABILITY MEASUREMENTS
    BELL, C
    SAUNDERS, MA
    EFFLUENT & WATER TREATMENT JOURNAL, 1971, 11 (10): : 554 - &
  • [27] An implantable microsystem for wireless multi-channel cortical recording
    Sodagar, Amir M.
    Perlin, Gayatri E.
    Yao, Ying
    Wise, Kensall D.
    Najafi, Khalil
    TRANSDUCERS '07 & EUROSENSORS XXI, DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, 2007,
  • [28] AUTOMATIC MULTI-CHANNEL SENSING AND RECORDING OF ANIMAL BEHAVIOR
    KAVANAU, JL
    ECOLOGY, 1962, 43 (01) : 161 - &
  • [29] A high precision analog multiplexer for multi-channel neural recording micro-systems
    Naghavi, Saeed
    Sharifi, Niloofar
    Abrishamifar, Adib
    MICROELECTRONICS JOURNAL, 2018, 71 : 8 - 18
  • [30] Entropy Analysis of Support Systems in Multi-Channel Cryogenic Lines
    Duda, P.
    Chorowski, M.
    Polinski, J.
    ADVANCES IN CRYOGENIC ENGINEERING, 2020, 755